ABSTRACT

In this paper, we describe development of a high-throughput, highly sensitive method based on Lab Chip CGE-SDS platform for purity determination and characterization of virus-like particle (VLP) vaccines. A capillary gel electrophoresis approach requiring about 41 s per sample for analysis and demonstrating sensitivity to protein initial concentrations as low as 20 µg/mL, this method has been used previously to evaluate monoclonal antibodies, but this application for lot release assay of VLPs using this platform is unique. The method was qualified and shown to be accurate for the quantitation of VLP purity. Assay repeatability was confirmed to be less than 2% relative standard deviation of the mean (% RSD) with interday precision less than 2% RSD. The assay can evaluate purified VLPs in a concentration range of 20-249 µg/mL for VEE and 20-250 µg/mL for EEE and WEE VLPs.


Subject(s)
Electrophoresis, Capillary/methods , Encephalitis Virus, Eastern Equine/isolation & purification , Encephalitis Virus, Venezuelan Equine/isolation & purification , Encephalitis Virus, Western Equine/isolation & purification , High-Throughput Screening Assays/methods , Vaccines, Virus-Like Particle/isolation & purification , Fluorescence , Fluorescent Dyes/chemistry , Humans , Sensitivity and Specificity , Vaccines, Virus-Like Particle/chemistry