J Heart Valve Dis ; 25(4): 508-511, 2016 07.
Artículo en Inglés | MEDLINE | ID: mdl-28009960

RESUMEN

A 59-year-old male with a history of mitral valve replacement several years previously was admitted to the authors' institution with symptoms of advanced heart failure. Echocardiography showed a severe paravalvular dehiscence and surgery was discouraged due to high perioperative risk. A first transcatheter leak closure was unsuccessful despite the correct deployment of various occluding devices. A second percutaneous attempt was carried out with implantation of three Amplatzer Vascular Plug® III devices (five occluders were positioned in total within the oblong defect), leading to a mild residual leakage. Device embolism occurred twice during the procedure but was resolved using adequate catheter retrieval techniques.


Asunto(s)
Insuficiencia de la Válvula Mitral/cirugía , Dispositivo Oclusor Septal , Ecocardiografía Transesofágica , Implantación de Prótesis de Válvulas Cardíacas , Humanos , Masculino , Persona de Mediana Edad , Válvula Mitral/cirugía , Insuficiencia de la Válvula Mitral/diagnóstico por imagen , Insuficiencia de la Válvula Mitral/etiología , Falla de Prótesis , Factores de Riesgo , Factores de Tiempo