Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters











Database
Language
Publication year range
1.
Mater Sci Eng C Mater Biol Appl ; 48: 213-9, 2015 Mar.
Article in English | MEDLINE | ID: mdl-25579916

ABSTRACT

We are introducing mercapto-mesoporous carbon modified carbon paste electrode (mercapto-MP-C-CPE) as a new sensor for trace determination of omeprazole (OM) in biological samples. The synthesized modifier was characterized by thermogravimetry analysis (TGA), differential thermal analysis (DTA), transmission electron microscopy (TEM), Fourier transform infrared spectrometry (FT-IR), X-ray diffraction (XRD), elemental analysis (CHN) and N2 adsorption surface area measurement (BET). The electrochemical response characteristic of the modified-CPE toward OM was investigated by cyclic and differential pulse voltammetry (CV and DPV). The proposed sensor displayed a good electrooxidation response to the OM, its linear range is 0.25nM to 25µM with a detection limit of 0.04nM under the optimized conditions. The prepared modified electrode shows several advantages such as high sensitivity, long-time stability, wide linear range, ease of preparation and regeneration of the electrode surface by simple polishing and excellent reproducibility.


Subject(s)
Carbon/chemistry , Electrochemical Techniques/instrumentation , Electrochemical Techniques/methods , Electrodes , Omeprazole/analysis , Humans , Limit of Detection , Microscopy, Electron, Transmission , Omeprazole/blood , Omeprazole/urine , Porosity , Reproducibility of Results , Spectroscopy, Fourier Transform Infrared , Thermogravimetry , X-Ray Diffraction
SELECTION OF CITATIONS
SEARCH DETAIL