Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 5 de 5
Filter
Add more filters










Language
Publication year range
1.
Braz Dent J ; 28(4): 461-466, 2017.
Article in English | MEDLINE | ID: mdl-29160398

ABSTRACT

This study aimed at evaluating the changes in surface characteristics and NiTi content of reciprocating and rotary instruments after continuous use. Thirty brand-new instruments were assigned to 3 groups (n=10): PNX1 Group - ProTaper Next system - instrument X1; PNX2 Group - ProTaper Next system - instrument X2 and WO Group - WaveOne system - Primary instrument 25.08. The instruments were used to prepare 60 simulated root canals. Analysis of surface characteristics by scanning electron microscopy (SEM) and chemical analysis of Ni and Ti content by energy-dispersive spectroscopy (EDS) were performed before and after the first and third use of the instruments. Only WO Group had significant increase in the quantity of defects and deformations after the third use (p<0.05). PNX1 Group had significant decrease in Ni content after the third use, in comparison with the unused instruments (p<0.05). PNX1 Group had no decrease in Ti content throughout the time of use (p>0.05), however, in PNX2 Group, there was significant decrease in the different time interval of analysis (p<0.05). Continuous use promoted increase in defects and deformations only for WaveOne instruments. Chemical composition presented significant changes according to the time of instruments use.


Subject(s)
Dental Instruments , Root Canal Preparation/instrumentation , Humans , Microscopy, Electron, Scanning , Spectrometry, X-Ray Emission , Surface Properties
2.
Braz. dent. j ; 28(4): 461-466, July-Aug. 2017. tab, graf
Article in English | LILACS | ID: biblio-888671

ABSTRACT

Abstract This study aimed at evaluating the changes in surface characteristics and NiTi content of reciprocating and rotary instruments after continuous use. Thirty brand-new instruments were assigned to 3 groups (n=10): PNX1 Group - ProTaper Next system - instrument X1; PNX2 Group - ProTaper Next system - instrument X2 and WO Group - WaveOne system - Primary instrument 25.08. The instruments were used to prepare 60 simulated root canals. Analysis of surface characteristics by scanning electron microscopy (SEM) and chemical analysis of Ni and Ti content by energy-dispersive spectroscopy (EDS) were performed before and after the first and third use of the instruments. Only WO Group had significant increase in the quantity of defects and deformations after the third use (p<0.05). PNX1 Group had significant decrease in Ni content after the third use, in comparison with the unused instruments (p<0.05). PNX1 Group had no decrease in Ti content throughout the time of use (p>0.05), however, in PNX2 Group, there was significant decrease in the different time interval of analysis (p<0.05). Continuous use promoted increase in defects and deformations only for WaveOne instruments. Chemical composition presented significant changes according to the time of instruments use.


Resumo Este estudo teve como objetivo avaliar as alterações nas características de superfície e teor de NiTi de instrumentos reciprocantes e rotatórios após uso contínuo. Trinta instrumentos novos foram separados em três grupos (n=10): Grupo PNX1 - ProTaper Next system - instrumento X1; Grupo PNX2 - ProTaper Next system - instrumento X2 e Grupo WO - WaveOne system - instrumento Primary 25.08. Os instrumentos foram utilizados para o preparo de 60 canais radiculares simulados. Análise por microscopia eletronica de varredura (MEV), a análise química do conteúdo de Ni e Ti por espectroscopia de energia dispersiva (EED) foram realizadas antes e após o primeiro e terceiro uso dos instrumentos. Apenas o Grupo WO apresentou aumento significativo na quantidade de defeitos e deformações após o terceiro uso (p<0,05). O grupo PNX1 teve diminuição significativa no teor de Ni após o terceiro uso, em comparação aos instrumentos não utilizados (p<0,05). O grupo PNX1 não apresentou diminuição do teor de Ti ao longo do tempo de uso (p>0,05), no entanto, no Grupo PNX2, houve diminuição significativa nos diferentes intervalos de tempo de análise (p<0,05). O uso contínuo promoveu aumento nos defeitos e deformações somente para os instrumentos WaveOne. A composição química apresentou alterações significativas de acordo com o tempo de uso dos instrumentos.


Subject(s)
Humans , Dental Instruments , Root Canal Preparation/instrumentation , Microscopy, Electron, Scanning , Spectrometry, X-Ray Emission , Surface Properties
4.
Clin Oral Investig ; 20(8): 1987-1993, 2016 Nov.
Article in English | MEDLINE | ID: mdl-26685850

ABSTRACT

OBJECTIVES: The aim of this study was to assess postoperative pain in a prospective randomized clinical trial comparing two groups, using the Reciproc® system in one group and the ProTaper® rotary system in the other. MATERIAL AND METHODS: The study included 78 male patients, aged 18-64 years (mean age of 26 years), with asymptomatic pulp necrosis in mandibular molar teeth (n = 78). The single-session endodontic treatment was performed by a single operator specialized in Endodontics. Mechanical preparation of the root canals was performed using the ProTaper® and Reciproc® instrumentation techniques. Postoperative pain was recorded using a verbal rating scale (VRS) and verbal description with well-defined categories at the three following time intervals: 24 h, 72 h, and 7 days after the endodontic procedure. The assessment of postoperative pain was recorded as no pain, mild pain, moderate pain, and severe pain or flare-up. Data were analyzed using the nonparametric Mann-Whitney test with the aid of the STATA® software. RESULTS: The incidence of postoperative pain in the ProTaper group (PT) 24 h after the endodontic procedure was 17.9 and 5.1 % after 72 h. In the Reciproc group (RP), the incidence after 24 h was 15.3 and 2.5 % after 72 h. No patients presented severe pain at the time intervals assessed. CONCLUSIONS: No significant difference (p > 0.05) in postoperative pain was found between the ProTaper® and Reciproc® instrumentation technique during endodontic treatment in this study. CLINICAL RELEVANCE: According to our findings and the results of the clinical trial, the occurrence of postoperative pain was low and similar between the reciprocating and rotary techniques during the time intervals assessed. These results are different from basic laboratory studies that affirm that the reciprocating techniques tend to promote more postoperative pain since extrusion of debris is greater.


Subject(s)
Pain, Postoperative/epidemiology , Root Canal Preparation/instrumentation , Adolescent , Adult , Brazil/epidemiology , Humans , Male , Middle Aged , Molar/surgery , Nickel , Pain Measurement , Prospective Studies , Titanium
5.
Restor Dent Endod ; 38(3): 167-71, 2013 Aug.
Article in English | MEDLINE | ID: mdl-24010084

ABSTRACT

Biomechanical preparation of root canals with accentuated curvature is challenging. New rotatory systems, such as Reciproc, require a shorter period of time to prepare curved canals, and became a viable alternative for endodontic treatment of teeth with root dilaceration. Thus, this study aimed to report a clinical case of endodontic therapy of root with accentuated dilaceration using Reciproc single-file system. Mandibular right second molar was diagnosed as asymptomatic irreversible pulpitis. Pulp chamber access was performed, and glide path was created with #10 K-file (Dentsply Maillefer) and PathFile #13, #16 and #19 (Dentsply Maillefer) up to the temporary working length. The working length measured corresponded to 20 mm in the mesio-buccal and mesio-lingual canals, and 22 mm in the distal canal. The R25 file (VDW GmbH) was used in all the canals for instrumentation and final preparation, followed by filling with Reciproc gutta-percha cones (VDW GmbH) and AH Plus sealer (Dentsply Maillefer), using thermal compaction technique. The case has been receiving follow-up for 6 mon and no painful symptomatology or periapical lesions have been found. Despite the difficulties, the treatment could be performed in a shorter period of time than the conventional methods.

SELECTION OF CITATIONS