Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Appl Radiat Isot ; 60(6): 921-7, 2004 Jun.
Article in English | MEDLINE | ID: mdl-15110358

ABSTRACT

A (82)Rb infusion system is described with two important features for imaging with 3D positron emission tomography. First, a generator bypass line is added to flush the patient infusion line at the end of an elution. Second, feedback control is implemented to permit 'slow-bolus' constant-activity elutions. A model for the activity eluted from a (82)Sr/(82)Rb generator based on a volume-activity empirical relationship, is used as the basis for performing simulations to demonstrate the efficacy of varying the flow rate through the generator to achieve desired eluted (82)Rb activity rate profiles. A (82)Rb infusion system was constructed to verify the accuracy of the simulations. The system can deliver accurate constant-activity elutions from 10% to 70% of the total generator activity.


Subject(s)
Rubidium Radioisotopes , Tomography, Emission-Computed/methods , Coronary Disease/diagnostic imaging , Humans , Infusions, Intravenous , Perfusion , Reproducibility of Results , Rubidium Radioisotopes/administration & dosage , Rubidium Radioisotopes/pharmacokinetics
SELECTION OF CITATIONS
SEARCH DETAIL
...