Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters











Database
Language
Publication year range
1.
J Pharm Biomed Anal ; 41(4): 1386-90, 2006 Jun 16.
Article in English | MEDLINE | ID: mdl-16597494

ABSTRACT

This study developed a simple method for the simultaneous determination of zonisamide (ZNS), carbamazepine (CBZ) and its active metabolite, carbamazepine-10,11-epoxide (CBZE) in infant serum using reversed-phase high-performance liquid chromatograph (HPLC). The method involves a single-step protein precipitation procedure that uses no solid-phase or liquid-liquid extraction. The HPLC separation was carried out on a Cadenza CD-C18 column (3 microm, 4.6 mm x 150 mm) with potassium phosphate buffer (pH 4.6; 25 mM)-methanol-acetonitrile (65:20:15 (v/v/v)) as a mobile phase at a 1.0 ml/min flow rate: ZNS was detectable using a UV detector at 235 nm, and both CBZ and CBZE were at 215 nm. The quantification limits were established in accordance with each therapeutic range at 2.5 microg/ml for ZNS, 0.5 microg/ml for CBZ, and 0.25 microg/ml for CBZE. The respective coefficients of variation were 1.3-6.0% and 2.2-7.7% for the intra- and inter-assay.


Subject(s)
Anticonvulsants/blood , Carbamazepine/analogs & derivatives , Carbamazepine/blood , Chromatography, High Pressure Liquid/methods , Isoxazoles/blood , Anticonvulsants/therapeutic use , Carbamazepine/therapeutic use , Epilepsy/blood , Epilepsy/drug therapy , Humans , Infant , Isoxazoles/therapeutic use , Zonisamide
SELECTION OF CITATIONS
SEARCH DETAIL