Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Ear Hear ; 31(5): 599-610, 2010 Oct.
Article in English | MEDLINE | ID: mdl-20520553

ABSTRACT

OBJECTIVES: To develop normative data for wideband middle-ear reflectance in a newborn hearing-screening population and to compare test performance with 1-kHz tympanometry for prediction of otoacoustic emission (OAE) screening outcome. DESIGN: Wideband middle-ear reflectance (using both tone and chirp stimuli from 0.2 to 6 kHz), 1-kHz tympanometry, and distortion-product (DP) OAEs were measured in 324 infants at two test sites. Ears were categorized into DP pass and DP refer groups. RESULTS: Normative reflectance values were defined over various frequency regions for both tone and chirp stimuli in ambient pressure conditions, and for reflectance area indices integrated over various frequency ranges. Receiver operating characteristic analyses showed that reflectance provides the best discriminability of DP status in frequency ranges involving 2 kHz and greater discriminability of DP status than 1-kHz tympanometry. Repeated-measures analyses of variance established that (a) there were significant differences in reflectance as a function of DP status and frequency but not sex or ear; (b) tone and chirp stimulus reflectance values are essentially indistinguishable; and (c) newborns from two geographic sites had similar reflectance patterns above 1 kHz. Birth type and weight did not contribute to differences in reflectance. CONCLUSIONS: Referrals in OAE-based infant hearing screening were strongly associated with increased wideband reflectance, suggesting middle-ear dysfunction at birth. Reflectance improved significantly during the first 4 days after birth with normalization of middle-ear function. Reflectance scores can be achieved within seconds using the same equipment used for OAE screening. Newborns with high reflectance scores at stage I screening should be rescreened within a few hours to a few days, because most middle-ear problems are transient and resolve spontaneously. If reflectance and OAE are not passed upon stage II screening, referral to an otologist for ear examination is suggested along with diagnostic testing. Newborns with normal reflectance and a refer result for the OAE screen should be referred immediately to an audiologist for diagnostic testing with threshold auditory brainstem response because of higher risk for permanent hearing loss.


Subject(s)
Hearing Disorders/diagnosis , Hearing Tests/methods , Hearing Tests/standards , Neonatal Screening/methods , Neonatal Screening/standards , Acoustic Impedance Tests , Acoustic Stimulation/methods , Ear, Middle/physiology , Female , Hearing Disorders/physiopathology , Humans , Infant, Newborn , Male , Otoacoustic Emissions, Spontaneous , Reference Values
SELECTION OF CITATIONS
SEARCH DETAIL
...