1.
Appl Radiat Isot ; 107: 191-194, 2016 Jan.
Article in English | MEDLINE | ID: mdl-26519921

ABSTRACT

This work introduces a new, versatile and robust X-ray detector with hemispherical 2π geometry, based on a set of 15 small cylindrical proportional counters located in a hexagonal and pentagonal fullerene C60 pattern, at the same distance from the center (where a sample is placed). The counteranode consists of stainless steel sewing needles with spherical tips measuring approximately 80 µm in diameter. The space between the counters and the sample could contain air, the same gas as the counters or vacuum. This allows a significant increase in the count rates by a factor approximately equal to the number of counters connected. It is shown that an energy resolution of 20% for 5.9 keV photons can be obtained, and a global counting rate of around 10(6)counts/s is achievable by the 15 Needle Anode Proportional Counters (NAPCs) operating in parallel mode, in our setup.


Subject(s)
Radiometry/instrumentation , Electrodes , Equipment Design , Humans , Photons , Radiometry/statistics & numerical data , X-Rays