1.
Med Phys ; 50(12): 7427-7440, 2023 Dec.
Article in English | MEDLINE | ID: mdl-37824821

ABSTRACT

PURPOSE: A comprehensive, centrally-monitored physics quality control (QC) program was developed for the Tomosynthesis Imaging Screening Trial (TMIST), a randomized controlled trial of digital breast tomosynthesis (TM) versus digital mammography (DM) for cancer screening. As part of the program, in addition to a set of phantom-based tests, de-identified data on image acquisition and processing parameters were captured from the DICOM headers of all individual patient images in the trial. These data were analyzed to assess the potential usefulness of header data from digital mammograms and tomosynthesis images of patients for quality assurance in breast imaging. METHODS: Data were automatically extracted from the headers of all de-identified patient mammograms and tomosynthesis images in the TMIST study. Image acquisition parameters and estimated radiation doses were tracked for individual sites, systems and across system types. These parameters included (among others) kV, target/filter use, number of acquired views per examination, AEC mode, compression thickness and force and detector temperature. Consistency of manually entered study data parameters (subject ID, screening time-point) from TMIST was evaluated. Preliminary observations from the program are presented. RESULTS: We report on data from 812 651 images from 135 525 examinations acquired between October, 2017 and December, 2022. Data came from 6 system models from 3 manufacturers. There was greater variability both in the number of views used and in the estimated (proxy) doses received in DM exams compared to TM. Mean proxy doses per examination varied among manufacturers from 2.76-4.54 mGy for DM and 3-4.84 mGy for the tomosynthesis component in the TM arm with maximum examination proxy doses of 20 and 26 mGy for DM and TM respectively. Mean proxy doses per examination for the combination examination in TM (tomosynthesis plus digital mammography) varied from 6.6 to 7.6 mGy among manufacturers with a maximum of 44.5 mGy. CONCLUSIONS: Overall, modern digital mammography and tomosynthesis systems used in TMIST have operated very reliably. Doses vary considerably due to variation in the number of views per examination, thickness and fibro-glandularity of the breast, and choices in the use of synthesized versus actual 2D mammography in the TM examination. These data may also be useful in predicting equipment problems. Header information is valuable not only for automated QC, but also for cross-checking accuracy and consistency of data in a clinical study.


Subject(s)
Breast Neoplasms , Early Detection of Cancer , Humans , Female , Radiation Dosage , Mammography/methods , Breast/diagnostic imaging , Phantoms, Imaging , Breast Neoplasms/diagnostic imaging
2.
Med Phys ; 50(12): 7441-7461, 2023 Dec.
Article in English | MEDLINE | ID: mdl-37830895

ABSTRACT

BACKGROUND: The Tomosynthesis Mammography Imaging Screening Trial (TMIST), EA1151 conducted by the Eastern Cooperative Oncology Group (ECOG)/American College of Radiology Imaging Network (ACRIN) is a randomized clinical trial designed to assess the effectiveness for breast cancer screening of digital breast tomosynthesis (TM) compared to digital mammography (DM). Equipment from multiple vendors is being used in the study. PURPOSE: For the findings of the study to be valid and capture the true capacities of the two technology types, it is important that all equipment is operated within appropriate parameters with regard to image quality and dose. A harmonized QC program was established by a core physics team. Since there are over 120 trial sites, a centralized, automated QC program was chosen as the most practical design. This report presents results of the weekly QC testing program. A companion paper will review quality monitoring based on data from the headers of the patient images. METHODS: Study images are collected centrally after de-identification using the "TRIAD" application developed by ACR. The core physics team devised and implemented a minimal set of quality control (QC) tests to evaluate the tomosynthesis and 2D mammography systems. Weekly, monthly and annual testing is performed by the site mammography technologists with images submitted directly to the physics core. The weekly physics QC tests are described: SDNR of a low-contrast mass object, artifact spread, spatial resolution, tracking of technical factors, and in-slice noise power spectra. RESULTS: As of December 31, 2022 (5 years), 145 sites with 411 machines had submitted QC data. A total of 136 742 TMIST participant screening imaging studies had been performed. The 5th and 95th percentile mean glandular doses for a single tomosynthesis exposure to a 4.0 cm thick PMMA phantom ("standard breast phantom") were 1.24 and 1.68 mGy respectively. The largest sources of QC non-conformance were: operator error, not following the QC protocol exactly, unreported software updates and preventive maintenance activities that affected QC setpoints. Noise power spectra were measured, however, standardization of performance targets across machine types and software revisions was difficult. Nevertheless, for each machine type, test measurement results were very consistent when the protocol was followed. Deviations in test results were mostly related to software and hardware changes. CONCLUSION: Most systems performed very consistently. Although this is a harmonized program using identical phantoms and testing protocols, it is not appropriate to apply universal threshold or target metrics across the machine types because the systems have different non-linear reconstruction algorithms and image display filters. It was found to be more useful to assess pass/fail criteria in terms of relative deviations from baseline values established when a system is first characterized and after equipment is changed. Generally, systems which needed repair failed suddenly, but in retrospect, for a few cases, drops in SDNR and increases in mAs were observed prior to tube failure. TMIST is registered as NCT03233191 by Clinicaltrials.gov.


Subject(s)
Breast Neoplasms , Mammography , Humans , Female , Mammography/methods , Breast , Breast Neoplasms/diagnostic imaging , Algorithms , Quality Control , Phantoms, Imaging
3.
Med Phys ; 48(7): 3623-3629, 2021 Jul.
Article in English | MEDLINE | ID: mdl-33931863

ABSTRACT

PURPOSE: In the reconstruction of volume breast images from x-ray projections in breast tomosynthesis, some tomographic systems truncate the image data presented to the radiologist such that a non-negligible amount of tissue may be missing from the breast image. QC tests were conducted to determine if this problem existed in imaging in the TMIST study. METHODS: Test tools developed for TMIST containing small objects at known heights were used in routine weekly and annual QC testing of tomosynthesis units to assess the degree to which phantom material that was irradiated in imaging was excluded from the reconstructed image. Results from 318 tests on five system types from three manufacturers are reported. RESULTS: The presence and extent of this problem varied among system types. The cause was most frequently related to machine errors in the determination of breast thickness or to deflection of components during breast compression. In particular, the problem occurred when a compression paddle other than the one calibrated for tomosynthesis was used for the tests. This was also verified to have occurred in some clinical imaging. CONCLUSIONS: Missing volume can be avoided by intentionally reconstructing additional image slices above and below the presumed locations of the breast support and compression plate. A compression paddle which has been calibrated for tomosynthesis should be used both for clinical imaging and testing. The prevalence of this phenomenon suggests that more frequent testing for volume coverage may be advisable.


Subject(s)
Breast , Data Compression , Breast/diagnostic imaging , Mammography , Phantoms, Imaging
4.
Med Phys ; 43(8): 4803, 2016 Aug.
Article in English | MEDLINE | ID: mdl-27487898

ABSTRACT

PURPOSE: The purpose of this work is to improve the repeatability of the measurement of the slice-sensitivity profile (SSP) in reconstructed breast tomosynthesis volumes. METHODS: A grid of aluminum ball-bearings (BBs) within a PMMA phantom was imaged on breast tomosynthesis systems from three different manufacturers. The full-width half-maximum (FWHM) values were measured for the SSPs of the BBs in the reconstructed volumes. The effect of transforming the volumes from a Cartesian coordinate system (CCS) to a cone-beam coordinate system (CBCS) on the variability in the FWHM values was assessed. RESULTS: Transforming the volumes from a CCS to a CBCS before measuring the SSPs reduced the coefficient of variation (COV) in the measurements of FWHM in repeated measurements by 56% and reduced the dependence of the FWHM values on the location of the BBs within the reconstructed volume by 76%. CONCLUSIONS: Measuring the SSP in the volumes in a CBCS improves the robustness of the measurement.


Subject(s)
Breast/diagnostic imaging , Polymethyl Methacrylate , Radiographic Image Enhancement/methods , Signal-To-Noise Ratio , Phantoms, Imaging