Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 5 de 5
Filter
Add more filters










Database
Language
Publication year range
1.
Artif Organs ; 24(8): 648-52, 2000 Aug.
Article in English | MEDLINE | ID: mdl-10971255

ABSTRACT

The authors studied avoidance of coagulation occurrence using computational fluid dynamics (CFD) analysis from the fluid dynamical point of view. Concerning centrifugal pumps, blood coagulation sometimes occurs at the region behind the impeller where the flow is generally stagnant. Therefore, we conducted a thorough study with the specimen pump with and without washout holes, mocking up the Nikkiso HPM-15. As the result, the model with washout holes indicated that the fluid rotates rapidly at the vicinity of the shaft and generates washout effects near the stationary rear casing. On the other hand, the model without washout holes showed that fluid cannot be quickly shipped out of the area behind the impeller and rotates mildly around the shaft. To clarify the moving relations between the impeller and the fluid, validation studies by comparing the results of CFD analysis and flow visualization experiments are ongoing; thus far, the studies show that CFD results are similar to the results from flow visualization experiments.


Subject(s)
Heart-Assist Devices , Centrifugation , Computer Simulation , Equipment Design , Humans , Rheology , Thrombosis/etiology , Thrombosis/prevention & control
2.
Artif Organs ; 23(8): 757-61, 1999 Aug.
Article in English | MEDLINE | ID: mdl-10463503

ABSTRACT

The purpose of this study was to examine the relationship between local flow conditions and the hemolysis level by integrating hemolysis tests, flow visualization, and computational fluid dynamics to establish practical design criteria for centrifugal blood pumps with lower levels of hemolysis. The Nikkiso centrifugal blood pump was used as a standard model, and pumps with different values of 3 geometrical parameters were tested. The studied parameters were the radial gap between the outer edge of the impeller vane and the casing wall, the position of the outlet port, and the discharge angle of the impeller vane. The effect of a narrow radial gap on hemolysis was consistent with no evidence that the outlet port position or the vane discharge angle affected blood trauma in so far as the Nikkiso centrifugal blood pump was concerned. The radial gap should be considered as a design parameter of a centrifugal blood pump to reduce blood trauma.


Subject(s)
Extracorporeal Circulation/instrumentation , Heart-Assist Devices , Hemolysis , Animals , Computer Simulation , Equipment Design , Goats , In Vitro Techniques , Rheology
3.
Artif Organs ; 23(8): 762-8, 1999 Aug.
Article in English | MEDLINE | ID: mdl-10463504

ABSTRACT

To establish an efficient design process for centrifugal blood pumps, the results of computational fluid dynamics (CFD) analysis were compared to the results of flow visualization tests and hemolysis tests, using the Nikkiso centrifugal blood pump. CFD analysis revealed that the radial gap greatly affected the shear stress in the outlet diffuser. The hemolysis study also indicated a similar tendency. To see the flow behind the impeller, we conducted a comparative study between models with and without washout holes using the CFD technique. CFD analysis indicated that flow and pressure distributions behind the impeller were different between both models, and a particle was observed to remain longer behind the impeller in the model without washout holes. In the future, CFD analysis could be a useful tool for developing blood pumps in comparison to flow visualization tests and hemolysis tests.


Subject(s)
Extracorporeal Circulation/instrumentation , Heart-Assist Devices , Computer Simulation , Equipment Design , Hemolysis , Rheology
4.
Artif Organs ; 22(6): 461-5, 1998 Jun.
Article in English | MEDLINE | ID: mdl-9650666

ABSTRACT

Regarding the development of a centrifugal blood pump to be connected directly with small diameter tubings for pediatric use while minimizing hemolysis, we have studied the inlet port side configurations of a pump using both a hemolysis test and computational fluid dynamics (CFD) analysis. We have conducted a hemolysis test on 2 models. The tapered shape inlet has proven to be lower in the index of hemolysis (IH) than the straight shape. CFD analyses utilizing expanded flow paths indicated that the flow velocity decreased as the fluid path became larger within the tapered nozzle. When entering the pump chamber, the flow rushed in at a greater velocity through the straight nozzle due to its small diameter. The straight shape showed an abrupt change in pressure around the entrance of the pump chamber while the tapered shape did not. The flow inlet angle of the straight model was observed to be larger than that of the tapered model because of its smaller turning radius.


Subject(s)
Extracorporeal Circulation/instrumentation , Blood Flow Velocity , Blood Pressure , Blood Viscosity , Blood Volume , Child , Computer Simulation , Equipment Design , Hemolysis , Hemorheology , Humans , Intubation/instrumentation , Rotation , Surface Properties
5.
Artif Organs ; 22(5): 381-5, 1998 May.
Article in English | MEDLINE | ID: mdl-9609345

ABSTRACT

To establish quantitative, efficient design theories for centrifugal blood pumps, computational fluid dynamics (CFD) analyses were compared to the results of flow visualization tests and hemolysis tests, mainly on the Nikkiso centrifugal blood pump. The results turned out to coincide in the velocity vector plots. CFD analysis revealed that the smaller the gap is, the greater the shear stress becomes. This tendency becomes even greater with a radial gap change. Hemolysis study also indicated that the smaller the gap is, the greater the hemolysis. CFD analysis in comparison with hemolysis tests could be a useful index for developing blood pumps in the future.


Subject(s)
Computer Simulation , Heart-Assist Devices , Hemolysis , Biomechanical Phenomena , Blood Flow Velocity , Centrifugation , Equipment Design , Models, Biological , Viscosity