Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add more filters










Database
Language
Publication year range
1.
Rev Sci Instrum ; 90(5): 053701, 2019 May.
Article in English | MEDLINE | ID: mdl-31153272

ABSTRACT

A transmission X-ray microscope has been designed and commissioned at the 18-ID Full-field X-ray Imaging beamline at the National Synchrotron Light Source II. This instrument operates in the 5-11 keV range, and, with the current set of optics, is capable of 30 nm spatial resolution imaging, with a field of view of about 40 µm. For absorption contrast, the minimum exposure time for a single projection image is about 20 ms and an entire 3D tomography data set can be acquired in under 1 min. The system enables tomographic reconstructions with sub-50 nm spatial resolution without the use of markers on the sample or corrections for rotation run-outs.

2.
J Synchrotron Radiat ; 18(Pt 6): 862-70, 2011 Nov.
Article in English | MEDLINE | ID: mdl-21997910

ABSTRACT

Characterization and testing of an L-shaped laterally graded multilayer mirror are presented. This mirror is designed as a two-dimensional collimating optics for the analyzer system of the ultra-high-resolution inelastic X-ray scattering (IXS) spectrometer at National Synchrotron Light Source II (NSLS-II). The characterization includes point-to-point reflectivity measurements, lattice parameter determination and mirror metrology (figure, slope error and roughness). The synchrotron X-ray test of the mirror was carried out reversely as a focusing device. The results show that the L-shaped laterally graded multilayer mirror is suitable to be used, with high efficiency, for the analyzer system of the IXS spectrometer at NSLS-II.

SELECTION OF CITATIONS
SEARCH DETAIL
...