Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters











Database
Language
Publication year range
1.
J Chromatogr A ; 1073(1-2): 309-15, 2005 May 06.
Article in English | MEDLINE | ID: mdl-15909535

ABSTRACT

A procedure was developed for the determination of caffeine and theophylline using a C18 column (5 microm, 250 mm x 4.6 mm) and micellar liquid chromatography using hybrid mobile phases containing sodium dodecyl sulfate (SDS) and propanol, butanol or pentanol as modifiers. Detection was performed with a variable wavelength UV-vis detector at 272 nm. After the application of an interpretative strategy for the selection of the optimimum mobile phase, caffeine and theophylline can be resolved and determined in serum samples by direct injection, using a mobile phase made up of 50 mM SDS-2.5% (v/v) propanol-10 mM KH2PO4, pH 7, with an analysis time below 5 min. Calibration was linear in the range 0.05 to 50 microg mL(-1) with r > 0.999. The statistical evaluation of the method was examined by performing intra-day (n = 6) and inter-day calibration (n = 7) and was found to be satisfactory, with highly accurate and precise results. The proposed method was suitably validated and applied to the determination of caffeine and theophylline in serum samples of patients treated with bronchodilators.


Subject(s)
Bronchodilator Agents/blood , Caffeine/blood , Theophylline/blood , Calibration , Chromatography, Liquid , Humans , Reproducibility of Results , Sensitivity and Specificity , Spectrophotometry, Ultraviolet
SELECTION OF CITATIONS
SEARCH DETAIL