Appl Opt ; 58(23): 6257-6267, 2019 Aug 10.
Article in English | MEDLINE | ID: mdl-31503767

ABSTRACT

We have constructed a Fourier-transform spectrometer (FTS) operating between 50 GHz and 330 GHz with minimum volume (355×260×64 mm) and weight (5.9 kg) while maximizing optical throughput (100 mm2 sr) and optimizing the spectral resolution (4 GHz). This FTS is designed as a polarizing Martin-Puplett interferometer with unobstructed input and output in which both input polarizations undergo interference. The instrument construction is simple with mirrors milled on the box walls and one motorized stage as the single moving element. We characterize the performance of the FTS, compare the measurements to an optical simulation, and discuss features that relate to details of the FTS design. The simulation is also used to determine the tolerance of optical alignments for the required specifications. We detail the FTS mechanical design and provide the control software as well as the analysis code online.