Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Int J Cardiovasc Intervent ; 4(1): 21-27, 2001 Mar.
Article in English | MEDLINE | ID: mdl-12431336

ABSTRACT

BACKGROUND: The mechanism for the disappointing late outcome following stenting of bifurcation lesions is unclear. This prospective observational study aims to evaluate culotte stent deployment and dimensions with intravascular ultrasound (IVUS). PATIENTS AND METHODS: Patients with bifurcation stenoses were treated using two stents in a culotte configuration. After optimizing the angiographic appearance of both stents, IVUS was used to evaluate both limbs of the culotte. The main outcome measures were cross-sectional area (CSA) and minimal lumen diameter (MLD) assessed by IVUS. RESULTS: Within the culotte stent, the final mean CSA in the main limb was 6.1 mm(2) (97% of reference) and in the side-limb was 5.9 mm(2) (97% of reference). However, in each case, the minimum CSA and IVUS MLD of both limbs was at the bifurcation point. For all patients, the final mean CSA at the bifurcation point of the main limb was 4.3 mm(2) (70% of main stent) and of the side-limb was 4.4 mm(2) (75% of side stent). The IVUS MLD at the bifurcation point of the main limb was 2.1 mm (78% of main stent) and of the side-limb was 2.1 mm (84% of the side stent). Importantly, this significant residual stenosis was not detectable with quantitative coronary angiography. CONCLUSIONS: IVUS evaluation of culotte stents is feasible. The minimum IVUS CSA and MLD of both limbs of the culotte stent is at the bifurcation point. Despite an optimal angiographic appearance a significant residual stenosis was noted with IVUS at each bifurcation point.

SELECTION OF CITATIONS
SEARCH DETAIL
...