Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Eur Surg Res ; 37(4): 204-9, 2005.
Article in English | MEDLINE | ID: mdl-16260869

ABSTRACT

OBJECTIVE: To prevent clamp injury that may occur during aortic surgery, we aimed to develop a special balloon occlusion (BO) device to lower the thromboembolic risk in patients with severe atherosclerosis during aortic aneurysm repair. METHODS: The study comprised two test phases: a laboratory-testing series focussing on flexible artificial aortas, and an experimental series conducted on 10 pigs. RESULTS: The device proved to be effective during the laboratory tests and the experiments on pigs. No complications such as intraoperative balloon rupture, dislocation, or occlusion leaks occurred. No damage to the aortic vessels was observed in further histological examinations. CONCLUSIONS: This BO device has the potential to become an alternative to cross-clamping for vascular surgeons in patients with severely atherosclerotic vessels.


Subject(s)
Aorta/surgery , Balloon Occlusion , Blood Vessel Prosthesis Implantation , Blood Vessel Prosthesis , Vascular Surgical Procedures , Animals , Aorta/pathology , Atherosclerosis/therapy , Balloon Occlusion/instrumentation , Balloon Occlusion/methods , Humans , Swine , Vascular Surgical Procedures/adverse effects , Vascular Surgical Procedures/instrumentation , Vascular Surgical Procedures/methods