Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
J Bone Joint Surg Am ; 72(9): 1307-15, 1990 Oct.
Article in English | MEDLINE | ID: mdl-2229105

ABSTRACT

Thirteen subjects who had normal knees and fifteen patients who had a chronic rupture of the anterior cruciate ligament were tested in order to compare the measurements of tibiofemoral displacement as recorded by four commercial devices: the Acufex knee-signature system, the Genucom knee-analysis system, the Medmetric KT-1000 arthrometer, and the Stryker knee-laxity tester. Anterior and posterior displacement were measured at forces of eighty-nine newtons (twenty pounds) and 133 newtons (thirty pounds). We found significant differences in reproducibility of measurement among the devices. The Acufex, Medmetric, and Stryker devices had more reproducible measurements, and they could be used to identify 80 to 90 per cent of the normal subjects and anterior cruciate-deficient patients. The Genucom device had poorer reproducibility of measurement, and it tended to register greater differences in displacement between the right and left knees of normal subjects.


Subject(s)
Anterior Cruciate Ligament Injuries , Joint Instability/pathology , Knee Joint/pathology , Orthopedic Equipment , Adult , Compliance , Female , Humans , Joint Instability/physiopathology , Knee Joint/anatomy & histology , Knee Joint/physiopathology , Male , Prospective Studies , Reference Values , Reproducibility of Results , Rupture/pathology , Rupture/physiopathology