Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 20 de 156
Filtrar
2.
Int Ophthalmol ; 44(1): 240, 2024 Jun 21.
Artículo en Inglés | MEDLINE | ID: mdl-38904711

RESUMEN

PURPOSE: To clarify the characteristics of intraocular lens (IOL) dislocation requiring IOL suture or intraocular scleral fixation. METHODS: This retrospective consecutive case series included 21 eyes (21 patients) who required sutured or sutureless intrascleral IOL fixation following IOL extraction owing to IOL dislocation at the outpatient clinic in the Department of Ophthalmology, Saitama Red Cross Hospital, Japan, between January and December 2019. Medical records were retrospectively reviewed for background diseases, location of the dislocated IOL (intracapsular/extracapsular), insertion of a capsular tension ring (CTR), and the period from IOL insertion to dislocation. RESULTS: We included 21 eyes of 21 patients who required IOL suture or intrascleral fixation for IOL dislocation at our clinic from January to December 2019 were included. The most common background disease was pseudoexfoliation syndrome (four cases), followed by atopic dermatitis, dysplasia/dehiscence of the zonule, post-retinal detachment surgery, high myopia, and uveitis (three cases each). At the time of dislocation, the IOLs were either intracapsular (16 cases, including 3 cases with CTR insertion) or extracapsular (5 cases). The time from IOL insertion to IOL dislocation was 13.7 ± 8.1 years (maximum: 31.3 years, minimum: 1.7 years). CONCLUSIONS: In this study, all 21 cases represented late IOL dislocations occurring after 3 months postoperatively. Among these late IOL dislocation cases, IOL dislocation occurred in a short-medium period of time, especially in those with CTR insertion and weakness/dehiscence of the zonule, with an average of 3 to 5 years postoperatively. We propose referring to these cases as intermediate-term IOL dislocation.


Asunto(s)
Migracion de Implante de Lente Artificial , Lentes Intraoculares , Humanos , Estudios Retrospectivos , Femenino , Masculino , Anciano , Persona de Mediana Edad , Migracion de Implante de Lente Artificial/cirugía , Migracion de Implante de Lente Artificial/etiología , Migracion de Implante de Lente Artificial/diagnóstico , Lentes Intraoculares/efectos adversos , Anciano de 80 o más Años , Factores de Tiempo , Agudeza Visual , Adulto , Esclerótica/cirugía , Técnicas de Sutura , Estudios de Seguimiento , Implantación de Lentes Intraoculares/métodos , Implantación de Lentes Intraoculares/efectos adversos , Complicaciones Posoperatorias
3.
Indian J Ophthalmol ; 72(7): 1064-1067, 2024 Jul 01.
Artículo en Inglés | MEDLINE | ID: mdl-38905465

RESUMEN

The aim of this study is to describe a modified technique for internal refixation of dislocated scleral-sutured polymethylmethacrylate (PMMA) intraocular lenses (IOLs) with eyelets. Three-port pars plana vitrectomy was performed. Through the scleral fixation site, a 30-gauge needle loaded with an 8-0 polypropylene suture was inserted into the vitreous cavity. The suture end was passed through the eyelet of IOL with 25-gauge forceps. Next, it was guided out of the eye through the original scleral fixation point. The end of the exterior suture was buried with a flapless intrascleral knotting technique. Six eyes of six patients were successfully treated with this technique and followed up for 6-12 months postsurgery. In all cases, there was significant improvement in uncorrected visual acuity. IOLs were stable with proper centration and no major complications. This modified technique offers an effective and minimally invasive surgical alternative for refixation of dislocated scleral-sutured PMMA IOLs with eyelets.


Asunto(s)
Migracion de Implante de Lente Artificial , Lentes Intraoculares , Polimetil Metacrilato , Reoperación , Esclerótica , Técnicas de Sutura , Agudeza Visual , Vitrectomía , Humanos , Masculino , Femenino , Técnicas de Sutura/instrumentación , Vitrectomía/métodos , Persona de Mediana Edad , Esclerótica/cirugía , Migracion de Implante de Lente Artificial/cirugía , Estudios de Seguimiento , Anciano , Suturas , Adulto
4.
Int Ophthalmol ; 44(1): 203, 2024 Apr 26.
Artículo en Inglés | MEDLINE | ID: mdl-38671195

RESUMEN

PURPOSE: This study aimed to observe the tilt and decentration of multifocal intraocular lens (IOL) with optic capture in Berger space within 2 years after pediatric cataract surgery. METHODS: This is a prospective observational study. The implantation of multifocal IOL (Tecnis ZMB00) with optic capture in Berger space was performed on 33 patients (48 eyes) with pediatric cataract at Qingdao Eye Hospital. Tilt and decentration of IOL was measured using Scheimpflug system (Pentacam) at 1 month and 2 years postoperatively. RESULTS: All the multifocal IOLs were successfully implanted in Berger space with optic capture and no visually significant complications were detected during the follow-up. The mean tilt of IOLs was 2.779° ± 0.950° in the vertical plane and 2.399° ± 0.898° in the horizontal plane at 1 month postoperatively, and the mean length of the decentration was 0.207 ± 0.081 mm in vertical plane and 0.211 ± 0.090 mm in the horizontal plane. Compared with 1 month after surgery, the angle of tilt decreased by a mean of 0.192° and decentration increased by a mean of 0.014 mm at the vertical meridian at 2 years postoperatively (P = 0.37 and P = 0.27, respectively), meanwhile, tilt increased by 0.265° and decentration increased by 0.012 mm at the horizontal meridian (P = 0.11 and P = 0.22, respectively). CONCLUSIONS: The follow-up results suggest the tilt and decentration of multifocal IOL implantation with optic capture in Berger space remain stable in an acceptable range within 2 years after cataract surgery in children above the age of 5. TRIAL REGISTRATION: The study was approved by the Ethics Committee of Qingdao Eye Hospital, and registered on Chinese Clinical Trial Registry (ChiCTR identifier: 1900023155).


Asunto(s)
Extracción de Catarata , Catarata , Lentes Intraoculares Multifocales , Agudeza Visual , Humanos , Masculino , Femenino , Estudios Prospectivos , Catarata/complicaciones , Catarata/fisiopatología , Preescolar , Niño , Extracción de Catarata/métodos , Extracción de Catarata/efectos adversos , Estudios de Seguimiento , Diseño de Prótesis , Migracion de Implante de Lente Artificial/diagnóstico , Migracion de Implante de Lente Artificial/fisiopatología , Migracion de Implante de Lente Artificial/etiología , Migracion de Implante de Lente Artificial/cirugía , Implantación de Lentes Intraoculares/métodos , Lactante
5.
Am J Ophthalmol ; 264: 229-234, 2024 Aug.
Artículo en Inglés | MEDLINE | ID: mdl-38461946

RESUMEN

PURPOSE: To compare 4 different secondary intraocular lens (IOL) fixation techniques regarding the least required force to dislocate a scleral fixated 3-piece IOL in human corneoscleral donor tissue. DESIGN: Experimental laboratory investigation. METHODS: The least required dislocation force (LRDF) of 4 different secondary IOL fixation techniques, namely, the techniques using transscleral tunnels (TTs; as described by Scharioth), glued haptics (GHs; Agarwal), flanged haptics (FHs; Yamane), and bent haptic ends (BH; Behera/Bolz), were investigated using 40 three-piece IOLs (Sensar AR40) fixated to human scleral tissue. The main outcome of the study, dislocation force between different techniques, was measured with a tensiometer. RESULTS: The force needed to dislocate the haptics was highest with the FH technique and was significantly higher than with all the other techniques (GH vs FH: -1.02±0.02 N, P < .001; TT vs FH: -1.08±0.21 N, P < .001; BH vs FH: -1.00±0.25 N, P = .044). There was no significant difference regarding the dislocation force between the other techniques: GH vs TT (-0.06±0.100 N, P = .988), GH vs BH (-0.02±0.03 N, P = .60), TT vs BH (-0.08±0.04 N, P > .99). CONCLUSIONS: The FH technique as described by Yamane proved to be the strongest form of secondary IOL fixation regarding dislocation force in this in vitro study. The other fixation techniques showed significantly less resistance to axial traction.


Asunto(s)
Migracion de Implante de Lente Artificial , Implantación de Lentes Intraoculares , Lentes Intraoculares , Esclerótica , Técnicas de Sutura , Humanos , Esclerótica/cirugía , Migracion de Implante de Lente Artificial/cirugía , Implantación de Lentes Intraoculares/métodos , Procedimientos Quirúrgicos sin Sutura/métodos , Donantes de Tejidos , Fenómenos Biomecánicos
6.
Am J Ophthalmol ; 262: 206-212, 2024 Jun.
Artículo en Inglés | MEDLINE | ID: mdl-38373583

RESUMEN

PURPOSE: To report and evaluate a multicenter series of 18 cases of severe, spontaneous IOL tilt involving the flanged intrascleral haptic fixation technique (FISHF). DESIGN: Clinical study with historical controls. METHODS: We report a cross-sectional study of 46 FISHF cases using the CT Lucia 602 IOL at a single academic center over a period of 24 weeks to determine the incidence of severe rotisserie-style rotational tilt. These rates were then compared with the same time-frame the prior year to help determine if this is a new phenomenon. Additional cases of severe tilt were solicited from another 4 academic centers. RESULTS: Among 46 FISHF cases at a single center, 5 developed severe tilt. No clear pattern in surgical technique, ocular history, or ocular anatomy was evident in these cases compared with controls, although the involved IOLs clustered within a narrow diopter range, indicative of a batch effect. In the same 24-week interval the year before, 33 FISHF cases were performed, none of which exhibited severe rotational tilt. In our multicenter dataset, 18 cases of tilt were identified. Surgeons included fellow and early-career physicians as well as surgeons with multiple years of experience with the Yamane technique. A variety of surgical approaches for FISHF were represented. In at least 8 of the cases, haptic rotation and/or dehiscence at the optic-haptic junction were documented. CONCLUSIONS: The identification of haptic rotation and dehiscence intraoperatively in several cases may reflect a new stability issue involving the optic-haptic junction.


Asunto(s)
Migracion de Implante de Lente Artificial , Implantación de Lentes Intraoculares , Lentes Intraoculares , Esclerótica , Humanos , Esclerótica/cirugía , Estudios Transversales , Implantación de Lentes Intraoculares/métodos , Femenino , Masculino , Anciano , Migracion de Implante de Lente Artificial/cirugía , Migracion de Implante de Lente Artificial/fisiopatología , Persona de Mediana Edad , Agudeza Visual/fisiología , Anciano de 80 o más Años , Facoemulsificación
7.
Eur J Ophthalmol ; 34(3): 884-887, 2024 May.
Artículo en Inglés | MEDLINE | ID: mdl-38115710

RESUMEN

INTRODUCTION: Single-haptic iris-claw intraocular lens (IOL) dislocation is not an uncommon complication. A few different surgical techniques are available for its refixation but usually involve a more invasive approach. We aim to demonstrate an original and simple approach for refixation of single haptic retropupillary iris-claw IOL subluxations. METHODS: We present a case of an 80-year-old male with a single haptic retropupillary iris-claw IOL subluxation in a previously vitrectomized eye. We used a new surgical approach to refix this type of subluxation. RESULTS: This new technique uses only two 30-gauge needles (one of them connected to a viscosurgical device) to re-enclavate the subluxated haptic of the retropupillary iris-claw IOL. By minimising surgical manipulation, the patient's postoperative period was uneventful. CONCLUSIONS: We highlight a new, quick, safe, and unusual surgical approach to single-haptic retropupillary iris-claw IOL dislocation in vitrectomized eyes. Because of its characteristics, we named it the "fencing" IOL repositioning technique.


Asunto(s)
Migracion de Implante de Lente Artificial , Iris , Lentes Intraoculares , Humanos , Masculino , Anciano de 80 o más Años , Iris/cirugía , Migracion de Implante de Lente Artificial/cirugía , Agudeza Visual/fisiología , Reoperación , Implantación de Lentes Intraoculares/métodos , Vitrectomía/métodos , Diseño de Prótesis
9.
Retina ; 41(10): 2035-2040, 2021 Oct 01.
Artículo en Inglés | MEDLINE | ID: mdl-34543241

RESUMEN

PURPOSE: To assess the incidence of cystoid macular edema (CME) associated with 4-point Gore-Tex suture intraocular lens (IOL) scleral fixation, before and after institution of routine intravitreal triamcinolone acetonide prophylaxis and long-term topical nonsteroidal anti-inflammatory drug usage. METHODS: Consecutive patients were included after IOL implantation with concurrent pars plana vitrectomy for spontaneous IOL dislocation due to pseudoexfoliation syndrome. We compared short-term prophylactic nonsteroidal anti-inflammatory drugs only (Group A) to prophylactic intravitreal triamcinolone acetonide and long-term nonsteroidal anti-inflammatory drugs (Group B). RESULTS: Twenty-six eyes of 26 patients with pseudoexfoliation syndrome and spontaneous IOL dislocation were studied. Mean logMAR visual acuity improved from 1.27 ± 0.80 (20/375 Snellen equivalent) preoperatively to 0.46 ± 0.39 (Snellen 20/43) postoperatively (P < 0.001). Visual outcomes were similar for Groups A and B. In Group A, 10/16 eyes had CME, 4/16 had chronic CME longer than 6 months, and 1 longer than 12 months. In Group B, 1/10 had CME (which was both chronic and refractory). CONCLUSION: In eyes with pseudoexfoliation syndrome and spontaneous IOL dislocation, 4-point Gore-Tex suture IOL ab externo fixation yielded good visual outcomes, although CME was observed more than reported elsewhere. Prophylactic intravitreal triamcinolone acetonide and long-term nonsteroidal anti-inflammatory drugs seem to reduce the risk of postoperative CME.


Asunto(s)
Implantación de Lentes Intraoculares/efectos adversos , Edema Macular/etiología , Esclerótica/cirugía , Anciano , Anciano de 80 o más Años , Migracion de Implante de Lente Artificial/etiología , Migracion de Implante de Lente Artificial/cirugía , Síndrome de Exfoliación/etiología , Femenino , Humanos , Incidencia , Edema Macular/epidemiología , Masculino , Persona de Mediana Edad , Politetrafluoroetileno , Estudios Retrospectivos , Técnicas de Sutura , Suturas , Agudeza Visual/fisiología , Vitrectomía
10.
Retina ; 41(12): 2485-2490, 2021 Dec 01.
Artículo en Inglés | MEDLINE | ID: mdl-34190728

RESUMEN

PURPOSE: To describe the clinical outcomes and surgical technique in transconjunctival sutureless intrascleral fixation of intraocular lenses, including the effectiveness of haptic flanging and peripheral iridotomy. METHOD: Retrospective series of patients who underwent sutureless intrascleral fixation of three-piece intraocular lenses by a single surgeon. RESULTS: A total of 488 eyes were included in this study. Mean follow-up was 444 days. Mean preoperative best-corrected visual acuity was 20/355, and mean postoperative best-corrected visual acuity was 20/39 (P < 0.001). Intraocular lens dislocation occurred during the postoperative period in 67 (13.7%), with the majority (65.7%) occurring within 3 months after surgery. Dislocation occurred in 13 of 196 (6.6%) flanged haptics versus 54 of 292 (18.5%) unflanged haptics (P < 0.001). Reverse pupillary block occurred in 7 of 231 eyes (3.0%) without intraoperative peripheral iridotomy but only in 1 of 257 eyes (0.4%) with iridotomy (P = 0.0297). Other complications included haptic exposure (1.2%), retinal detachment (1.0%), and endophthalmitis (0.4%). CONCLUSION: This is the largest reported series of sutureless intrascleral fixation of intraocular lenses using trocar cannulas. This technique is an effective surgical option with low complication rates. The authors recommend that haptic flanging and peripheral iridotomy be performed in all cases.


Asunto(s)
Implantación de Lentes Intraoculares/métodos , Lentes Intraoculares , Esclerótica/cirugía , Procedimientos Quirúrgicos sin Sutura/métodos , Migracion de Implante de Lente Artificial/cirugía , Humanos , Complicaciones Posoperatorias , Seudofaquia/fisiopatología , Estudios Retrospectivos , Instrumentos Quirúrgicos , Resultado del Tratamiento , Agudeza Visual/fisiología , Vitrectomía
11.
Digit J Ophthalmol ; 26(2): 7-16, 2021 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-33867876

RESUMEN

We report a case of late breakage of a 9-0 polypropylene transscleral suture used for fixation of a dislocated capsular bag-intraocular lens-modified capsular tension ring complex in a 52-year-old woman with Marfan syndrome. Breakage occurred despite use of a cow-hitch technique for external and internal fixation. We believe breakage was caused by the suture chafing on the sharp edges of the modified capsular tension ring eyelet. Cross-sectional analysis of Malyugin-modified capsular tension rings from two different manufacturers revealed a difference with respect to radius of curvature. Suturing intraocular implants with relatively sharp edges may cause suture breakage; further studies are needed to identify the critical parameters for the surface quality of sutured intraocular implants.


Asunto(s)
Migracion de Implante de Lente Artificial/etiología , Lentes Intraoculares/efectos adversos , Polipropilenos , Complicaciones Posoperatorias , Esclerótica/cirugía , Técnicas de Sutura/efectos adversos , Suturas/efectos adversos , Migracion de Implante de Lente Artificial/cirugía , Falla de Equipo , Femenino , Estudios de Seguimiento , Humanos , Persona de Mediana Edad , Reoperación/métodos , Factores de Tiempo
12.
Medicine (Baltimore) ; 100(10): e24986, 2021 Mar 12.
Artículo en Inglés | MEDLINE | ID: mdl-33725870

RESUMEN

INTRODUCTION: Implantable collamer lens have been used widely worldwide, and have been accepted by more and more doctors and patients due to good safety, stability, and effectiveness. However, there is still a problem of crystal rotation. The large angle rotation (over 10°) would weaken the original astigmatism correction effect and even induce irregular astigmatism, seriously affecting the visual quality of patients. Herein, we reported a case who had 2 times of crystal rotations after toric implantable collamer lens (TICL) implantation. PATIENT CONCERNS: The patient was a 38-year-old man who underwent TICL implantation for the correction of high myopic astigmatism in eyes. He presented a sudden decrease in the visual acuity (VA) of the left eye 4 months after the TICL implantation. The uncorrected visual acuity (UCVA) was 8/20 (refraction, +2.25 -5.25 × 68). DIAGNOSIS: Rotation of TICL was diagnosed. The toric marks with a rotation of 75° counter-clockwise from the original position were observed. INTERVENTIONS: The TICL was re-set to the original position, leading to the UCVA of 12/20 in the left eye (refraction, -0.00 -0.75 × 131), with the vaulting of 589 µm. Ten months after the TICL relocation, the patient again presented a sudden decrease in the VA of the left eye, with the UCVA of 2/20 (refraction, +2.25 -5.00 × 66). Again, the toric marks with a rotation of 75° counter-clockwise from the original position was observed, just at the same position as the last rotation. This time, the TICL was removed. The axis and power were recalculated, and a new TICL was implanted, with the rotation of 73° counter-clockwise from the horizontal line of the temporal side. OUTCOMES: The patient obtained a final UCVA of 12/20 in the left eye (refraction, +0.50 -0.50 × 26), which remained stable in the 6-month follow-up period, with the vaulting of 602 µm. LESSONS: Rotation is a common complication after TICL surgery. Relocation or replacement of TICL are safe and efficient ways to recover VA due to TICL rotation.


Asunto(s)
Migracion de Implante de Lente Artificial/diagnóstico , Astigmatismo/cirugía , Implantación de Lentes Intraoculares/efectos adversos , Miopía/cirugía , Lentes Intraoculares Fáquicas/efectos adversos , Adulto , Migracion de Implante de Lente Artificial/cirugía , Astigmatismo/complicaciones , Remoción de Dispositivos , Humanos , Implantación de Lentes Intraoculares/instrumentación , Masculino , Miopía/complicaciones , Recurrencia , Reoperación , Resultado del Tratamiento , Agudeza Visual
13.
Ocul Immunol Inflamm ; 29(7-8): 1553-1558, 2021 Nov 17.
Artículo en Inglés | MEDLINE | ID: mdl-32643989

RESUMEN

Purpose: To describe practical approaches to the management of subluxed or dislocated intraocular lenses (IOL) in patients with uveitis.Patients and methods: Retrospective case series from a specialist uveitis clinicResults: Fifteen IOLs in 13 patients were subluxed inferiorly (12) or dislocated into anterior chamber (2) or vitreous (1) at a mean delay of 12 years after cataract surgery. Six eyes required vitrectomy and seven IOL explantation. A dislocated IOL was repositioned by scleral fixation in one, and a new IOL was implanted in three (two scleral-sutured, one iris-claw). Eight were observed without surgery and 7 were left functionally aphakic (4 corrected with contact lens). The mean final best-corrected visual acuity was 0.6 LogMAR.Conclusions: There are several management choices for IOL dislocation which should take into account the degree of uveitis, patient age and expectations. We present a pragmatic approach: surgery can often be avoided in this high-risk group.


Asunto(s)
Migracion de Implante de Lente Artificial/cirugía , Complicaciones Posoperatorias/cirugía , Uveítis/complicaciones , Adulto , Anciano , Migracion de Implante de Lente Artificial/etiología , Remoción de Dispositivos , Femenino , Estudios de Seguimiento , Humanos , Implantación de Lentes Intraoculares , Masculino , Persona de Mediana Edad , Facoemulsificación , Reoperación , Estudios Retrospectivos , Factores de Tiempo , Agudeza Visual/fisiología , Adulto Joven
14.
Retin Cases Brief Rep ; 15(5): 622-624, 2021 Sep 01.
Artículo en Inglés | MEDLINE | ID: mdl-30932995

RESUMEN

PURPOSE: To describe a novel surgical technique for rescuing a dislocated "in-the-bag" accommodative intraocular lens (Crystalens; Bausch and Lomb Inc, Rochester NY) and report its postoperative outcome. METHODS: Interventional case report. An 82-year-old patient with history of pseudoexfoliation syndrome presented with a posteriorly dislocated "in-the-bag" Crystalens intraocular lens. A pars plana vitrectomy was performed. The Crystalens intraocular lens was rescued, and scleral sulcus fixated using prolene suture. RESULTS: No intraoperative or postoperative complications were noted. The preoperative visual acuity was 6/200. The vision improved to 20/30, 8 weeks after the surgery. CONCLUSION: Rescuing, instead of exchanging, a Crystalens intraocular lens might be a safe alternative in patients with a dislocated "in-the-bag" Crystalens intraocular lens.


Asunto(s)
Migracion de Implante de Lente Artificial , Implantación de Lentes Intraoculares , Procedimientos Quirúrgicos Oftalmológicos , Anciano de 80 o más Años , Migracion de Implante de Lente Artificial/cirugía , Humanos , Procedimientos Quirúrgicos Oftalmológicos/métodos , Resultado del Tratamiento
15.
Am J Ophthalmol ; 224: 150-157, 2021 04.
Artículo en Inglés | MEDLINE | ID: mdl-33309811

RESUMEN

PURPOSE: The study was performed to evaluate the refractive and visual outcome of patients with misaligned toric intraocular lenses (IOLs) after operative realignment, with and without back-calculation of the toric axis after implantation of the IOL. DESIGN: Institutional, retrospective case-control study. METHODS: This is a retrospective case series of 39 patients who underwent a second operation to realign a misaligned toric IOL from August 2013 to December 2019 at the Department of Ophthalmology, Goethe University, Frankfurt, Germany. Ideal toric axis was calculated using the back-calculator astigmatismfix.com. RESULTS: The study consists of 39 treated eyes (20 [51%] right eyes). The toric IOLs showed a postoperative misalignment of 25.69 ± 26.06°. Postrotational, uncorrected distance visual acuity (UDVA) improved from 0.39 ± 0.29 logMAR to 0.27 ± 0.18 logMAR. Refractive outcome showed a reduction of residual sphere and cylinder. The postoperative UDVA when performing alignment to the preoperative calculated axis (51%) was 0.24 ± 0.16 logMAR with a cylinder of 0.90 ± 0.90 diopter (D). In the group with alignment to a back-calculated axis (49%), the UDVA was 0.32 ± 0.20 logMAR with a cylinder of 0.76 ± 0.72 D. High cylinder power IOLs (≥2 D) showed a higher decrease in residual cylinder when back-calculation was performed than low cylinder power IOLs (<2 D) (27% vs 9%). The mean spherical equivalent prediction error of the back-calculator was 0.54 ± 0.55 D. CONCLUSION: Realignment of misaligned toric IOLs improves visual acuity and reduces residual refractive errors. Especially for high cylinder power IOLs, better refractive outcome can be seen when performing a back-calculation before realignment.


Asunto(s)
Migracion de Implante de Lente Artificial/fisiopatología , Astigmatismo/fisiopatología , Implantación de Lentes Intraoculares , Seudofaquia/fisiopatología , Refracción Ocular/fisiología , Agudeza Visual/fisiología , Anciano , Migracion de Implante de Lente Artificial/cirugía , Astigmatismo/cirugía , Biometría , Estudios de Casos y Controles , Femenino , Humanos , Masculino , Persona de Mediana Edad , Facoemulsificación , Reoperación , Estudios Retrospectivos , Resultado del Tratamiento
16.
J Fr Ophtalmol ; 43(10): 1062-1068, 2020 Dec.
Artículo en Inglés | MEDLINE | ID: mdl-32811657

RESUMEN

PURPOSE: To evaluate the predisposing factors, management and visual prognosis of intraocular Lens (IOL) dislocation into the posterior segment. METHODS: The cases of posterior IOL dislocation from January 2012 to May 2017 at 2 centers were reviewed. Only eyes with dislocations requiring IOL explantation or repositioning were included. Predisposing factors, interval between cataract surgery and IOL dislocation, circumstances of onset, management, and postoperative complications are reported. RESULTS: 72 eyes of 72 patients were included. The mean age was 67.6 years. 47 patients (68%) were men. The mean time interval from cataract surgery to IOL dislocation was significantly shorter in the out-of-the bag group than the in-the-bag IOL dislocation group (3.8 months vs 132 months, P=0.002). Predisposing factors for out-of-the-bag IOL dislocation were mainly capsular rupture and/or zonular dehiscence (83%) after complicated cataract surgery. The predisposing factors for in-the-bag IOL dislocation were high myopia (40%), pseudoexfoliation syndrome (40%), previous vitrectomy (38%), or Marfan syndrome (3%) with uneventful cataract surgery. The type of luxated implant was mainly a 3-piece foldable IOL (50%), followed by foldable one-piece IOL (28%) and a rigid one-piece IOL (17%). Most cases of posterior chamber IOL dislocation occurred spontaneously (80%) without a trigger event. Management consisted of a posterior approach in 24 cases (33%) or an anterior approach in 48 cases (67%), associated with IOL repositioning in 20 eyes (28%), and IOL replacement in 34 eyes (47%). Finally, 18 eyes (25%) were left aphakic. Postoperative complications occurred in 7 cases (9.7%). CONCLUSIONS: Predisposing factors and time from cataract surgery to IOL dislocation were different for out-of-the bag versus in-the-bag IOL dislocation. Management of IOL dislocation varied considerably, depending on surgeon preference and experience. Surgery for IOL dislocation significantly improved best corrected visual acuity and was associated with a low complication rate.


Asunto(s)
Migracion de Implante de Lente Artificial , Remoción de Dispositivos , Falla de Prótesis/etiología , Adulto , Anciano , Anciano de 80 o más Años , Migracion de Implante de Lente Artificial/diagnóstico , Migracion de Implante de Lente Artificial/epidemiología , Migracion de Implante de Lente Artificial/etiología , Migracion de Implante de Lente Artificial/cirugía , Remoción de Dispositivos/métodos , Remoción de Dispositivos/estadística & datos numéricos , Síndrome de Exfoliación/complicaciones , Síndrome de Exfoliación/diagnóstico , Síndrome de Exfoliación/epidemiología , Síndrome de Exfoliación/cirugía , Femenino , Humanos , Cápsula del Cristalino/patología , Cápsula del Cristalino/cirugía , Implantación de Lentes Intraoculares/efectos adversos , Implantación de Lentes Intraoculares/estadística & datos numéricos , Subluxación del Cristalino/diagnóstico , Subluxación del Cristalino/epidemiología , Subluxación del Cristalino/etiología , Subluxación del Cristalino/cirugía , Lentes Intraoculares/efectos adversos , Masculino , Persona de Mediana Edad , Miopía/complicaciones , Miopía/diagnóstico , Miopía/epidemiología , Miopía/cirugía , Complicaciones Posoperatorias/diagnóstico , Complicaciones Posoperatorias/epidemiología , Complicaciones Posoperatorias/etiología , Complicaciones Posoperatorias/cirugía , Pronóstico , Reoperación/métodos , Reoperación/estadística & datos numéricos , Estudios Retrospectivos , Factores de Riesgo , Resultado del Tratamiento , Vitrectomía/efectos adversos , Vitrectomía/métodos , Vitrectomía/estadística & datos numéricos
17.
Am J Ophthalmol ; 218: 28-39, 2020 10.
Artículo en Inglés | MEDLINE | ID: mdl-32445701

RESUMEN

PURPOSE: To compare outcomes of patient who underwent surgery using perfluorooctane (PFO; C8F18; Ala Octa) with those who underwent surgery with perfluorodecalin (F-Decalin). DESIGN: Retrospective, consecutive, comparative, interventional case series. METHODS: A total of 48 eyes that underwent vitrectomy with PFO were compared to 29 eyes that underwent vitrectomy with perfluorodecalin. Two experienced surgeons performed vitrectomies at the Geneva University Eye Clinic. Visual acuity before, at 8 and 24 weeks after surgery, was documented, and spectral domain optical coherence tomography (SD-OCT) images were analyzed for abnormalities. RESULTS: Two patients experienced severe retinal toxicity, including 1 with severe vision loss. However, no statistical differences in VA were observed between the PFO and perfluorodecalin patients. Analysis of SD-OCT images showed differences in occurrence of several abnormalities, for example, inner segment-outer segment alterations were found in 60.4% of eyes treated with PFO and in 10.3% of perfluorodecalin-treated eyes; retinal atrophic areas were found in 41.7% of PFO and in none of the perfluorodecalin eyes; inner limiting membrane contraction was found in 58.4% of PFO and in none of perfluorodecalin eyes; inner retina cystic alterations were found in 58.3% of PFO eyes and 17.2% of perfluorodecalin eyes; outer retina cystic alterations were found in 39.6% of PFO eyes and 13.8% of perfluorodecalin eyes; retinal holes were found in 14.6% of PFO eyes and in none of the perfluorodecalin eyes; and outer retinal inclusions were found in 20.8% of PFO eyes and in 3.45% of perfluorodecalin eyes. CONCLUSIONS: Perfluorooctane caused significantly more toxic damage than perfluorodecalin. Special consideration should be given to develop a central European Union (EU) control agency for medical devices and to reevaluate safety procedures currently accepted by the EU and International Organization for Standardization for intraocular surgery.


Asunto(s)
Migracion de Implante de Lente Artificial/cirugía , Cuerpos Extraños en el Ojo/cirugía , Fluorocarburos/toxicidad , Retina/efectos de los fármacos , Desprendimiento de Retina/cirugía , Enfermedades de la Retina/inducido químicamente , Adulto , Anciano , Anciano de 80 o más Años , Endotaponamiento , Femenino , Humanos , Masculino , Persona de Mediana Edad , Retina/diagnóstico por imagen , Enfermedades de la Retina/diagnóstico por imagen , Enfermedades de la Retina/fisiopatología , Estudios Retrospectivos , Suiza , Tomografía de Coherencia Óptica , Agudeza Visual/fisiología , Vitrectomía
18.
Am J Ophthalmol ; 218: 314-319, 2020 10.
Artículo en Inglés | MEDLINE | ID: mdl-32222364

RESUMEN

PURPOSE: To evaluate the outcome and viability of a modified self-sealing scleral pocket technique for scleral fixation of an intraocular lens (IOL). DESIGN: Retrospective interventional case series. METHODS: A retrospective, interventional case series in which 81 eyes underwent this modified technique, performed in the previous 4 years. Cases included a dislocated IOL and an absent posterior capsule or subluxated cataract. Two diagonally opposed paralimbal, curved self-sealing scleral pockets were made 3 mm away from the limbus along with a vitrectomy. A multipiece IOL was used, and the haptics were fixed under the scleral pockets inside a linear scleral tunnel underneath the superficial scleral flap. The conjunctiva was opposed by using cautery. Postoperatively, anterior segment optical coherence tomography (AS-OCT) was performed in order to detect intrascleral hyper-reflective cross-section of the haptics and optic tilt. The optic position was re-evaluated using ultrasound biomicroscopy (UBM). Main outcomes were mean pre- and postoperative best corrected visual acuity (BCVA), postoperative astigmatism. Postoperative AS-OCT was performed to detect the position of the haptics and optics. UBM was done to recheck any optic tilt. RESULTS: Mean pre- and postoperative BCVA were LogMAR 1.2 ± 0.59 and LogMAR 0.47 ± 0.3 (P < 0.001). Mean postoperative astigmatism was 1.27 ± 1.02 Dcyl. In AS-OCT, an intrascleral hyper-reflective shadow signified stable haptics without any optic tilt. UBM also showed a stable position of the IOL optic without any tilt. CONCLUSIONS: This modified, simple way of scleral fixation of an IOL decreases the duration of surgery with minimal complication.


Asunto(s)
Implantación de Lentes Intraoculares/métodos , Esclerótica/cirugía , Adolescente , Adulto , Anciano , Afaquia Poscatarata/fisiopatología , Afaquia Poscatarata/cirugía , Migracion de Implante de Lente Artificial/fisiopatología , Migracion de Implante de Lente Artificial/cirugía , Niño , Preescolar , Femenino , Humanos , Masculino , Persona de Mediana Edad , Estudios Retrospectivos , Colgajos Quirúrgicos , Procedimientos Quirúrgicos sin Sutura , Adhesivos Tisulares , Agudeza Visual/fisiología , Vitrectomía
19.
Eur J Ophthalmol ; 30(4): 680-684, 2020 Jul.
Artículo en Inglés | MEDLINE | ID: mdl-30841757

RESUMEN

PURPOSE: To compare the incidence of re-orientating surgery to improve misalignment of three models of acrylic toric intraocular lenses: AcrySof toric intraocular lens (Alcon Laboratories, Inc.), TECNIS toric intraocular lens (Johnson & Johnson Vision, Inc.) and HOYA 355 toric intraocular lens (HOYA). METHODS: In this retrospective, multicenter case series, medical charts were reviewed for collecting data on realignment surgery of toric intraocular lenses at 10 ophthalmic surgical sites in Japan. RESULTS: Over all, intraocular lens repositioning surgery was conducted in 89 of 9430 eyes (0.944%) at an average of 10.5 ± 9.7 days after the initial cataract surgery. The incidence was 0.213% (11/5155), 1.797% (62/3451) and 1.942% (16/824) with AcrySof, TECNIS and HOYA toric intraocular lenses, respectively. The incidence was significantly lower with AcrySof than with other two brands of toric intraocular lenses (p < 0.0001). In those eyes which underwent reorientation surgery, the amount of misalignment was 26.4 ± 21.9°, 29.7 ± 15.4° and 28.1 ± 20.7° with AcrySof, TECNIS and HOYA toric intraocular lenses, respectively; there was no significant difference among groups (p = 0.821). The repositioning surgery significantly reduced misalignment in all three groups. CONCLUSION: The rotational stability was considerably different among toric intraocular lenses of different manufacturers. The incidence of repositioning surgery was significantly lower with AcrySof than with TECNIS and HOYA toric intraocular lenses.


Asunto(s)
Migracion de Implante de Lente Artificial/epidemiología , Migracion de Implante de Lente Artificial/cirugía , Implantación de Lentes Intraoculares , Lentes Intraoculares , Complicaciones Posoperatorias , Adolescente , Adulto , Anciano , Anciano de 80 o más Años , Astigmatismo/fisiopatología , Longitud Axial del Ojo , Femenino , Humanos , Incidencia , Masculino , Persona de Mediana Edad , Facoemulsificación , Seudofaquia/fisiopatología , Reoperación , Estudios Retrospectivos , Agudeza Visual/fisiología , Adulto Joven
20.
Eur J Ophthalmol ; 30(5): NP79-NP81, 2020 Sep.
Artículo en Inglés | MEDLINE | ID: mdl-30897947

RESUMEN

PURPOSE: Fracture of a three-piece polyimide-elastimide intraocular lens at the optic-haptic junction following blunt trauma is a possible complication after cataract surgery. We report a case of an isolated posterior chamber intraocular lens fracture caused by direct ocular blunt trauma. CASE: A 51-year-old patient underwent blunt orbital trauma caused by a raw chicken egg that was thrown at him, as he was walking. The posterior chamber intraocular lens was found in the anterior chamber, with a broken optic-haptic junction. The intraocular lens was exchanged, followed by iris fixation and pupilloplasty, with satisfactory postoperative anatomical and optical outcomes. DISCUSSION: The three-piece polyimide-elastimide intraocular lens was fractured at the optic-haptic junction. Although the fractured intraocular lens surface had a regular appearance, it has been previously reported that the polyimide haptic's durability is probably inferior to that of polymethylmethacrylate. Therefore, it is highly susceptible to shear stress induced by a blunt trauma. CONCLUSION: To our knowledge, this is an uncommon report of an implanted posterior chamber polyimide-elastimide intraocular lens fracture following blunt orbital trauma. Intraocular lens exchange and fixation had successful optical and anatomical results.


Asunto(s)
Migracion de Implante de Lente Artificial/etiología , Lesiones Oculares/etiología , Lentes Intraoculares , Falla de Prótesis/etiología , Heridas no Penetrantes/etiología , Cámara Anterior/cirugía , Migracion de Implante de Lente Artificial/diagnóstico , Migracion de Implante de Lente Artificial/cirugía , Humanos , Implantación de Lentes Intraoculares , Edema Macular/diagnóstico por imagen , Edema Macular/etiología , Edema Macular/cirugía , Masculino , Persona de Mediana Edad , Facoemulsificación , Reoperación , Resinas Sintéticas , Tomografía de Coherencia Óptica
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA