Acta Neurochir (Wien) ; 160(2): 385-387, 2018 02.
Artículo en Inglés | MEDLINE | ID: mdl-29116383

RESUMEN

Deep brain stimulation for Parkinson's disease has become an established treatment option in recent years. The method and its application in clinical practice has proved to be safe and effective. Nevertheless, procedure-related and hardware-related complications occur. We present a rare case of a patient with an allergic reaction to the impulse generator. The patient suffered from delayed wound-healing deficits with several wound revisions and generator repositionings. After diagnosis of an allergic reaction to components of the generator, a custom-made silicon-coated model was implanted. Hereafter, no wound healing-deficit occurred throughout long-term follow-up. Allergic reaction to hardware components may lead to wound-healing deficits. In such cases, custom-made silicon-coated models may be an effective treatment option.


Asunto(s)
Estimulación Encefálica Profunda/efectos adversos , Electrodos Implantados/efectos adversos , Hipersensibilidad/etiología , Enfermedad de Parkinson/cirugía , Complicaciones Posoperatorias/etiología , Estimulación Encefálica Profunda/instrumentación , Estimulación Encefálica Profunda/métodos , Electrodos Implantados/normas , Humanos , Masculino , Persona de Mediana Edad , Siliconas