Your browser doesn't support javascript.
loading
: 20 | 50 | 100
1 - 20 de 33
1.
J Med Eng Technol ; 47(2): 141-146, 2023 Feb.
Article En | MEDLINE | ID: mdl-36426804

The mechanical properties of the myocardium in the left ventricle and right atrium were estimated by simultaneously measuring the local impedance (LI) and contact force (CF) using an ablation catheter. Radiofrequency catheter ablation (RFCA) is a well-established arrhythmia treatment. Monitoring the RF power, CF and properties of myocardium during RFCA are necessary to estimate the effect of ablation. Indices, such as CF, lesion size index and ablation index, do not include the myocardium mechanical properties. Therefore, there is the risk of side effects, such as cardiac tamponade, by excessive catheter indentation into vulnerable areas. We propose the simultaneous measurement of LI and CF for estimating the myocardial mechanical properties to reduce the side effects. In this study, an in vitro experimental system was constructed to measure LI and CF via the catheter. The relationship between the porcine myocardial tissue thickness and CF-LI curve was investigated using the left ventricle and right atrium. Power function coefficients approximating the CF-LI curve increased with thicker left ventricle. The thickness of the myocardium can be estimated by simultaneously measuring LI and CF. Intraoperative measurement of the myocardial mechanical properties can be used to determine the ablation conditions at each site.


Arrhythmias, Cardiac , Catheter Ablation , Swine , Animals , Electric Impedance , Heart Ventricles/surgery , Heart Ventricles/pathology , Catheters , Catheter Ablation/adverse effects
2.
J Interv Card Electrophysiol ; 66(2): 485-492, 2023 Mar.
Article En | MEDLINE | ID: mdl-36074285

BACKGROUND: The lesion size index (LSI) predicts radiofrequency (RF) ablation lesion size and is an established parameter for pulmonary vein isolation. However, the effectiveness and safety of LSI for cavotricuspid isthmus (CTI) linear ablation remain unclear. METHODS: This single-center retrospective study included 50 of patients (67 ± 10 years, 68% male) who underwent de novo CTI linear ablation between July 2020 and December 2020. The LSI target was set at 5.0 and 4.0 for the anterior 2/3 and posterior 1/3 segments, respectively. Acute procedural parameters of ablation were evaluated. RESULTS: Acute bidirectional CTI block was achieved in all patients with an RF application time of 4.0 min (3.1-5.0 min), RF application number of 15 ± 7, and length of CTI of 36.9 ± 9.3 mm. First-pass bidirectional conduction block of the CTI was achieved in 39/50 (78%) patients. No major complications were observed. The contact force (CF) per application was significantly lower in the gap tag group than in the non-gap tag group (7 g [7-8 g] vs. 10 g [7-12 g], P = 0.0284). CONCLUSIONS: LSI-guided CTI linear ablation is an effective and safe treatment approach. CF affects gap formation, even when the target LSI is the same.


Atrial Flutter , Catheter Ablation , Humans , Male , Female , Atrial Flutter/surgery , Treatment Outcome , Retrospective Studies , Catheter Ablation/adverse effects , Tricuspid Valve/surgery
3.
J Arrhythm ; 38(6): 991-996, 2022 Dec.
Article En | MEDLINE | ID: mdl-36524028

Background: Previous studies have identified noninvasive methods for predicting atrial fibrillation (AF) recurrence after catheter ablation (CA). We assessed the association between AF recurrence and atrial late potentials (ALPs), which were measured using P-wave signal-averaged electrocardiography (P-SAECG). Methods: Consecutive patients with paroxysmal AF who underwent their first CA at our institution between August 2015 and August 2019 were enrolled. P-SAECG was performed before CA. Two ALP parameters were evaluated: the root-mean-square voltage during the terminal 20 ms (RMS20) and the P-wave duration (PWD). Positive ALPs were defined as an RMS20 <2.2 µV and/or a PWD >115 ms. Patients were allocated to either the recurrence or nonrecurrence group based on the presence of AF recurrence at the 1-year follow-up post-CA. Results: Of the 190 patients (age: 65 ± 11 years, 37% women) enrolled in this study, 21 (11%) had AF recurrence. The positive ALP rate was significantly higher in the recurrence group than in the nonrecurrence group (86% vs. 64%, p = .04), despite the absence of differences in other baseline characteristics between the two groups. In the multivariate analysis, positive ALP was an independent predictor of AF recurrence (odds ratio: 3.83, 95% confidence interval: 1.05-14.1, p = .04). Conclusions: Positive ALP on pre-CA P-SAECG is associated with AF recurrence after CA.

4.
Pediatr Surg Int ; 39(1): 23, 2022 Nov 30.
Article En | MEDLINE | ID: mdl-36449184

PURPOSE: Although the impairment of regulatory T-cells (Tregs) has been shown in the liver or portal area of biliary atresia (BA) the frequency and function of circulating Tregs in BA patients is poorly understood. We aimed to investigate the frequency and function of circulating Tregs in BA patients. METHODS: Peripheral blood mononuclear cells were collected from 25 BA patients and 24 controls. Treg frequency was measured by flow cytometry; function was determined by T-cell proliferation assay. We also assessed the association between Treg frequency/function and clinical parameters in BA cases. RESULTS: There was no significant difference between the two groups in both frequency (BA: 3.4%; control: 3.2%; p = 0.97) and function (BA: 22.0%; control: 7.5%; p = 0.23) of Tregs. We further focused on 13 preoperative BA patients and 14 age-matched controls. Neither Treg frequency nor function were significantly different (frequency: BA: 4.6%; control: 3.4%; p = 0.38, function: BA: 2.7%; control: 7.6%; p = 0.89). There was no association between Treg frequency/function and clinical parameters. CONCLUSION: Neither the frequency nor function of circulating Tregs was affected in BA patients, suggesting the negative role of circulating Tregs in the pathogenesis of BA. Further investigation of local Treg profiles is warranted.


Biliary Atresia , Humans , Biliary Atresia/surgery , T-Lymphocytes, Regulatory , Leukocytes, Mononuclear , Liver , Flow Cytometry
5.
J Cardiol Cases ; 26(5): 371-374, 2022 Nov.
Article En | MEDLINE | ID: mdl-36312773

A 48-year-old male with ischemic cardiomyopathy was admitted because of cardiogenic shock due to incessant ventricular tachycardia (VT). Despite the introduction of multiple mechanical cardiac support (MCS) devices, his hemodynamics were unstable; therefore, VT ablation was performed. Due to the limited vascular and left ventricle (LV) access with the multiple MCS devices, only the left femoral artery and vein were accessible. In addition, the transseptal approach as an LV access was difficult because of the MitraClip (Abbott Park, IL, USA). Posterior papillary muscle-origin VT was successfully treated via a transaortic approach under the MCS. After the VT ablation, the patient recovered from the MCS. Learning objective: Mechanical cardiac support (MCS) devices are critical during ventricular tachycardia (VT) ablation in patients with left ventricular (LV) dysfunction. However, the ablation strategy for cases with limited access due to multiple MCS devices has not been established. In this case, VT was successfully treated via a transaortic approach using intracardiac echocardiography. The number of cases requiring VT ablation for patients with limited vascular and LV endocardial access would increase; therefore, the accumulation of those cases is required to find a better strategy.

7.
J Cardiovasc Electrophysiol ; 33(6): 1160-1166, 2022 06.
Article En | MEDLINE | ID: mdl-35488745

INTRODUCTION: Direct comparison studies about the incidence of esophagogastric complications between radiofrequency (RF) and cryoballoon (CB) catheter ablation (CA) for atrial fibrillation (AF) have been scarce. We sought to elucidate the relationship between the pulmonary vein isolation (PVI) modalities and esophagogastric complications. METHODS: The study population consisted of 254 patients who underwent CA for AF from November 2017 to October 2018. Finally, 160 patients were enrolled and divided into the RF and CB groups. Esophageal ulcers, gastric hypomotility, and exfoliative esophagitis detected by esophagogastroduodenoscopy were defined as esophagogastric complications in this study. RESULTS: The median age was 68 years old, with 34% being females. Esophagogastric complications were observed in 42.5% of patients who underwent CA. According to the detailed esophagogastric complications, the RF group had a higher prevalence of esophageal ulcers than the CB group (19% vs. 0%, p < .0001). There was no significant difference between the two groups regarding gastric hypomotility and exfoliative esophagitis (18% vs. 28%; p = .15 and 16% vs. 21%; p = .42, respectively). CONCLUSION: Asymptomatic esophagogastric complications were common in CA for AF. The incidence of esophageal ulcers was higher in the RF group than in the CB group, whereas the other esophagogastric complications did not significantly differ.


Atrial Fibrillation , Catheter Ablation , Cryosurgery , Esophagitis , Pulmonary Veins , Aged , Atrial Fibrillation/diagnosis , Atrial Fibrillation/epidemiology , Atrial Fibrillation/surgery , Catheter Ablation/adverse effects , Cryosurgery/adverse effects , Esophagitis/etiology , Esophagitis/surgery , Female , Humans , Male , Pulmonary Veins/surgery , Recurrence , Treatment Outcome , Ulcer/etiology , Ulcer/surgery
8.
J Arrhythm ; 38(1): 160-162, 2022 Feb.
Article En | MEDLINE | ID: mdl-35222764

We performed cavotricuspid isthmus (CTI) linear ablation for atrial flutter; however, the tachycardia cycle length was not changed at all. In such cases, repeated or broad line ablation is usually performed. We presented that high-density three-dimensional mapping after the first CTI linear ablation, which revealed the complex tachycardia circuit with the epicardial and endocardial breakthrough.

9.
J Cardiovasc Electrophysiol ; 33(3): 380-388, 2022 03.
Article En | MEDLINE | ID: mdl-35018687

BACKGROUND: Local impedance (LI) can indirectly measure catheter contact and tissue temperature during radiofrequency catheter ablation (RFCA). However, data on the effects of catheter contact angle on LI parameters are scarce. This study aimed to evaluate the influence of catheter contact angle on LI changes and lesion size with two different LI-sensing catheters in a porcine experimental study. METHODS: Lesions were created by the INTELLANAV MiFi™ OI (MiFi) and the INTELLANAV STABLEPOINT™ (STABLEPOINT). RFCA was performed with 30 W and a duration of 30 s. The contact force (CF) (0, 5, 10, 20, and 30 g) and catheter contact angle (30°, 45°, and 90°) were changed in each set (n = 8 each). The LI rise, LI drop, and lesion size were evaluated. RESULTS: The LI rise increased as CF increased. There was no angular dependence with the LI rise under all CFs in the MiFi. On the other hand, the LI rise at 90° was lower than at 30° under 5 and 10 g of CF in STABLEPOINT. The LI drop increased as CF increased. Regarding the difference in catheter contact angles, the LI drop at 90° was lower than that at 30° for both catheters. The maximum lesion widths and surface widths were smaller at 90° than at 30°, whereas there were no differences in lesion depths. CONCLUSION: The LI drop and lesion widths at 90° were significantly smaller than those at 30°, although the lesion depths were not different among the 3 angles for the MiFi and STABLEPOINT.


Catheter Ablation , Animals , Catheter Ablation/adverse effects , Catheters , Electric Impedance , Equipment Design , Swine
10.
Pediatr Rep ; 14(1): 20-25, 2022 Jan 04.
Article En | MEDLINE | ID: mdl-35076592

The clinical features and risk factors of acute appendicitis in infants are unclear. Our aim was to evaluate the association between anthropometrics and the occurrence of infantile appendicitis. This was a retrospective study of infants (<6 years of age) and school-age children (6-10 years of age) of Asian ethnicity who required hospitalization for appendicitis at our two participating institutions between 2004 and 2018. The Z-score for height, body weight, and body mass index (BMI) was compared between the two groups, as well as between patients presenting with perforated and non-perforated appendicitis. The analysis included data from 73 infants and 362 school-age children. Z-scores were greater in infants than in school-age children for height (0.37 versus -0.03, p = 0.003) and body weight (0.12 versus -0.36, p = 0.023), with no between-group difference for the Z-score of BMI. There was no difference in Z-scores for height, weight, and BMI between the perforated and non-perforated appendicitis infant groups. Infants presenting with acute appendicitis were characterized by a larger physique but with normal proportion. This trend was not observed in school-age children. Therefore, larger infants presenting with abdominal pain should be screened for appendicitis.

11.
Heart Vessels ; 37(4): 628-637, 2022 Apr.
Article En | MEDLINE | ID: mdl-34613425

The recurrence of atrial fibrillation (AF) after catheter ablation (CA) is still an unsolved issue. Although structural remodeling is relatively well defined, the method to assess electrical remodeling of the atrium is not well established. In this study, we evaluated the relationship between atrial conduction properties and recurrence after CA for AF. One hundred six consecutive patients (66 ± 11 years old, male: 68%) who underwent CA for AF with a CARTO system from July 2016 to July 2019 were enrolled in this study. An activation map of both atria was constructed to precisely evaluate the total conduction time, distance, and conduction velocity between the earliest and latest activation sites during sinus rhythm. All parameters were compared between the patients with or without AF recurrence. Of the patients, 27 had an AF recurrence (Rec group). The left atrial (LA) conduction velocity was significantly slower in the Rec group than in the non-Rec group (101.2 ± 17.9 vs. 116.9 ± 18.0 cm/s, P < 0.01). Likewise, the right atrial (RA) conduction velocity was significantly slower in the Rec group than in the non-Rec group (81.1 ± 17.5 vs. 103.6 ± 25.4 cm/s, P < 0.01). A multivariate logistic analysis demonstrated that the LA and RA conduction velocities were independent predictors of AF recurrence, with adjusted odds ratios of 0.95 (95% confidential interval: 0.91-0.98, P < 0.01) and 0.94 (0.89-0.98, P < 0.01), respectively. In conclusion, slower conduction velocity of the atrium was associated with AF recurrence after pulmonary vein isolation.


Atrial Fibrillation , Catheter Ablation , Pulmonary Veins , Aged , Atrial Fibrillation/complications , Atrial Fibrillation/diagnosis , Atrial Fibrillation/surgery , Catheter Ablation/adverse effects , Catheter Ablation/methods , Heart Atria , Humans , Male , Middle Aged , Pulmonary Veins/surgery , Recurrence , Treatment Outcome
12.
Circ Rep ; 3(9): 497-503, 2021 Sep 10.
Article En | MEDLINE | ID: mdl-34568628

Background: The incidence of new-onset atrial high-rate episode (AHRE) is higher among patients with cardiac implantable electronic devices (CIEDs) than in the general population. We sought to elucidate the clinical factors associated with AHRE in CIED patients, including P-wave dispersion (PWD) in sinus rhythm. Methods and Results: In all, 101 patients with CIEDs newly implanted between 2010 and 2014 were included in the study. PWD was measured at the time of device implantation via a body-surface electrocardiogram. AHRE was defined as any episode of sustained atrial tachyarrhythmia (>170 beats/min) recorded in the device's memory. Patients were divided into an AHRE (n=34) and non-AHRE (n=67) group based on the presence or absence of AHRE within 1 year of device implantation and compared. Mean (±SD) patient age was 75±11 years. A greater incidence of sick sinus syndrome (P=0.05) and longer PWD (62.6±13.1 vs. 38.2±13.9 ms; P<0.0001) were apparent in the AHRE than non-AHRE group. Multivariate analysis revealed that PWD was an independent predictor of new-onset AHRE (odds ratio 1.11; 95% confidence interval 1.06-1.17; P<0.0001). In logistic regression analysis, receiver-operating characteristic curve analysis (area under the curve 0.90; P<0.001) suggested the best cut-off value for PWD was 48 mm (sensitivity 73.8%, specificity 77.9%). Conclusions: PWD is a simple but feasible predictor of new-onset AHRE in patients with CIEDs.

13.
J Arrhythm ; 37(3): 558-565, 2021 Jun.
Article En | MEDLINE | ID: mdl-34141008

BACKGROUND: Although the lesion size index (LSI) has been well established, it is sometimes difficult to achieve first-pass pulmonary vein isolation (PVI) and to avoid acute pulmonary vein reconnections, even with LSI-guided procedures. The purpose of this study was to assess the predictive accuracy of a novel parameter, the optimized lesion size index (o-LSI), to perform PVI. METHODS: The voltage maps created by the Advisor™ high-density (HD) grid catheter before PVI in 35 atrial fibrillation (AF) patients were examined for an association between the voltage amplitude and insufficient ablation sites (IAS), which were defined as either (i) spontaneous reconnection sites, (ii) dormant PV conduction sites unmasked with 20 mg of adenosine triphosphate disodium hydrate (ATP) injection, or (iii) PV-LA gap sites after the initial PVI. RESULTS: IAS was observed in 25/1417 of the total ablation sites. IAS was significantly associated with higher bipolar voltage areas (4.20 ± 2.68 vs 2.43 ± 1.93 mV, P < .0001) but not with LSI. A novel index, o-LSI (defined as LSI/bipolar voltage), was significantly lower in IAS than in others (1.14 [0.82, 1.81] vs 2.35 [1.31, 4.80] LSI/mV). By receiver operating characteristic analysis, an o-LSI of 2.04 was the best cutoff value for the prediction of IAS (88% sensitivity and 55% specificity, P < .0001, areas under the curve: 0.742). CONCLUSION: Low o-LSI was strongly associated with IAS, potentially providing a novel index to improve first-pass PV isolation.

14.
J Arrhythm ; 37(1): 203-211, 2021 Feb.
Article En | MEDLINE | ID: mdl-33664904

BACKGROUND: Right ventricular (RV) pacing causes left ventricular (LV) dyssynchrony sometimes resulting in pacing-induced cardiomyopathy. However, RV pacing for hypertrophic obstructive cardiomyopathy is one of the treatment options. LV flow energy loss (EL) using vector flow mapping (VFM) is a novel hemodynamic index for assessing cardiac function. Our study aimed to elucidate the impact of RV pacing on EL in normal LV function and hypertrophic cardiomyopathy (HCM) patients. METHODS: A total of 36 patients with dual-chamber pacemakers for sick sinus syndrome or implantable cardioverter defibrillators for fatal ventricular tachyarrhythmias were enrolled. All patients were divided into two groups: 16 patients with HCM (HCM group) and others (non-HCM group). The absolute changes in EL under AAI (without RV pacing) and DDD (with RV pacing) modes were assessed using VFM on color Doppler echocardiography. RESULTS: In the non-HCM group, the mean systolic EL significantly increased from the AAI to DDD modes (14.0 ± 7.7 to 17.0 ± 8.6 mW/m, P = .003), whereas the mean diastolic EL did not change (19.0 ± 12.3 to 17.0 ± 14.8 mW/m, P = .231). In the HCM group, the mean systolic EL significantly decreased from the AAI to DDD modes (26.7 ± 14.2 to 21.6 ± 11.9 mW/m, P < .001), whereas the mean diastolic EL did not change (28.7 ± 16.4 to 23.9 ± 19.7 mW/m, P = .130). CONCLUSIONS: RV pacing increased the mean systolic EL in patients without HCM. Conversely, RV pacing decreased the mean systolic EL in patients with HCM.

15.
J Am Heart Assoc ; 10(2): e017483, 2021 01 19.
Article En | MEDLINE | ID: mdl-33399004

Background Recent clinical trials have demonstrated the possible pleiotropic effects of SGLT2 (sodium-glucose cotransporter 2) inhibitors in clinical cardiovascular diseases. Atrial electrical and structural remodeling is important as an atrial fibrillation (AF) substrate. Methods and Results The present study assessed the effect of canagliflozin (CAN), an SGLT2 inhibitor, on atrial remodeling in a canine AF model. The study included 12 beagle dogs, with 10 receiving continuous rapid atrial pacing and 2 acting as the nonpacing group. The 10 dogs that received continuous rapid atrial pacing for 3 weeks were subdivided as follows: pacing control group (n=5) and pacing+CAN (3 mg/kg per day) group (n=5). The atrial effective refractory period, conduction velocity, and AF inducibility were evaluated weekly through atrial epicardial wires. After the protocol, atrial tissues were sampled for histological examination. The degree of reactive oxygen species expression was evaluated by dihydroethidium staining. The atrial effective refractory period reduction was smaller (P=0.06) and the degree of conduction velocity decrease was smaller in the pacing+CAN group compared with the pacing control group (P=0.009). The AF inducibility gradually increased in the pacing control group, but such an increase was suppressed in the pacing+CAN group (P=0.011). The pacing control group exhibited interstitial fibrosis and enhanced oxidative stress, which were suppressed in the pacing+CAN group. Conclusions CAN and possibly other SGLT2 inhibitors might be useful for preventing AF and suppressing the promotion of atrial remodeling as an AF substrate.


Atrial Fibrillation , Atrial Remodeling/drug effects , Canagliflozin/pharmacology , Heart Atria , Oxidative Stress/drug effects , Sodium-Glucose Transporter 2/metabolism , Animals , Atrial Fibrillation/metabolism , Atrial Fibrillation/pathology , Atrial Fibrillation/physiopathology , Dogs , Electrophysiologic Techniques, Cardiac/methods , Heart Atria/pathology , Heart Atria/physiopathology , Heart Conduction System/metabolism , Heart Conduction System/physiopathology , Reactive Oxygen Species/analysis , Sodium-Glucose Transporter 2 Inhibitors/pharmacology , Treatment Outcome
16.
J Interv Card Electrophysiol ; 62(1): 123-131, 2021 Oct.
Article En | MEDLINE | ID: mdl-32975734

PURPOSE: Ablation index (AI) is a useful tool of the CARTO® system to make effective lesions during pulmonary vein isolation (PVI) for atrial fibrillation (AF). However, the optimal distance between neighboring ablation points (interlesion distance (ILD)) is still unclear. Here, we evaluated the optimal ILDs in the AI-guided PVI. METHODS: Forty-nine AF patients who underwent AI-guided PVI in our institute from July 2018 to March 2019 were retrospectively enrolled in this study. Target AI was set at 500 and 400 for anterior and posterior walls, respectively, and we compared the ILDs with and without electrical gaps after a first encircling PVI. RESULTS: In both PV, the ILDs with electrical gaps were longer than those without electrical gaps. The best cutoff values of ILD to detect the electrical gaps using the ROC curve were 5.4 mm for the RPV anterior wall (AUC, 0.67; sensitivity, 0.42; specificity, 0.84, P < 0.01) and 4.4 mm for the RPV posterior wall (AUC, 0.68; sensitivity, 0.91; specificity, 0.39, P < 0.01). Similarly, the best cutoff values of ILD were 5.5 mm for the LPV anterior wall (AUC, 0.74; sensitivity, 0.65; specificity, 0.82, P < 0.01) and 5.1 mm for the LPV posterior wall (AUC, 0.67; sensitivity, 0.79; specificity, 0.53, P =0.03). CONCLUSION: The optimal interlesion distances for PVI were different in each PV segment. To achieve the first-pass isolation, less than 5.4/4.4 mm for the RPV anterior/posterior and 5.5/5.1 mm for the LPV anterior/posterior walls of interlesion distances were the best cutoff values in the patients with AF.


Atrial Fibrillation , Catheter Ablation , Pulmonary Veins , Atrial Fibrillation/diagnostic imaging , Atrial Fibrillation/surgery , Humans , Pulmonary Veins/diagnostic imaging , Pulmonary Veins/surgery , Recurrence , Retrospective Studies , Treatment Outcome
17.
Int J Cardiol ; 326: 103-108, 2021 03 01.
Article En | MEDLINE | ID: mdl-33130261

BACKGROUND: Gastric hypomotility (GH) is a possible complication of catheter ablation (CA) for atrial fibrillation (AF). However, it is unclear which factors are associated with GH. We sought to elucidate the relationship between the CA procedure and GH. METHODS: The study population consisted of 254 patients who underwent CA for AF from November 2017 to October 2018. Finally, 119 patients were enrolled and divided into two groups: with or without GH (GH or non-GH groups). To evaluate the association with GH, the clinical backgrounds and procedure characteristics of the radiofrequency CA (RFCA) were compared between the two groups. RESULTS: The median age was 69 years old with 34% of female. GH were observed in 27.7% of patients who underwent RFCA, which was significantly higher than that in the cohort of patients who underwent esophago-gastro-duodenoscopy during the same time period (1.9%: 151 in 8063 patients, p < 0.0001). According to the detailed RFCA procedure, additional posterior wall isolation with pulmonary vein isolation (PVI) had a higher prevalence of GH than that with only PVI (54.8% vs. 18.2%; odds ratio 5.46, 95%CI 2.24-13.32, p = 0.0002). After an adjustment using a multivariate logistic analysis, a posterior wall isolation with the PVI was identified as the only independent predictor for GH (odds ratio 5.01, 95%CI 1.94-13.43, p = 0.0009). CONCLUSIONS: Additional posterior wall isolation with PVI was associated with gastric hypomotility.


Atrial Fibrillation , Catheter Ablation , Pulmonary Veins , Aged , Atrial Fibrillation/diagnostic imaging , Atrial Fibrillation/epidemiology , Catheter Ablation/adverse effects , Female , Humans , Pulmonary Veins/surgery , Recurrence , Treatment Outcome
18.
Int Heart J ; 61(5): 927-935, 2020 Sep 29.
Article En | MEDLINE | ID: mdl-32879263

We prospectively collected device and heart rate data through remote monitoring (RM) of patients with an implantable cardioverter defibrillator (ICD). The objective was to identify the predictors of lethal arrhythmic events (VT/VF).Thirty-three patients (mean age: 50 years) with ICDs [with functionality of heart rate variability (HRV) analysis] were divided into two groups [VT/VF (+), VT/VF (-) ]. Clinical, device (ventricular lead impedance; amplitude of ventricular electrogram), and HRV data were compared between the two groups. The NN interval-index (SDNNi) was calculated for every 5 minutes, and the mean, maximum, minimum, and standard deviation of SDNNi during the 24-hour period were used.During the observation period of 13 ± 10 months, 10 patients experienced VT/VF events. Total mean, max, and min SDNNi were higher in the VT/VF (+) than the VT/VF (-) group (132.9 ± 9.3 versus 93.5 ± 6.1, P = 0.0013; 214.6 ± 10.6 versus 167.0 ± 7.0, P = 0.0007; 71.2 ± 7.5 versus 43.9 ± 4.9, P = 0.0047). On logistic regression analysis, a total mean SDNNi of 100.1, max SDNNi of 185.0 and min SDNNi of 52.0 as cut-off values for prediction of a VT/VF event demonstrated significant receiver operating characteristic (ROC) curves (AUC = 0.86, P = 0.0007; AUC = 0.84, P = 0.0005; AUC = 0.78, P = 0.0030). The max ΔSDNNi, i.e., difference from baseline SDNNi, and min ΔSDNNi in 7 and 28 days preceding VT/VF events were significant predictors of VT/VF events.Time-domain HRV analysis through a RM system may help identify patients at high risk of lethal arrhythmic events; in addition, it may help predict the occurrence of lethal arrhythmic events in specific cases.


Death, Sudden, Cardiac/prevention & control , Defibrillators, Implantable , Heart Rate , Tachycardia, Ventricular/epidemiology , Ventricular Fibrillation/epidemiology , Adult , Aged , Brugada Syndrome/physiopathology , Cardiomyopathies/physiopathology , Female , Humans , Long QT Syndrome/physiopathology , Male , Middle Aged , Myocardial Ischemia/physiopathology , Remote Sensing Technology , Tachycardia, Ventricular/physiopathology , Ventricular Fibrillation/physiopathology
19.
Int Heart J ; 61(4): 838-842, 2020 Jul 30.
Article En | MEDLINE | ID: mdl-32684601

Complex atrial tachycardias (ATs) after catheter ablation or a MAZE procedure is sometimes difficult to determine the circuits of the tachycardia. A high-density, grid-shapes mapping catheter has been launched, which can be useful for detecting the detail circuits of tachycardias on three-dimensional mapping systems. The signal quality is also important for performing electrophysiological studies (EPSs), such as entrainment mapping, to identify the circuit. This unique mapping catheter has 1 mm electrodes on 2.5 Fr shafts, which improve the signal quality. The high-quality intracardiac electrograms facilitate differentiating small critical potentials, which allows us to perform detailed entrainment mapping in targeted narrow areas. Here, we describe a patient with a perimetral AT with epi-endocardium breakthrough after a MAZE surgery and catheter ablation, which was treated successfully along with detailed entrainment mapping using the HD Grid. This catheter with high-quality signals could be a significant diagnostic tool for a classic EPS as well as for the construction of 3D mapping.


Cardiac Catheters , Electrophysiologic Techniques, Cardiac/instrumentation , Postoperative Complications/diagnostic imaging , Tachycardia, Supraventricular/diagnostic imaging , Humans , Male , Maze Procedure , Middle Aged
20.
Circ Rep ; 2(3): 137-142, 2020 Feb 20.
Article En | MEDLINE | ID: mdl-33693220

Background: The wearable cardioverter defibrillator (WCD) has been available since 2014 in Japan, and its benefit in the in-hospital acute phase at high risk of ventricular tachyarrhythmia (VTA) has been established, but its clinical use in the outpatient setting remains unclear, especially in Japan. Methods and Results: The subjects consisted of 43 consecutive patients with WCD use in the outpatient setting from April 2014 to October 2019 at the present institute. Event alerts and wearing compliance were checked via the remote monitoring system, and a dedicated WCD training team contacted the patients if necessary. The median observation period was 51 days (IQR, 37-68 days) and the median daily wearing time was 23.1 h/day (IQR, 22.0-23.6 h/day). WCD was prescribed for primary prevention of VTA in 7 patients (16%), and for secondary prevention in 36 (84%). The common reason for WCD use was preventive therapy and/or clinical observation. Two appropriate and one inappropriate shock were observed. Eleven patients were not indicated for ICD because of successful catheter ablation optimal medical therapy, VTA in early onset of heart disease and refusal. The remaining 32 patients, however, underwent ICD implantation. Conclusions: In the present real-world study, the WCD wearing compliance was well-maintained in the outpatient setting. WCD is useful for patients at high risk of VTA.

...