1.
Phys Med ; 32(1): 42-51, 2016 Jan.
Artículo en Inglés | MEDLINE | ID: mdl-26508013

RESUMEN

A comparison of calibration results and procedures in terms of air kerma length product, PKL, and air kerma, K, was conducted between eight dosimetry laboratories. A pencil-type ionization chamber (IC), generally used for computed tomography dose measurements, was calibrated according to three calibration methods, while its residual signal and other characteristics (sensitivity profile, active length) were assessed. The results showed that the "partial irradiation method" is the preferred method for the pencil-type IC calibration in terms of PKL and it could be applied by the calibration laboratories successfully. Most of the participating laboratories achieved high level of agreement (>99%) for both dosimetry quantities (PKL and K). Estimated relative standard uncertainties of comparison results vary among laboratories from 0.34% to 2.32% depending on the quantity, beam quality and calibration method applied. Detailed analysis of the assigned uncertainties have been presented and discussed.


Asunto(s)
Iones , Laboratorios/normas , Radiometría/métodos , Tomografía Computarizada por Rayos X/métodos , Aire , Calibración , Europa (Continente) , Cooperación Internacional , Modelos Estadísticos , Variaciones Dependientes del Observador , Dosis de Radiación , Reproducibilidad de los Resultados
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA