Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters











Database
Language
Publication year range
1.
Rev Sci Instrum ; 95(8)2024 Aug 01.
Article in English | MEDLINE | ID: mdl-39101791

ABSTRACT

The ITER Radial Gamma-Ray Spectrometer (RGRS) consists of three gamma-ray detectors observing the plasma through three collimated, coplanar, radial lines of sight (LoS). The system was initially designed to monitor the runaway electron emission and the alpha-particle density profile [Nocente et al., Nucl. Fusion 57, 076016 (2017)]. This work presents a novel technique for measuring the fusion power during D-T operation using the RGRS. This method is based on the absolute measurement of the 17 MeV fusion gamma-rays and a semi-analytical computation of their transport from the plasma source to the detectors. This approach was initially developed and tested at JET during the second D-T campaign (DTE2) on a single LoS diagnostic [Dal Molin et al., Phys. Rev. Lett. (submitted) (2024); Rebai et al., Phys. Rev. C (submitted) (2024); and Marcer et al., Nucl. Fusion (unpublished) (2024)]. This work exploits the multiple LoS of the RGRS to create a combined virtual diagnostic whose detected fraction of the total plasma emission is less affected by variations in the plasma emission profile, reducing systematic uncertainties on the estimated total emission, compared to the individual detectors.