Anal Chim Acta ; 1054: 95-103, 2019 Apr 25.
Article in English | MEDLINE | ID: mdl-30712597

ABSTRACT

The precise quantification of mercury (Hg) stable isotope compositions in low concentration or dilute samples poses analytical challenges due to Hg mass limitations. Common Hg pre-concentration procedures require extended processing times, making rapid Hg stable isotope measurements challenging. Here we present a modified pre-concentration method that combines commonly used Hg reduction and gold trap amalgamation followed by semi-rapid thermal desorption (less than 1 h) and chemical trapping. This custom designed system was demonstrated to perform adequately on multiple trapping matrices including a new bromine monochloride (BrCl) wet oxidant trap (40% 3HNO3:BrCl), capable of trapping consistently in 2 mL volume over a wide range of Hg masses (5-200 ng). The procedure was also shown to work effectively on natural matrices, waters and sediments, producing comparable isotope results to the direct digestion analyses. Here, we present a method that can effectively triple sample throughput in comparison to traditional procedures, and also access lower concentration matrices without compromising the accuracy or precision of Hg isotope measurements.