Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 20 de 228
Filtrar
2.
Respir Care ; 69(7): 790-798, 2024 Jun 28.
Artículo en Inglés | MEDLINE | ID: mdl-38744474

RESUMEN

BACKGROUND: During invasive ventilation, external flow jet nebulization results in increases in displayed exhaled tidal volumes (VT). We hypothesized that the magnitude of the increase is inaccurate. An ASL 5000 simulator measured ventilatory parameters over a wide range of adult settings: actual VT, peak inspiratory pressure (PIP), and time to minimum pressure. METHODS: Ventilators with internal and external flow sensors were tested by using a variety of volume and pressure control modes (the target VT was 420 mL). Patient conditions (normal, COPD, ARDS) defined on the ASL 5000 were assessed at baseline and with 3.5 or 8 L/min of added external flow. Patient-triggering was assessed by reducing muscle effort to the level that resulted in backup ventilation and by changing ventilator sensitivity to the point of auto-triggering. RESULTS: Results are reported as percentage change from baseline after addition of 3.5 or 8 L/min external flow. For ventilators with internal flow sensors, changes in displayed exhaled VT ranged from 10% to 118%, however, when using volume control, actual increases in actual VT and PIP were only 4%-21% (P = .063, .031) and 6%-24% (P = .25, .031), respectively. Changes in actual VT correlated closely with changes in PIP (P < .001; R2 = 0.68). For pressure control, actual VT decreased by 3%-5% (P = .031) and 4%-9% (P = .031) with 3.5 and 8 L/min respectively, PIP was unchanged. With external flow sensors at the distal Y-piece junction, volume and pressure changes were statistically insignificant. The time to minimum pressure increased at most by 8% (P = .02) across all modes and ventilators. The effects on muscle pressure were minimal (∼1 cm H2O), and ventilator sensitivity effects were nearly undetectable. CONCLUSIONS: External flow jet nebulization resulted in much smaller changes in volume than indicated by the ventilator display. Statistically significant effects were confined primarily to machines with internal flow sensors. Differences approached the manufacturer-reported variation in ventilator baseline performance. During nebulizer therapy, effects on VT can be estimated at the bedside by monitoring PIP.


Asunto(s)
Nebulizadores y Vaporizadores , Volumen de Ventilación Pulmonar , Ventiladores Mecánicos , Humanos , Diseño de Equipo , Enfermedad Pulmonar Obstructiva Crónica/fisiopatología , Enfermedad Pulmonar Obstructiva Crónica/terapia , Síndrome de Dificultad Respiratoria/terapia , Síndrome de Dificultad Respiratoria/fisiopatología , Presiones Respiratorias Máximas , Ventilación con Chorro de Alta Frecuencia/instrumentación , Ventilación con Chorro de Alta Frecuencia/métodos , Adulto
3.
BMC Anesthesiol ; 21(1): 65, 2021 03 02.
Artículo en Inglés | MEDLINE | ID: mdl-33653271

RESUMEN

BACKGROUND: High frequency jet ventilation (HFJV) is an open ventilating technique to maintain ventilation for emergency or difficult airway. However, whether jet ventilation or conventional oxygen therapy (COT) is more effective and safe in maintaining adequate oxygenation, is unclear among patients with airway stenosis during bronchoscopic intervention (BI) under deep sedation. METHODS: A prospective randomized cohort study was conducted to compare COT (high flow oxygen) with normal frequency jet ventilation (NFJV) and HFJV in oxygen supplementation during BI under deep sedation from March 2020 to August 2020. Patients receiving BI under deep sedation were randomly divided into 3 parallel groups of 50 patients each: the COT group (fractional inspired oxygen (FiO2) of 1.0, 12 L/min), the NFJV group (FiO2 of 1.0, driving pressure of 0.1 MPa, and respiratory rate (RR) 15 bpm) and the HFJV Group (FiO2 of 1.0, driving pressure of 0.1 MPa, and RR of 1200 bpm). Pulse oxygen saturation (SpO2), mean arterial blood pressure and heart rate were recorded during the whole procedure. Arterial blood gas was examined and recorded 15 min after the procedure was initiated. The procedure duration, dose of anesthetics, and adverse events during BI in the three groups were also recorded. RESULTS: A total of 161 patients were enrolled, with 11 patients excluded. The clinical characteristics were similar among the three groups. PaO2 of the COT and NFJV groups was significantly lower than that of the HFJV group (P < 0.001). PaO2 was significantly correlated with ventilation mode (P < 0.001), body mass index (BMI) (P = 0.019) and procedure duration (P = 0.001). Multiple linear regression showed that only BMI and procedure duration were independent influencing factors of arterial blood gas PaO2 (P = 0.040 and P = 0.002, respectively). The location of airway lesions and the severity of airway stenosis were not statistically correlated with PaCO2 and PaO2. CONCLUSIONS: HFJV could effectively and safely improve intra-operative PaO2 among patients with airway stenosis during BI in deep sedation, and it did not increase the intra-operative PaCO2 and the risk of hypercapnia. PaO2 was correlated with ventilation mode, BMI and procedure duration. Only BMI and procedure duration were independent influencing factors of arterial blood gas PaO2. PaCO2 was not correlated with any preoperative factor. TRIAL REGISTRATION: Chinese Clinical Trial Registry. Registration number, ChiCTR2000031110 , registered on March 22, 2020.


Asunto(s)
Broncoscopía/métodos , Sedación Profunda/métodos , Ventilación con Chorro de Alta Frecuencia/instrumentación , Ventilación con Chorro de Alta Frecuencia/métodos , Terapia por Inhalación de Oxígeno/métodos , Saturación de Oxígeno/fisiología , Estudios de Cohortes , Femenino , Humanos , Masculino , Máscaras , Persona de Mediana Edad , Estudios Prospectivos
4.
Laryngoscope ; 131(10): 2292-2297, 2021 10.
Artículo en Inglés | MEDLINE | ID: mdl-33609043

RESUMEN

OBJECTIVES/HYPOTHESIS: To evaluate the safety and complications of endoscopic airway surgery using supraglottic jet ventilation with a team-based approach. STUDY DESIGN: Retrospective cohort study. METHODS: Subjects at two academic institutions diagnosed with laryngotracheal stenosis who underwent endoscopic airway surgery with jet ventilation between January 2008 and December 2018 were identified. Patient characteristics (age, gender, race, follow-up duration) and comorbidities were extracted from the electronic health record. Records were reviewed for treatment approach, intraoperative data, and complications (intraoperative, acute postoperative, and delayed postoperative). RESULTS: Eight hundred and ninety-four patient encounters from 371 patients were identified. Intraoperative complications (unplanned tracheotomy, profound or severe hypoxic events, barotrauma, laryngospasm) occurred in fewer than 1% of patient encounters. Acute postoperative complications (postoperative recovery unit [PACU] rapid response, PACU intubation, return to the emergency department [ED] within 24 hours of surgery) were rare, occurring in fewer than 3% of patient encounters. Delayed postoperative complications (return to the ED or admission for respiratory complaints within 30 days of surgery) occurred in fewer than 1% of patient encounters. Diabetes mellitus, active smoking, and history of previous tracheotomy were independently associated with intraoperative, acute, and delayed complications. CONCLUSIONS: Employing a team-based approach, jet ventilation during endoscopic airway surgery demonstrates a low rate of complications and provides for safe and successful surgery. LEVEL OF EVIDENCE: 4 Laryngoscope, 131:2292-2297, 2021.


Asunto(s)
Ventilación con Chorro de Alta Frecuencia/efectos adversos , Complicaciones Intraoperatorias/epidemiología , Laparoscopía/efectos adversos , Laringoestenosis/cirugía , Complicaciones Posoperatorias/epidemiología , Estenosis Traqueal/cirugía , Adulto , Comorbilidad , Diabetes Mellitus/epidemiología , Femenino , Estudios de Seguimiento , Ventilación con Chorro de Alta Frecuencia/instrumentación , Humanos , Complicaciones Intraoperatorias/etiología , Laparoscopía/instrumentación , Laringoestenosis/epidemiología , Masculino , Persona de Mediana Edad , Grupo de Atención al Paciente , Complicaciones Posoperatorias/etiología , Estudios Retrospectivos , Factores de Riesgo , Fumar/epidemiología , Estenosis Traqueal/epidemiología , Resultado del Tratamiento
5.
Minerva Pediatr ; 72(1): 60-64, 2020 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-29479941

RESUMEN

Infants born prematurely require external respiratory support device like ventilation for the purpose of life saving. However, these ventilation machines have complications that sometimes unfortunately result in morbidity. New ventilation techniques have been developed to prevent morbidity, but have yet to be fully evaluated. The present review article would discuss current aspects of this life saving gear especially for pediatric patients in clinical setting. Besides basic ventilation apparatus, advancements in the filed like proportional assist ventilation, volume targeted ventilation would be discussed.


Asunto(s)
Recién Nacido de Bajo Peso , Respiración Artificial/efectos adversos , Ventiladores Mecánicos/efectos adversos , Displasia Broncopulmonar/etiología , Diafragma , Ventilación con Chorro de Alta Frecuencia/instrumentación , Ventilación de Alta Frecuencia/efectos adversos , Ventilación de Alta Frecuencia/instrumentación , Ventilación de Alta Frecuencia/métodos , Humanos , Recién Nacido , Soporte Ventilatorio Interactivo/métodos , Neumotórax/etiología , Enfisema Pulmonar/etiología , Trastornos Respiratorios/etiología , Respiración Artificial/instrumentación , Respiración Artificial/métodos , Respiración Artificial/tendencias
6.
Paediatr Anaesth ; 29(8): 799-807, 2019 08.
Artículo en Inglés | MEDLINE | ID: mdl-31233654

RESUMEN

BACKGROUND: Self-made Y-connector jet-oxygenation devices with wide-bore expiratory port have been described but not evaluated in infant models. Little is known about the effect of oxygen flow rates on jet oxygenation via transtracheal cannula. AIMS: The aim of this study was to compare two self-made Y-connector jet-oxygenation devices against the ENK oxygen flow modulator™, and the effects of three different oxygen flow rates based on body weight, in both unobstructed and obstructed airways, on the time to re-oxygenate in a rabbit infant model. The aim was also to assess the effectiveness of an oxygen flow rate of 1 L/min, for re-oxygenation using ENK oxygen flow modulator™. METHODS: Nine rabbits were grouped in threes: Group 1 had a Y-connector attached to an intravenous infusion tubing, Group 2 the same Y-connector attached to a perfusion oxygenator tubing and Group 3, ENK oxygen flow modulator™. From oxygen saturations of 75%, the rabbits were jet oxygenated using their assigned device for 10 minutes at each flow rate of 1 L/kg/min, 1.5 L/kg/min and 2 L/kg/min with their airways unobstructed and later, obstructed. Group 3 had additional experiments involving an absolute oxygen flow rate of 1 L/min. RESULTS: All devices resulted in rapid re-oxygenation within 40 seconds at flow rates of 1 L/kg/min. Oxygen flow rates beyond 1 L/kg/min in obstructed airways resulted in high airway pressures. All rabbits in Group 3 with obstructed airways died from barotrauma when jet oxygenated at a flow rate of 1.5 L/kg/min. When an oxygen flow rate of 1 L/min was used in Group 3, there was a failure to re-oxygenate to SpO2 90% within 120 seconds in some rabbits. CONCLUSION: Our animal model results suggest that self-made Y-connector jet-oxygenation devices with wide-bore expiratory port are efficacious and perhaps safer than ENK oxygen flow modulator™ in obstructed airways, and jet oxygenation with minimal oxygen flow rates starting at 1 L/kg/min or (age [years] + 4) L/min, whichever lower, should be considered.


Asunto(s)
Ventilación con Chorro de Alta Frecuencia/instrumentación , Intubación Intratraqueal/métodos , Respiración Artificial/métodos , Obstrucción de las Vías Aéreas , Animales , Modelos Animales de Enfermedad , Diseño de Equipo , Conejos
7.
Br J Anaesth ; 119(1): 158-166, 2017 Jul 01.
Artículo en Inglés | MEDLINE | ID: mdl-28974061

RESUMEN

BACKGROUND: Hypoventilation is the main reason for hypoxia during upper gastrointestinal endoscopy procedures with sedation. The key to preventing hypoxia is to maintain normal ventilation during the procedure. We introduced supraglottic jet oxygenation and ventilation (SJOV) through a new Wei nasal jet tube (WNJ) to reduce the incidence of hypoxia in patients sedated with propofol during upper gastrointestinal endoscopy procedures. METHODS: In a multicentre, prospective randomized single-blinded study, 1781 outpatients undergoing routine upper gastrointestinal endoscopy who were sedated with propofol by an anaesthetist were randomized into the following three groups: the supplementary oxygen via nasal cannula group [nasal cannula oxygen: O 2 (2 litres min -1 ) was administered via a nasal cannula]; the supplementary oxygen via WNJ group [WNJ oxygen: O 2 (2 litres min -1 ) was administered through a WNJ]; and the SJOV via WNJ group (WNJ SJOV: SJOV was administered via WNJ) at three centres from March 2015 to July 2016. The primary outcome of interest was the incidence of hypoxia (peripheral oxygen saturation of 75-89%). Other adverse events were also recorded. RESULTS: Supraglottic jet oxygenation and ventilation decreased the incidence of hypoxia from 9 to 3% ( P <0.0001). No severe hypoxia occurred in the WNJ SJOV group, one instance occurred in the WNJ oxygen group, and two instances were observed in the nasal cannula oxygen supply control group. Supraglottic jet oxygenation and ventilation-related minor adverse events increased significantly within 1 min after the procedure but decreased 30 min later. CONCLUSIONS: The use of SJOV during upper gastrointestinal endoscopy for patients who are sedated with propofol reduces the incidence of hypoxia, with minor and tolerable adverse events. Supraglottic jet oxygenation and ventilation has a favourable risk-to-benefit ratio and may improve patient safety. CLINICAL TRIAL REGISTRATION: NCT02436018.


Asunto(s)
Endoscopía Gastrointestinal , Ventilación con Chorro de Alta Frecuencia/métodos , Hipnóticos y Sedantes/farmacología , Oxígeno/metabolismo , Propofol/farmacología , Adolescente , Adulto , Anciano , Anciano de 80 o más Años , Femenino , Ventilación con Chorro de Alta Frecuencia/instrumentación , Humanos , Hipoxia/prevención & control , Masculino , Persona de Mediana Edad , Estudios Prospectivos , Método Simple Ciego , Adulto Joven
8.
Biomed Res Int ; 2016: 4234861, 2016.
Artículo en Inglés | MEDLINE | ID: mdl-27847813

RESUMEN

The indications for rigid bronchoscopy for interventional pulmonology have increased and include stent placements and transbronchial cryobiopsy procedures. The shared airway between anesthesiologist and pulmonologist and the open airway system, requiring specific ventilation techniques such as jet ventilation, need a good understanding of the procedure to reduce potentially harmful complications. Appropriate adjustment of the ventilator settings including pause pressure and peak inspiratory pressure reduces the risk of barotrauma. High frequency jet ventilation allows adequate oxygenation and carbon dioxide removal even in cases of tracheal stenosis up to frequencies of around 150 min-1; however, in an in vivo animal model, high frequency jet ventilation along with normal frequency jet ventilation (superimposed high frequency jet ventilation) has been shown to improve oxygenation by increasing lung volume and carbon dioxide removal by increasing tidal volume across a large spectrum of frequencies without increasing barotrauma. General anesthesia with a continuous, intravenous, short-acting agent is safe and effective during rigid bronchoscopy procedures.


Asunto(s)
Anestesia General , Broncoscopía/métodos , Ventilación con Chorro de Alta Frecuencia/métodos , Pulmón/fisiopatología , Adulto , Barotrauma/fisiopatología , Broncoscopía/efectos adversos , Broncoscopía/instrumentación , Dióxido de Carbono/química , Ventilación con Chorro de Alta Frecuencia/efectos adversos , Ventilación con Chorro de Alta Frecuencia/instrumentación , Humanos , Mediciones del Volumen Pulmonar
9.
Br J Anaesth ; 117 Suppl 1: i28-i38, 2016 Sep.
Artículo en Inglés | MEDLINE | ID: mdl-27566790

RESUMEN

BACKGROUND: Transtracheal jet ventilation (TTJV) is recommended in several airway guidelines as a potentially life-saving procedure during the 'Can't Intubate Can't Oxygenate' (CICO) emergency. Some studies have questioned its effectiveness. METHODS: Our goal was to determine the complication rates of TTJV in the CICO emergency compared with the emergency setting where CICO is not described (non-CICO emergency) or elective surgical setting. Several databases of published and unpublished literature were searched systematically for studies describing TTJV in human subjects. Complications were categorized as device failure, barotrauma (including subcutaneous emphysema), and miscellaneous. Device failure was defined by the inability to place and/or use the TTJV device, not patient survival. RESULTS: Forty-four studies (428 procedures) met the inclusion criteria. Four studies included both emergency and elective procedures. Thirty studies described 132 emergency TTJV procedures; 90 were CICO emergencies. Eighteen studies described 296 elective TTJV procedures. Device failure occurred in 42% of CICO emergency vs 0% of non-CICO emergency (P<0.001) and 0.3% of elective procedures (P<0.001). Barotrauma occurred in 32% of CICO emergency vs 7% of non-CICO emergency (P<0.001) and 8% of elective procedures (P<0.001). The total number of procedures with any complication was 51% of CICO emergency vs 7% of non-CICO emergency (P<0.001) and 8% of elective procedures (P<0.001). Several reports described TTJV-related subcutaneous emphysema hampering subsequent attempts at surgical airway or tracheal intubation. CONCLUSIONS: TTJV is associated with a high risk of device failure and barotrauma in the CICO emergency. Guidelines and recommendations supporting the use of TTJV in CICO should be reconsidered.


Asunto(s)
Manejo de la Vía Aérea/métodos , Obstrucción de las Vías Aéreas/terapia , Ventilación con Chorro de Alta Frecuencia/métodos , Manejo de la Vía Aérea/efectos adversos , Barotrauma/etiología , Urgencias Médicas , Falla de Equipo , Ventilación con Chorro de Alta Frecuencia/efectos adversos , Ventilación con Chorro de Alta Frecuencia/instrumentación , Humanos , Intubación Intratraqueal/métodos
10.
J Clin Anesth ; 33: 233-5, 2016 Sep.
Artículo en Inglés | MEDLINE | ID: mdl-27555171

RESUMEN

The safety of high pressure source ventilation (jet ventilation) is dependent upon upper airway patency to facilitate adequate passive expiration and prevent increasing intrathoracic pressure and its associated deleterious sequelae. Distortions in airway anatomy may make passive expiration inadequate or impossible in some patients. We report the elective use of the Ventrain device to provide ventilation in a clinical setting of upper airway obstruction in a patient with post radiation fibrosis that had previously prevented passive expiration during attempted high pressure source ventilation.


Asunto(s)
Obstrucción de las Vías Aéreas/terapia , Ventilación con Chorro de Alta Frecuencia/instrumentación , Ventilación con Chorro de Alta Frecuencia/métodos , Obstrucción de las Vías Aéreas/etiología , Carcinoma de Células Escamosas/complicaciones , Carcinoma de Células Escamosas/radioterapia , Neoplasias de Cabeza y Cuello/complicaciones , Neoplasias de Cabeza y Cuello/radioterapia , Humanos , Masculino , Persona de Mediana Edad , Radioterapia/efectos adversos , Tiroidectomía
11.
Zhongguo Yi Xue Ke Xue Yuan Xue Bao ; 38(2): 131-5, 2016 Apr.
Artículo en Chino | MEDLINE | ID: mdl-27181886

RESUMEN

OBJECTIVE: To compare the effectiveness of high-frequency jet ventilation via Wei jet nasal airway and controlled ventilation with improved laryngeal mask airway during bronchial thermoplasty. METHODS: Twenty-eight patients undergoing bronchial thermoplasty were equally divided into two groups: group A (high-frequency jet ventilation through Wei jet nasal airway) and group B (controlled ventilation with improved laryngeal mask airway). Pulse oxygenation,heart rate,and mean arterial blood pressure were recorded after entering the operating room (T0), 1 minute after administration/induction (T1), bronchoscope inserting (T2), 15 minutes (T3)/30 minutes (T4)/45 minutes (T5) after ventilation,at the end of the operation (T6), and at the recovery of patients' consciousness (T7). The pH,arterial oxygen partial pressure,and arterial carbon dioxide partial pressure were recorded at T0, T4, and T6. The endoscope indwelling duration,operative time,patients' awakening time,adverse events during anesthesia,satisfactions of patients and operators, anesthesic effectiveness were also recorded. RESULTS: The arterial carbon dioxide partial pressur in group A at T4 and T6 were significantly higher than in group B (P<0.05). The pH in group A at T4 and T6 was significantly lower than in group B (P<0.05). The endoscope indwelling duration and the operative time in group B were significantly shorter than in group A (P<0.05) while the recovery of consciousness in group B was significantly longer than in group A (P<0.05). The satisfaction for operators and the efficacy of anesthesia in group B were better than in group A (P<0.05). The number of adverse events in group B was significantly smaller than in group A (P<0.05). CONCLUSION: The improved laryngeal mask airway with controlled ventilation is more suitable for bronchial thermoplasty.


Asunto(s)
Broncoscopía , Ablación por Catéter , Ventilación con Chorro de Alta Frecuencia/instrumentación , Máscaras Laríngeas , Análisis de los Gases de la Sangre , Frecuencia Cardíaca , Humanos
13.
J Clin Anesth ; 26(7): 570-3, 2014 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-25439421

RESUMEN

A novel method in the management of refractory severe hypoxemia during one-lung ventilation (OLV) in a patient who presented with myasthenia gravis, asthma, a symptomatic mediastinal mass, hiatal hernia, and a moderate pericardial effusion is presented. The patient was scheduled for excision of a large anterior mediastinal mass and creation of a pericardial window through a left thoracotomy. One-lung ventilation was achieved using an Arndt bronchial blocker. High-frequency jet ventilation (HFJV) was applied to the surgical nondependent lung through the lumen of the Arndt endobronchial blocker with titration of positive end-expiratory pressure to the dependent lung. Oxygenation improved significantly. The use of HFJV through the Arndt blocker offers an effective method for treatment of refractory hypoxemia during OLV.


Asunto(s)
Asma/complicaciones , Ventilación con Chorro de Alta Frecuencia/instrumentación , Hipoxia/terapia , Miastenia Gravis/complicaciones , Ventilación Unipulmonar/instrumentación , Anciano , Femenino , Ventilación con Chorro de Alta Frecuencia/métodos , Humanos , Hipoxia/etiología , Ventilación Unipulmonar/métodos
15.
Br J Anaesth ; 113(6): 1039-45, 2014 Dec.
Artículo en Inglés | MEDLINE | ID: mdl-24980421

RESUMEN

BACKGROUND: Temporizing oxygenation by percutaneous transtracheal ventilation (PTV) is a recommended emergency technique in 'can't intubate, can't oxygenate' (CICO) situations. Barotrauma risk increases if expiration is obstructed. The Ventrain(®) is a new PTV device that assists expiration. Our aim was to compare key physiological outcomes after PTV with the Ventrain and the Manujet(®) in a large animal obstructed airway model. METHODS: Five anaesthetized sheep had post-apnoea PTV performed for 15 min using the Ventrain or Manujet with the proximal airway completely or critically obstructed, yielding four ventilation protocols per sheep. After apnoeic desaturation ([Formula: see text]70%), a 4 s rescue breath was delivered. Subsequent 2 s breaths were delivered whenever the airway pressure fell <10 cm H2O. RESULTS: Both devices achieved rapid re-oxygenation. There were marked device differences (Ventrain vs Manujet) in peak airway pressures with rescue (16 vs 40 cm H2O) breaths, minute ventilation (4.7 vs 0.1 litre min(-1)), and end-protocol pH (7.34 vs 7.01). There was no clinical evidence of barotrauma in any sheep after any ventilation protocol. An equilibration phase prevented large subatmospheric intrathoracic pressure development with Ventrain ventilation. CONCLUSIONS: The Ventrain provided stable oxygenation and effective ventilation at low airway pressures during emergency PTV in critically obstructed airways. The Manujet provided effective temporizing oxygenation in this situation with hypoventilation necessary to minimize barotrauma risk. The nature and extent of airway obstruction may not be known in a CICO emergency but an understanding of device differences may help inform optimal ventilation device and method selection.


Asunto(s)
Obstrucción de las Vías Aéreas/terapia , Apnea/terapia , Ventilación con Chorro de Alta Frecuencia/instrumentación , Obstrucción de las Vías Aéreas/sangre , Obstrucción de las Vías Aéreas/fisiopatología , Resistencia de las Vías Respiratorias , Animales , Apnea/sangre , Apnea/fisiopatología , Barotrauma/etiología , Barotrauma/prevención & control , Modelos Animales de Enfermedad , Urgencias Médicas , Diseño de Equipo , Femenino , Ventilación con Chorro de Alta Frecuencia/efectos adversos , Ventilación con Chorro de Alta Frecuencia/métodos , Intubación Intratraqueal/instrumentación , Intubación Intratraqueal/métodos , Oxígeno/sangre , Terapia por Inhalación de Oxígeno/instrumentación , Terapia por Inhalación de Oxígeno/métodos , Presión Parcial , Oveja Doméstica
16.
Laryngorhinootologie ; 93(7): 455-60, 2014 Jul.
Artículo en Alemán | MEDLINE | ID: mdl-24676871

RESUMEN

BACKGROUND: High-frequency jet ventilation (HFJV) through thin catheters creates more room for surgical procedures in laryngotracheal surgery, while few contraindications exist. In contrast, the anaesthesiologist has to cope with reduced monitoring and numerous contraindications. Therefore, every participating discipline has to thoroughly know the method. We report our experiences with HFJV and focus on relevant points, which contribute to a safe and cooperative procedure. MATERIAL AND METHODS: We performed a retrospective chart review for patients, who were operated for benign laryngotracheal lesions using HFJV at the Hannover Medical School, Department of Otorhino-Laryngology between June 2012 and February 2013. We analyze patient characteristics, complications and important anaesthesiologic and operative steps. RESULTS: A total of 46 cases of microlaryngoscopies and tracheoscopies were included in this study. The median body mass index was 27 kg/m(2). According to the ASA-classification, 24% of patients were categorized class I, 54% class II and 22% class III. In 2 cases we had reversible, peripheral desaturations down to 70%. In one case, emergent re-intubation with an endotracheal tube was inevitable due to rapid desaturation below 50%. Further complications were not observed. CONCLUSION: HFJV represents a safe ventilation approach in laryngotracheal surgery, when patients are thoroughly selected. Especially the surgeon benefits from more room in the operating field. Obesity, reflux and cardiopulmonary diseases have to be evaluated well, as these can represent relative contraindications. Individual decisions can be made under consideration of all comorbidities.


Asunto(s)
Anestesia General , Endoscopía , Ventilación con Chorro de Alta Frecuencia , Enfermedades Otorrinolaringológicas/cirugía , Adolescente , Adulto , Anciano , Anciano de 80 o más Años , Niño , Preescolar , Contraindicaciones , Endoscopía/instrumentación , Diseño de Equipo , Ventilación con Chorro de Alta Frecuencia/instrumentación , Humanos , Lactante , Laringoscopía/instrumentación , Masculino , Microcirugia/instrumentación , Persona de Mediana Edad , Procedimientos Quirúrgicos Otorrinolaringológicos/instrumentación , Seguridad del Paciente , Selección de Paciente , Estudios Retrospectivos , Tráquea/cirugía , Adulto Joven
17.
Paediatr Anaesth ; 24(2): 208-13, 2014 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-24387148

RESUMEN

INTRODUCTION: Limited information is available on the management of the 'cannot intubate, cannot ventilate' (CICV) situation in infants. We compared the time to achieve adequate oxygenation following rescue ventilation using the Enk oxygen flow modulator (OFM) with a jet ventilator in a simulated CICV situation using the rabbit as an infant respiratory model. METHODS: Following institutional ethics committee approval, needle cricothyrotomy was performed under direct vision in nine anesthetized rabbits following surgical exposure of the larynx. After ensuring adequate level of anesthesia and analgesia, and confirming proper positioning of the 18G cannula, apnea was induced by the administration of myorelaxant and the SpO2 was allowed to drop to 75% before initiating rescue ventilation via either the OFM or jet ventilator. RESULTS: Five rabbits were ventilated with the OFM and four with the jet ventilator. Ventilation was maintained with either device for 15 min. All rabbits were successfully rescued using either device. There was no statistical difference in the time required for SpO2 to return to 80%, 85%, 90%, and 95%. CONCLUSIONS: Both devices facilitated successful rescue ventilation through a needle cricothyrotomy.


Asunto(s)
Cartílago Cricoides/cirugía , Ventilación con Chorro de Alta Frecuencia/métodos , Oxígeno/sangre , Respiración Artificial/métodos , Tiroidectomía/métodos , Ventiladores Mecánicos , Animales , Apnea/terapia , Análisis de los Gases de la Sangre , Presión Sanguínea/fisiología , Dióxido de Carbono/sangre , Estudios de Factibilidad , Frecuencia Cardíaca/fisiología , Ventilación con Chorro de Alta Frecuencia/instrumentación , Concentración de Iones de Hidrógeno , Hipoxia/terapia , Conejos , Resultado del Tratamiento
18.
Cardiovasc Intervent Radiol ; 37(1): 140-6, 2014 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-23636246

RESUMEN

PURPOSE: To report feasibility and potential benefits of high-frequency jet ventilation (HFJV) in tumor ablations techniques in liver, kidney, and lung lesions. METHODS: This prospective study included 51 patients (14 women, mean age 66 years) bearing 66 tumors (56 hepatic, 5 pulmonary, 5 renal tumors) with a median size of 16 ± 8.7 mm, referred for tumor ablation in an intention-to-treat fashion before preoperative anesthesiology visit. Cancellation and complications of HFJV were prospectively recorded. Anesthesia and procedure duration, as well as mean CO2 capnea, were recorded. When computed tomography guidance was used, 3D spacial coordinates of an anatomical target <2 mm in diameter on 8 slabs of 4 slices of 3.75-mm slice thickness were registered. RESULTS: HFJV was used in 41 of 51 patients. Of the ten patients who were not candidate for HFJV, two patients had contraindication to HFJV (severe COPD), three had lesions invisible under HFJV requiring deep inspiration apnea for tumor targeting, and five patients could not have HFJV because of unavailability of a trained anesthetic team. No specific complication or hypercapnia related to HFJV were observed despite a mean anesthetic duration of 2 h and ventilation performed in procubitus (n = 4) or lateral decubitus (n = 6). Measured internal target movement was 0.3 mm in x- and y-axis and below the slice thickness of 3.75 mm in the z-axis in 11 patients. CONCLUSIONS: HFJV is feasible in 80 % of patients allowing for near immobility of internal organs during liver, kidney, and lung tumor ablation.


Asunto(s)
Ablación por Catéter/instrumentación , Ventilación con Chorro de Alta Frecuencia/instrumentación , Neoplasias Renales/cirugía , Neoplasias Hepáticas/cirugía , Neoplasias Pulmonares/cirugía , Adulto , Anciano , Anciano de 80 o más Años , Medios de Contraste/administración & dosificación , Diseño de Equipo , Estudios de Factibilidad , Femenino , Humanos , Neoplasias Renales/diagnóstico por imagen , Neoplasias Hepáticas/diagnóstico por imagen , Neoplasias Pulmonares/diagnóstico por imagen , Masculino , Persona de Mediana Edad , Estudios Prospectivos , Radiografía Intervencional , Tomografía Computarizada por Rayos X
19.
Thorac Surg Clin ; 24(1): 13-25, 2014 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-24295656

RESUMEN

Tracheobronchial surgery constitutes a challenge to the anesthetist because it involves the anatomic structures dedicated to bulk gas transport. Common approaches to airway management and gas exchange for extrathoracic and intrathoracic airway surgery are reviewed, with due regard to less common methods thought crucial for specific procedures. Tracheal surgery, beyond sharing the airways, requires sharing with the surgeon ideas on preoperative assessment, on the impact on gas exchange of induction across compromised airways, and of emergence from anesthesia with airways altered by surgical repair. Mutual understanding is essential to prevent, rapidly identify, and correct imminent loss of airway viability.


Asunto(s)
Anestesia Epidural/métodos , Anestesia General/métodos , Intercambio Gaseoso Pulmonar , Procedimientos Quirúrgicos Torácicos , Tráquea/cirugía , Anestesia General/instrumentación , Bronquios/cirugía , Ventilación con Chorro de Alta Frecuencia/instrumentación , Humanos , Intercambio Gaseoso Pulmonar/fisiología
20.
Respiration ; 85(1): 59-63, 2013.
Artículo en Inglés | MEDLINE | ID: mdl-23107666

RESUMEN

BACKGROUND: To determine the suitability of different superimposed high-frequency jet ventilation (SHFJV) application methods during tracheal bleeding. OBJECTIVE: To determine the effect of SHFJV on the aspiration of blood during tracheal bleeding. METHODS: A test lung was ventilated using SHFJV via a rigid endoscope, a jet laryngoscope and a 4-lumen jet catheter. Packed red blood cells (PRBCs) were injected into the artificial trachea caudally to the rigid endoscope and jet laryngoscope ventilation, and both caudally and cranially during ventilation via the 4-lumen jet catheter, and the migration of PRBCs during ventilation was studied using continuous video recording. RESULTS: Migration of blood into the lower respiratory tract did not occur during SHFJV via the rigid endoscope and jet laryngoscope and via the 4-lumen jet catheter with the bleeding caudal to ventilation source. If the bleeding was cranial to the 4-lumen jet catheter ventilation, migration of blood into the lower respiratory tract was seen when reflux of blood reached the entrainment area. From this area, blood is transported within the jet stream into the lower respiratory tract. CONCLUSIONS: SHFJV protects the lower respiratory tract from blood aspiration in case of tracheal bleeding. During SHFJV via the 4-lumen jet catheter, aspiration of blood only occurs if bleeding is localized cranial to the 4-lumen jet catheter ventilation. In case of heavy tracheal bleeding, the jet sources should be positioned cranial to the site of bleeding.


Asunto(s)
Asfixia/prevención & control , Hemorragia/complicaciones , Ventilación con Chorro de Alta Frecuencia/instrumentación , Modelos Biológicos , Enfermedades de la Tráquea/complicaciones , Asfixia/etiología , Diseño de Equipo , Análisis de Falla de Equipo , Humanos , Laringoscopios
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...