Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters











Database
Language
Publication year range
1.
IEEE Trans Biomed Eng ; 48(4): 495-9, 2001 Apr.
Article in English | MEDLINE | ID: mdl-11322538

ABSTRACT

Designs for a low-cost dual wavelength transducer based on light transmittance and a disposable cuvette for monitoring oxygen saturation (SO2) in extracorporeal arterial and venous blood are presented. The transducer utilizes red and infrared light-emitting diodes and a large photodiode; it is designed to attach to a flow-through cuvette modified from 3/8-in x3/8-in bypass-circuit connectors. A mock extracorporeal circulation system was assembled to evaluate operation of the transducer at a controlled blood SO2 and the relationship between light transmittance and hemodilution. SO2 was calculated based on multiple linear regression analyses. The results show a high correlation between the SO2 obtained with the equipment designed and values measured with commercial gasometric equipment in the range of 50% to 100% (r2 = 0.976, error <2%). The method presented allows continuous and real time measurement of whole blood SO2 with a low-cost transmittance transducer.


Subject(s)
Extracorporeal Circulation/instrumentation , Oxygen/blood , Transducers , Animals , Cattle , Cost Control , Equipment Design , Hemoglobins , Regression Analysis , Transducers/economics
SELECTION OF CITATIONS
SEARCH DETAIL