Sensors (Basel) ; 23(19)2023 Oct 09.
Article En | MEDLINE | ID: mdl-37837170

There is a worldwide need to improve blood pressure (BP) measurement error in order to correctly diagnose hypertension. Cardiovascular diseases cause 17.9 million deaths annually and are a substantial monetary strain on healthcare. The current measurement uncertainty of 3 mmHg should be improved upon. Dynamic pressure measurement standards are lacking or non-existing. In this study we propose a novel method of measuring air pressure inside the sphygmomanometer tubing during BP measurement using a condenser microphone. We designed, built, and tested a system that uses a radiofrequency (RF) modulation method to convert changes in capacitance of a condenser microphone into pressure signals. We tested the RF microphone with a low-frequency (LF) sound source, BP simulator and using a piezoresistive pressure sensor as a reference. Necessary tests were conducted to assess the uncertainty budget of the system. The RF microphone prototype has a working frequency range from 0.5 Hz to 280 Hz in the pressure range from 0 to 300 mmHg. The total expanded uncertainty (k = 2, p = 95.5%) of the RF microphone was 4.32 mmHg. The proposed method could establish traceability of BP measuring devices to acoustic standards described in IEC 61094-2 and could also be used in forming dynamic BP standards.


Blood Pressure Determination , Hypertension , Humans , Sphygmomanometers , Hypertension/diagnosis , Sound , Acoustics , Blood Pressure
...