Ultrasound Med Biol
; 41(3): 890-7, 2015 Mar.
Article
in English
| MEDLINE
| ID: mdl-25542496
ABSTRACT
There are currently very few test objects suitable for high-frequency ultrasound scanners that can be rapidly manufactured, have appropriate acoustic characteristics and are suitably robust. Here we describe techniques for the creation of a wall-less flow phantom using a physically robust konjac and carrageenan-based tissue-mimicking material. Vessel dimensions equivalent to those of mouse and rat arteries were achieved with steady flow, with the vessel at a depth of 1.0 mm. We then employed the phantom to briefly investigate velocity errors using pulsed wave Doppler with a commercial preclinical ultrasound system. This phantom will provide a useful tool for testing preclinical ultrasound imaging systems.