1.
Interv Neuroradiol ; : 15910199241245156, 2024 Apr 05.
Artículo en Inglés | MEDLINE | ID: mdl-38576332

RESUMEN

PURPOSE: Pulsatile tinnitus can be caused by a high-riding jugular bulb (HRJB), characterized by the superior position of the jugular bulb in the petrous temporal bone. The anatomical position and morphology of this entity make it challenging for endovascular treatment. We report our experience with two patients successfully treated with a stent-assisted Woven EndoBridge (WEB; Microvention, Tustin, CA, USA) device. MATERIALS AND METHODS: We describe two cases of HRJB in patients presenting with disabling pulsatile tinnitus. Temporary balloon occlusion of the jugular bulb prior to the intervention reduced tinnitus intensity. Both patients were subsequently treated under general anesthesia with the WEB device deployed in the HRJB which was held by a stent deployed in the sigmoid sinus. RESULTS: Both procedures were successful with good positioning of the WEB device and no procedural complications. Both patients had complete resolution of pulsatile tinnitus immediately after the procedure. Follow-up imaging showed successful occlusion of the venous cavity with a widely patent stent. CONCLUSION: Among patients with pulsatile tinnitus caused by an ipsilateral HRJB, a stent-assisted WEB device seems to be a viable endovascular option with angiographic and clinical success.