ABSTRACT
Daily volume loss of the stump leads to a poor fit of the prosthetic socket. A method of preventing this volume loss and maintaining a good fit was developed. A vacuum (-78 kPa) was drawn on the expulsion port of a total surface-bearing suction socket to hold the liner tightly against the socket. Stump volume of 10 trans-tibial amputees was measured prior to and immediately after a 30 minute walk with normal and vacuum socket conditions. Under the normal condition, the limb lost an average of 6.5% of its volume during the walk. In contrast, with the liner held tightly by vacuum, the limb gained an average of 3.7% in volume. It is believed that the difference observed between conditions resulted from a greater negative pressure developed during the swing phase of gait with the vacuum condition. X-rays revealed that the limb and tibia pistoned 4 mm and 7 mm less, respectively, under the vacuum condition. The combination of reduced pistoning and maintenance of volume is thought to account for the more symmetrical gait observed with the vacuum.