Biosensors (Basel) ; 13(3)2023 Feb 23.
Article in English | MEDLINE | ID: mdl-36979522

ABSTRACT

In this paper, we present a novel and cost-effective lab-on-paper microfluidics platform for performing ELISA autonomously, with no user intervention beyond adding the sample. The platform utilizes two Bi-Material Cantilever Valves placed in a specially designed housing. The integration of these valves in a specific channel network forms a complete fluidic logic circuit for performing ELISA on paper. The housing also incorporates an innovative reagent storage and release mechanism that minimizes variability in the volume of reagents released into the reagent pads. The platform design was optimized to minimize variance in the time of fluid wicking from the reagent pad, using a randomized design of experiment. The platform adheres to the World Health Organization's ASSURED principles. The optimized design was used to conduct an ELISA for detecting rabbit immunoglobulin G (IgG) in a buffer, with a limit of detection of 2.27 ng/mL and a limit of quantification of 8.33 ng/mL. This represents a 58% improvement over previous ELISA methods for detecting rabbit IgG in buffer using portable microfluidic technology.


Subject(s)
Microfluidic Analytical Techniques , Microfluidics , Animals , Rabbits , Enzyme-Linked Immunosorbent Assay , Immunoglobulin G/analysis , Lab-On-A-Chip Devices , Microfluidic Analytical Techniques/methods , Microfluidics/methods