Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros










Base de datos
Asunto principal
Intervalo de año de publicación
1.
J Assist Reprod Genet ; 36(11): 2271-2278, 2019 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-31650453

RESUMEN

PURPOSE: To determine liquid nitrogen evaporation rates of intact liquid nitrogen storage tanks and tanks with their vacuum removed. METHODS: Donated storage tank performance (LN2 evaporation) was evaluated before and after induced vacuum failure. Vacuum of each tank was removed by drilling through the vacuum port. Temperature probes were placed 2 in. below the bottom of the styrofoam cap/plug, and tanks were weighed every 3 h. Evaporation rate and time from failure to the critical temperature was determined. RESULT: Storage tanks with failed vacuum have a much higher evaporation rate than those with intact vacuum; evaporation rates increased dramatically within 3 to 6 h in the smaller tanks, and time to complete depletion varied according to starting LN2 volume. Tanks with storage racks/specimens may have altered evaporation profiles compared to tanks without. Locating temperature probes 2 in. below the styrofoam cap/plug suggests that for most applications, alarms would sound approximately 1 h prior to reaching the critical warming temperature, approximately - 130 °C. External signs of vacuum loss were dramatic: vapor, frost, and audible movement of air. CONCLUSION: For the first time, we have data on how liquid nitrogen storage tanks behave when their vacuum is removed. These findings are conservative; each lab must consider starting volume, tank size/capacity, function (storage or shipping), age, and pre-existing evaporation behavior in order to develop an emergency response to critical tank failure. Times to complete failure/evaporation and critical warming temperature after vacuum loss are different; these data should be considered when evaluating tank alarm systems.


Asunto(s)
Criopreservación/métodos , Crioprotectores/química , Humanos , Vehículos a Motor , Nitrógeno/química , Temperatura , Vacio
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...