Lab Med ; 49(1): e1-e8, 2017 Dec 22.
Article in English | MEDLINE | ID: mdl-29253199

ABSTRACT

BACKGROUND: To validate the performance, reproducibility, and reliability of BD automated instruments in order to establish a fully automated clinical microbiology laboratory. METHODS: We used control strains and clinical samples to assess the accuracy, reproducibility, and reliability of the BD Kiestra WCA, the BD Phoenix, and BD Bruker MALDI-Biotyper instruments and compared them to previously established conventional methods. The following processes were evaluated: sample inoculation and spreading, colony counts, sorting of cultures, antibiotic susceptibility test, and microbial identification. RESULTS: The BD Kiestra recovered single colonies in less time than conventional methods (e.g. E. coli, 7h vs 10h, respectively) and agreement between both methodologies was excellent for colony counts (κ=0.824) and sorting cultures (κ=0.821). Antibiotic susceptibility tests performed with BD Phoenix and disk diffusion demonstrated 96.3% agreement with both methods. Finally, we compared microbial identification in BD Phoenix and Bruker MALDI-Biotyper and observed perfect agreement (κ=1) and identification at a species level for control strains. Together these instruments allow us to process clinical urine samples in 36h (effective time). CONCLUSION: The BD automated technologies have improved performance compared with conventional methods, and are suitable for its implementation in very busy microbiology laboratories.


Subject(s)
Automation, Laboratory , Bacteriological Techniques , Automation, Laboratory/instrumentation , Automation, Laboratory/methods , Automation, Laboratory/standards , Bacterial Load , Bacteriological Techniques/instrumentation , Bacteriological Techniques/methods , Bacteriological Techniques/standards , Humans , Microbial Sensitivity Tests , Molecular Typing , Reproducibility of Results , Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization
SELECTION OF CITATIONS
SEARCH DETAIL