ABSTRACT
OBJECTIVES: Visualizing myocardium with near field ultrasound (NFUS) transducers in the tip of the catheter might provide an image of the evolving pathological lesion during energy delivery. BACKGROUND: Radiofrequency (RF) catheter ablation has been effective in arrhythmia treatment, but no technology has allowed lesion formation to be visualized in real time in vivo. METHODS: RF catheter ablations were performed in vivo with the goal to create transmural atrial lesions and large ventricular lesions. RF lesion formation was imaged in real time using M-mode, tissue Doppler, and strain rate information from the NFUS open irrigated RF ablation catheter incorporating 4 ultrasound transducers (1 axial and 3 radial), and growth kinetics were analyzed. Nineteen dogs underwent ablation in the right and left atria (n = 185), right ventricle (n = 67), and left ventricle (n = 66). Lesions were echolucent with tissue strain rate by NFUS. RESULTS: Lesion growth frequently progressed from epicardium to endocardium in thin-walled tissue. The half time of lesion growth was 5.5 ± 2.8 s in thin-walled and 9.7 ± 4.3 s in thick-walled tissue. Latency of lesion onset was seen in 57% of lesions ranging from 1 to 63.8 s. Tissue edema (median 25% increased wall thickness) formed immediately upon lesion formation in 83%, and intramyocardial steam was seen in 71% of cases. CONCLUSIONS: NFUS was effective in imaging RF catheter ablation lesion formation in real time. It was useful in assessing the dynamics of lesion growth and could visualize impending steam pops. It may be a useful technology to improve both safety and efficacy of RF catheter ablation.
Subject(s)
Catheter Ablation , Ultrasonography, Interventional , Animals , Atrial Fibrillation , Dogs , Heart Atria/diagnostic imaging , Heart Atria/surgeryABSTRACT
Commercially available intracardiac echo (ICE) catheters face a trade-off between viewing depth and resolution. Frequency-tunable ICE probes would offer versatility of choice between penetration or resolution imaging within a single device. In this phantom study, the imaging performance of a novel, frequency-tunable, 32-element, 1-D CMUT array integrated with front-end electronics is evaluated. Phased-array ultrasound imaging with a forward-looking CMUT probe prototype operated beyond collapse mode at voltages up to three times higher than the collapse voltage (-65â¯V) is demonstrated. Imaging performance as a function of bias voltage (-70â¯V to -160â¯V), transmit pulse frequency (5-25â¯MHz), and number of transmit pulse cycles (1-3) is quantified, based on which penetration, resolution, and generic imaging modes are identified. It is shown that by utilizing the concept of frequency tuning, images with different characteristics can be generated trading-off the resolution and penetration depth. The penetration mode provides imaging up to 71â¯mm in the tissue-mimicking phantom, axial resolution of 0.44â¯mm, and lateral resolution of 0.12â¯rad. In the resolution mode, axial resolution of 0.055â¯mm, lateral resolution of 0.035â¯rad, and penetration depth of 16â¯mm are measured. These results show what this CMUT array has the potential versatile characteristics needed for intracardiac imaging, despite its relatively small transducer aperture size of 2â¯mmâ¯×â¯2â¯mm imposed by the clinical application.
ABSTRACT
BACKGROUND: Safe and successful radiofrequency catheter ablation depends on creation of transmural lesions without collateral injury to contiguous structures. Near-field ultrasound (NFUS) imaging through transducers in the tip of an ablation catheter may provide important information about catheter contact, wall thickness, and ablation lesion formation. METHODS AND RESULTS: NFUS imaging was performed using a specially designed open-irrigated radiofrequency ablation catheter incorporating 4 ultrasound transducers. Tissue/phantom thickness was measured in vitro with varying contact angles. In vivo testing was performed in 19 dogs with NFUS catheters positioned in 4 chambers. Wall thickness measurements were made at 222 sites (excluding the left ventricle) and compared with measurements from intracardiac echocardiography. Imaging was used to identify the epicardium with saline infusion into the pericardial space at 39 sites. In vitro, the measured exceeded actual tissue/phantom thickness by 13% to 20%. In vivo, NFUS reliably visualized electrode-tissue contact, but sensitivity of epicardial imaging was 92%. The chamber wall thickness measured by NFUS correlated well with intracardiac echocardiography (r=0.86; P<0.0001). Sensitivity of lesion identification by NFUS was 94% for atrial and 95% for ventricular ablations. NFUS was the best parameter to predict lesion depth in right and left ventricle (r=0.47; P<0.0001; multiple regression P=0.0025). Lesion transmurality was correctly identified in 87% of atrial lesions. CONCLUSIONS: NFUS catheter imaging reliably assesses electrode-tissue contact and wall thickness. Its use during radiofrequency catheter ablation may allow the operator to assess the depth of ablation required for transmural lesion formation to optimize power delivery.
Subject(s)
Catheter Ablation/methods , Echocardiography/methods , Pericardium/diagnostic imaging , Pericardium/surgery , Animals , Cardiac Catheterization , Dogs , Fluoroscopy , Image Processing, Computer-Assisted , Phantoms, Imaging , Sensitivity and Specificity , TransducersABSTRACT
In intracardiac echocardiography (ICE) it may be beneficial to generate ultrasound images acquired at multiple frequencies, having the possibility of high penetration or high-resolution imaging in a single device. The objective of the presented work is to test two frequency-tunable probe prototypes in a preclinical setting: a rigid probe having a diameter of 11 mm and a new flexible and steerable 12-Fr ICE catheter. Both probes feature a forward-looking 32-element capacitive micromachined ultrasonic transducer array (aperture of 2 × 2 mm2) operated in collapse mode, which allows for frequency tuning in the 6-MHz-18-MHz range. The rigid probe prototype is tested ex vivo in a passive heart platform. Images of an aortic valve acquired in high-penetration (6 MHz), generic (12 MHz) and high-resolution (18 MHz) mode combine satisfying image quality and penetration depth between 2.5 cm and 10 cm. The ICE catheter prototype is tested in vivo using a porcine animal model. Images of an aortic valve are acquired in the 3 imaging modes with the ICE catheter placed in an ascending aorta at multiple depths. It was found that the combination of the forward-looking design and frequency-tuning capability allows visualizing intracardiac structures of various sizes at different distances relative to the catheter tip, providing both wide overviews and detailed close-ups.
Subject(s)
Aortic Valve/diagnostic imaging , Ultrasonography/instrumentation , Ultrasonography/methods , Animals , Equipment Design , Models, Animal , Swine , TransducersABSTRACT
INTRODUCTION: Steam pops are a risk of irrigated RF ablation even when limiting power delivery. There is currently no way to predict gas formation during ablation. It would be useful to visualize intramyocardial gas formation prior to a steam pop occurring using near-field ultrasound integrated into a RF ablation catheter. METHODS AND RESULTS: In an in vivo open-chest ovine model (n = 9), 86 lesions were delivered to the epicardial surface of the ventricles. Energy was delivered for 15-60 seconds, to achieve lesions with and without steam pops, based on modeling data. The ultrasound image was compared to a digital audio recording from within the pericardium by a blinded observer. Of 86 lesions, 28 resulted in an audible steam pop. For lesions that resulted in a steam pop compared to those that did not (n = 58), the mean power delivered was 8.0 ± 1.8 W versus 6.7 ± 2.0 W, P = 0.006. A change in US contrast due to gas formation in the tissue occurred in all lesions that resulted in a steam pop. In 4 ablations, a similar change in US contrast was observed in the tissue and RF delivery was stopped; in these cases, no pop occurred. The mean depth of gas formation was 0.9 ± 0.8 mm, which correlated with maximal temperature predicted by modeling. Changes in US contrast occurred 7.6 ± 7.2 seconds before the impedance rise and 7.9 ± 6.2 seconds (0.1-17.0) before an audible pop. CONCLUSION: Integrated US in an RF ablation catheter is able to visualize gas formation intramyocardially several seconds prior to a steam pop occurring. This technology may help prevent complications arising from steam pops.