Eye Contact Lens ; 48(11): 485-488, 2022 Nov 01.
Article in English | MEDLINE | ID: mdl-35916554

ABSTRACT

ABSTRACT: This case reports on the use of wavefront-guided (wfg) optics on custom ocular impression-based scleral lenses (IBSLs) for visual improvement in a patient with keratoconus (KC). A 28-year-old man with KC, who had previously failed a traditional, diagnostically fit scleral lens (tSL), was fit with IBSLs with traditional optics. Using a system that included a dot matrix on the IBSL and a wavefront aberrometer with pupil and dot registration software, a wfgIBSL was created. When compared with the IBSL, the wfgIBSL reduced the total higher-order root mean square (HORMS) 67% and 64% in the right and left eye, respectively, resulting in a 2-line improvement in best-contact lens visual acuity (BCLVA) for both eyes. This case demonstrates the successful creation and application of a wfgIBSL resulting in a stable lens, a reduction in HORMS, and an improvement in BCLVA, after failure with a diagnostically fit tSL.


Subject(s)
Contact Lenses , Keratoconus , Lens, Crystalline , Male , Humans , Adult , Keratoconus/complications , Keratoconus/therapy , Sclera , Visual Acuity