Appl Radiat Isot ; 208: 111307, 2024 Jun.
Article En | MEDLINE | ID: mdl-38564840

Early works that used thermoluminescent dosimeters (TLDs) to measure absorbed dose from alpha particles reported relatively high variation (10%) between TLDs, which is undesirable for modern dosimetry applications. This work outlines a method to increase precision for absorbed dose measured using TLDs with alpha-emitting radionuclides by applying an alpha-specific chip factor (CF) that individually characterizes the TLD sensitivity to alpha particles. Variation between TLDs was reduced from 21.8% to 6.7% for the standard TLD chips and 7.9% to 3.3% for the thin TLD chips. It has been demonstrated by this work that TLD-100 can be calibrated to precisely measure the absorbed dose to water from alpha-emitting radionuclides.


Radiation Dosimeters , Thermoluminescent Dosimetry , Thermoluminescent Dosimetry/methods , Radioisotopes , Radiometry/methods , Calibration
...