Ther Apher Dial ; 28(1): 89-95, 2024 Feb.
Article in English | MEDLINE | ID: mdl-37583361

ABSTRACT

INTRODUCTION: PD catheter tip migration is a common complication and a significant cause of catheter malfunction. In this perspective, we present our experience with a new catheter and a new technique that involves the use of a new triple cuff PD catheter and a low entry site in an attempt to prevent PD catheter migration. METHODS: A total of 503 incident PD patients have been studied in more than one PD center over a period of 5 years. RESULTS: During the 5-year follow up we recorded zero percent catheter migration. Other technical complications were poor drainage in 3.4%, omental wrap in 2.8%, early leakage in 3.4%, and catheter replacement in 2.4%. By the end of the study, the one-year PD catheter survival was 97.6%. CONCLUSION: Our new triple cuff PD catheter and our low-entry approach seem to be effective in preventing PD catheter migration and minimizing other mechanical complications.


Subject(s)
Catheters, Indwelling , Peritoneal Dialysis , Humans , Prospective Studies , Catheters, Indwelling/adverse effects , Nephrologists , Catheterization/adverse effects , Catheterization/methods