1.
Invest Radiol ; 55(5): 285-292, 2020 05.
Article in English | MEDLINE | ID: mdl-32011572

ABSTRACT

INTRODUCTION: Magnetic resonance imaging (MRI) of the abdomen increasingly incorporates diffusion-weighted imaging (DWI) sequences. Whereas DWI can substantially aid in detecting and characterizing suspicious findings, it remains unclear to what extent the use of ultra-high b-value DWI might further be of aid for the radiologist especially when using DWI sequences with advanced processing. The target of this study was therefore to compare high and ultra-high b-value DWI in abdominal MRI examinations. METHODS: This institutional review board-approved, prospective study included abdominal MRI examinations of 70 oncologic patients (mean age, 58 years; range, 21-90 years) examined with a clinical 1.5 T MRI scanner (MAGNETOM Aera, Siemens Healthcare, Erlangen, Germany) with an advanced echo planar DWI sequence (b = 0, 50, 900, and 1500 s/mm) after ex vivo phantom and in vivo volunteer investigations. High b900 and ultra-high b1500 DWIs were compared by a qualitative reading for image quality and lesion conspicuity using a 5-point Likert scale with 2 radiologists as readers. The ratios of apparent signal intensities of suspicious lesions/normal tissue of the same organ (LNTRs) were calculated. Appropriate methods were used for statistical analysis, including Wilcoxon signed-rank test and κ statistic for interreader agreement analysis (P < 0.05/0.0125/0.005 after Bonferroni correction). RESULTS: Image quality was significantly increased with b900 as compared with b1500 DWI (P < 0.001) despite using an advanced DWI sequence. A total of 153 suspicious lesions were analyzed. Overall reader confidence for characterization/detection of malignant lesions and, correspondingly, the LNTR (mean, 2.7 ± 1.8 vs 2.4 ± 1.6) were significantly higher with b900 than with b1500 DWI (P < 0.001 and P < 0.001). The increased confidence of lesion recognition and LNTR in the b900 DWI remained significant qualitatively in lymphatic and hepatic lesions and quantitatively in lymphatic, pulmonal, and osseous lesions. CONCLUSIONS: Using high b-value DWI (900 s/mm) provided an improved image quality and also lesion conspicuity as compared with ultra-high b-value DWI (1500 s/mm) in oncologic abdominal examinations despite using advanced processing. Consequently, the value for additional ultra-high b-value DWI in oncologic examinations should be critically evaluated in future studies.


Subject(s)
Abdomen/diagnostic imaging , Abdominal Neoplasms/diagnosis , Diffusion Magnetic Resonance Imaging/methods , Adult , Aged , Aged, 80 and over , Female , Humans , Male , Middle Aged , Phantoms, Imaging , Prospective Studies , Reproducibility of Results , Young Adult
2.
Invest Radiol ; 54(11): 704-711, 2019 11.
Article in English | MEDLINE | ID: mdl-31356384

ABSTRACT

INTRODUCTION: Diffusion-weighted imaging (DWI) is an important part of oncological magnetic resonance imaging (MRI) examinations, especially for tertiary cancer prevention in terms of early detection of recurrent disease. However, abdominal studies can be challenged by motion artifacts, poor signal-to-noise ratios, and visibility of retroperitoneal structures, which necessitates sequence optimization depending on the investigated region. This study aims at prospectively evaluating an adapted DWI sequence ex vivo and in vivo in oncologic patients undergoing abdominal MRI. METHODS: This institutional review board-approved, prospective study included phantom measurements, volunteer examinations, and oncologic patient examinations of the abdomen. Fifty-seven MRI examinations in 54 patients (mean age, 58 years; range, 21-90 years) were included into the analysis. The MRI examination were performed at a 1.5 T MRI scanner (MAGNETOM Aera; Siemens Healthcare, Erlangen, Germany) and included both a standard EPI-DWI (s-DWI; b = 50, 900 s/mm) and an adapted DWI (opt-DWI; EPI-DWI with b = 0, 50, 900, 1500 s/mm, acquisition with higher spatial resolution and optimized processing for the abdomen including motion correction, adaptive image combination, and background suppression). For b = 900 s/mm, the ratio of signal intensity in the normal tissue and the standard deviation of the noise in the surrounding air was quantitatively calculated; image quality and tissue differentiation parameters were rated by 2 independent, blinded readers using a 5-point Likert scale. Statistics included Wilcoxon signed-rank test and kappa statistic (P < 0.05/0.0125 after Bonferroni correction). RESULTS: The DWI phantom demonstrated an optimized contour sharpness and inlay differentiation for opt-DWI. The apparent ratio of normal tissue signal/standard deviation of background noise at b = 900 s/mm of the right/left hemiabdomen was significantly increased in opt-DWI (mean, 71.9 ± 23.5/86.0 ± 43.3) versus s-DWI (mean, 51.4 ± 15.4/63.4 ± 36.5; P < 0.001). Image quality parameters (contour sharpness and tissue differentiation of upper abdominal and retroperitoneal structures) were significantly increased in opt-DWI versus s-DWI (P < 0.001). Interreader reliability test showed good agreement (kappa = 0.768; P < 0.001). DISCUSSION: This study prospectively evaluated the potential of adapted DWI for screening in tertiary prevention of oncologic patients. An optimized DWI protocol with advanced processing achieved improved image quality in quantitative and qualitative analyses. Oncological optimization of DWI should be performed before its application in cancer patients to improve both screening and follow-up examinations, to better unleash the diagnostic potential of DWI.


Subject(s)
Abdominal Neoplasms/diagnostic imaging , Diffusion Magnetic Resonance Imaging/methods , Image Interpretation, Computer-Assisted/methods , Tertiary Prevention/methods , Adult , Aged , Aged, 80 and over , Artifacts , Female , Humans , Male , Middle Aged , Motion , Phantoms, Imaging , Prospective Studies , Reproducibility of Results , Signal-To-Noise Ratio , Young Adult