Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 3 de 3
Filter
Add more filters










Database
Language
Publication year range
1.
Radiat Prot Dosimetry ; 200(8): 802-807, 2024 Jun 03.
Article in English | MEDLINE | ID: mdl-38773775

ABSTRACT

Interventional cardiology is characterized by high radiation exposure for both the patient and the operator. Adequate shielding and monitoring of the operator are fundamental to comply with radiation protection principles. In a previous work, the effect on the dose of the dosemeter position on the chest was studied. In this paper, the investigation has been completed, employing an anthropomorphic thorax phantom, equipped with arms. Although there are differences between the Monte Carlo simulations and the measurements, similar trends are observed, showing that the reduction in dose, due to the arms, is between 20 and 60%, compared with the situation without arms. For that reason, considering a dosemeter placed on the chest, the upper position, which is the least affected by the arms, should be preferred while the extreme lateral position, near the armpit, should be avoided.


Subject(s)
Computer Simulation , Monte Carlo Method , Occupational Exposure , Phantoms, Imaging , Protective Clothing , Radiation Dosage , Radiation Monitoring , Radiation Protection , Humans , Occupational Exposure/analysis , Occupational Exposure/prevention & control , Radiation Protection/methods , Radiation Monitoring/methods , Cardiology/methods , Medical Staff , Thorax/radiation effects
2.
Radiat Prot Dosimetry ; 180(1-4): 360-364, 2018 Aug 01.
Article in English | MEDLINE | ID: mdl-29053837

ABSTRACT

The TOP-IMPLART, a new proton therapy facility, is under development in Frascati ENEA Laboratories, near Rome. The project is centered on a medium-energy proton accelerator designed as a sequence of modular linear accelerators (the final energy will be 230 MeV). Being not a commercial product, measurements and simulation are fundamental to characterize the system and the radiation field, even during its construction. In this work some preliminary evaluations of the neutron contamination have been tried. The simulations were validated through some measurements obtaining a satisfactory agreement. A more detailed calculations and measurements campaign is scheduled for the next future.


Subject(s)
Facility Design and Construction , Neutrons , Particle Accelerators/instrumentation , Proton Therapy/instrumentation , Proton Therapy/methods , Computer Simulation , Humans , Monte Carlo Method , Radiation Protection , Radiotherapy Dosage
3.
Radiat Prot Dosimetry ; 170(1-4): 145-9, 2016 Sep.
Article in English | MEDLINE | ID: mdl-26170436

ABSTRACT

Since the publication of International Commission on Radiological Protection statement in 2011 on tissue reaction, eye lens radiation protection played an important role in exposed personnel dosimetry. For this reason, the Italian National Agency for New Technologies, Energy and Sustainable Economic Development (ENEA) Individual Monitoring Service decided to study a prototype to fulfil specific requests (e.g. for survey in interventional department and intercomparisons). On the basis of such preliminary investigation, a new eye lens dosemeter was developed. The new dosemeter, named EDEL (ENEA Dosemeter for Eye Lens), was characterised in terms of Hp(3), the operational quantity related to eye lens monitoring. The investigation was performed experimentally and optimised using the Monte Carlo MCNP6 code. The new prototype was thought to fulfil two main requests: the reliability of the dosimetric data and the portability of the dosemeter itself. The new dosemeter will soon be supplied to the collaborating hospitals for workplace test measurements.


Subject(s)
Lens, Crystalline/radiation effects , Occupational Exposure/analysis , Radiation Dosimeters , Radiation Monitoring/instrumentation , Radiation Protection/instrumentation , Thermoluminescent Dosimetry/instrumentation , Calibration , Hospitals , Humans , Monte Carlo Method , Occupational Exposure/prevention & control , Radiation Dosage , Radiation Monitoring/methods , Radiation Protection/methods , Reproducibility of Results , Software , Thermoluminescent Dosimetry/methods
SELECTION OF CITATIONS
SEARCH DETAIL