Cardiol J ; 31(1): 124-132, 2024.
Article in English | MEDLINE | ID: mdl-37519054

ABSTRACT

BACKGROUND: The aim of the study was to evaluate bio-functionality of a novel, proprietary balloon-expandable biological transcatheter aortic valve implantation (TAVI) system (InFlow, CardValve Consortium, Poland) in an ovine model of aortic banding. METHODS: Surgical ascending aorta banding was created in 21 sheep. Two weeks later, 18 biological valves were implanted within the model using 15-16 F InFlow TAVI systems and carotid cut-down approach. Follow-up transthoracic echocardiography was performed at 30, 90, and 180-day. At designated time, animals were euthanized and valves harvested for analysis. RESULTS: All sheep survived the banding procedure. There were 4 (22%) procedure related deaths within a 7-day period. During the observation an additional 2 sheep died. In one, the valve dislocated after the procedure - the animal was excluded. Two animals completed 30-day follow up, five 90-day follow-up and four terminal follow-up of 180 days. Valves examined via transesophageal echocardiography showed proper hemodynamic parameters without evidence of structural valve deterioration. The maximum and average flow gradients at 180 days were 31.4 (23.3-37.7) and 17.5 (13.1-20.2) mmHg, respectively. There was one case of moderate insufficiency and no case of perivalvular leaks. By histopathology, there were no inflammation, thrombosis, nor calcifications in any tested valves at long-term follow-up. Neointimal coverage of stent struts increased with time from basal part in "early" groups to nearly 3/4 of stent length in the 180-day group. The pannus tissue showed maturation that increased with time with no stenotic "collar" visible in orthotopically implanted valves. CONCLUSIONS: The study showed good hemodynamic performance, durability and biocompatibility of the novel biological THV.


Subject(s)
Aortic Valve Stenosis , Heart Valve Prosthesis , Transcatheter Aortic Valve Replacement , Animals , Sheep , Aortic Valve/diagnostic imaging , Aortic Valve/surgery , Aortic Valve Stenosis/surgery , Prosthesis Design , Transcatheter Aortic Valve Replacement/methods , Treatment Outcome
2.
Kardiol Pol ; 2023 Apr 25.
Article in English | MEDLINE | ID: mdl-37096947

ABSTRACT

BACKGROUND: The Coordinated Care in Myocardial Infarction Program (KOS-MI) was introduced to improve prognosis for patients after myocardial infarction (MI). The program includes complete revascularization followed by unrestricted access to rehabilitation, electrotherapy and cardiac care. AIM: The aim of this study was to assess major adverse cardiac and cerebrovascular events (MACCE) of patients enrolled in the KOS-MI at 3-year follow-up. METHODS: This is a retrospective, multicenter registry of patients treated for MI. Study group (KOS-MI) of 963 patients was compared to the control group (standard of care) of 1009 patients. At 3-year follow-up MACCE including death, MI, stroke and repeated revascularization were reported. Additionally, hospitalization due to heart failure (HF) was analyzed. Propensity score matching (PSM) was utilized for group baseline characteristics adjustment. RESULTS: Patients in the KOS-MI group were younger (65 vs. 68; P < 0.001), mostly men (70% vs. 62.9%; P < 0.001), admitted with ST-elevation myocardial infarction (STEMI) (44.6% vs. 36.2%; P < 0.001). Patients in the control group had more comorbidities and were admitted more often with non ST-elevation myocardial infarction (63.8% vs. 55.4%; P < 0.001) and acute HF (5.1% vs. 2.7%; P = 0.007). Following PSM 530 well matched pairs were selected. At three years (92.3% follow-up completeness), the relative risk reduction was: 25% in MACCE (P = 0.008), 38% in mortality (P = 0.008), 29% in repeated revascularization(P = 0.04) and 28% (P = 0.0496) in hospitalization for HF in the KOS-MI group. CONCLUSIONS: The combination of contemporary invasive techniques, complete revascularization, cardiac rehabilitation and ambulatory care included in the KOS-MI Program improves long-term prognosis of patients after MI up to 3-year follow-up.

3.
JACC Basic Transl Sci ; 7(5): 486-495, 2022 May.
Article in English | MEDLINE | ID: mdl-35663633

ABSTRACT

The aim of the study was to evaluate a balloon expandable transcatheter heart valve (THV) system (Myval) at 6-month follow-up in ovine banding model. Eleven THV systems were implanted via carotid approach. There were 2 procedure-related deaths and 2 premature deaths. At 6 months all valves that completed follow-up (n = 7) were functional, with no significant regurgitation, calcification, thrombi, or vegetation. Mean pressure gradient was 21.9 ± 11 mm Hg, maximum velocity = 3.3 ± 1 m/s, and ejection fraction was 53.3 ± 6%. Myval THV showed optimal hemodynamic performance and biocompatibility.

4.
Front Cardiovasc Med ; 9: 977006, 2022.
Article in English | MEDLINE | ID: mdl-36606288

ABSTRACT

Objectives: The aim of the study is to evaluate the functionality, durability, and temporal biocompatibility of a novel, balloon-expandable polymeric transcatheter heart valve (ATHV) system (InFlow, CardValve Consortium, Poland). Along with expanding TAVI indications, the demand for new transcatheter valves is increasing. Methods: A surgical ascending aortic banding model was created in 20 sheep. Two weeks later, 16 sheep were implanted with ATHV systems (15-16F). Three animals were euthanized after a 30-day follow-up, four animals after a 90-day follow-up, and six animals after a 180-day follow-up. A follow-up transthoracic echocardiography (TTE) was performed. Results: There was one procedure-related (6,25%) and two model-related deaths (12,5%; banding site calcification with subsequent infection originating externally from banding). TTE revealed the flow gradients (max/average) of 30,75/17,91; 32,57/19,21; and 21,34/10,63 mmHg at 30, 90, and 180 days, respectively. There were two cases of low-degree regurgitation after 180 days with no perivalvular leak observed. Histopathological analysis showed no valve degeneration at terminal follow-up with optimal healing. Small thrombi were present at the aortic wall adjacent to the base of the leaflets, and between the aortic wall and the stent in most of the valves; however, leaflets remained free from thrombi in all cases. Scanty calcifications of leaflets were reported in three animals evaluated 180 days after implantation. Conclusion: This preclinical study in the aortic banding model showed good hemodynamic performance, durability, and biocompatibility of the novel ATHV. Furthermore, regulatory studies with longer follow-ups are warranted.

5.
Vasc Med ; 26(4): 401-408, 2021 08.
Article in English | MEDLINE | ID: mdl-33686879

ABSTRACT

The aim of the BIOPAC trial was to determine long-term safety and efficacy of a novel microcrystalline paclitaxel-coated balloon (mcPCB) with a biocompatible polymer as an excipient in the treatment of occlusive femoropopliteal lesions. In this first-in-human prospective controlled randomized trial, 66 patients with femoropopliteal, symptomatic (Rutherford stages 2B to 5) occlusive arterial disease were randomized to either mcPCB (study group) or POBA (plain old balloon angioplasty) (control group) on a 1:1 basis. Late lumen loss (LLL) at 6 months was the primary endpoint of the study and serious adverse events (SAE: death, amputation, repeated revascularization) were considered a composite secondary endpoint. Routine angiography was scheduled for all study subjects at 6-month follow-up; outpatient appointments were scheduled at 12 and 36 months after intervention. At 6 months, the LLL was 63% lower in the mcPCB group compared to the POBA group (0.52 ± 1.2 vs 1.39 ± 1.1 mm; psup < 0.01). Binary restenosis occurred in 23% vs 52% of patients (p = 0.02). At 3 years, the prevalence of SAE was significantly lower in the mcPCB group (33.3 vs 63.3%; p = 0.02), which mainly resulted from a twofold reduction in target vessel revascularization rate (28.6 vs 59.3%; p = 0.02). The difference in mortality was nonsignificant (7.4 vs 14.3%; p = 0.42). Patients with mcPCB were less symptomatic and less likely to adhere to secondary prevention measures. In this pivotal trial, a novel mcPCB proved superior to POBA concerning LLL at 6-month follow-up, and SAE at 12 months. This result was sustained up to 3 years. There was no difference between groups regarding mortality. ClinicalTrials.gov Identifier: NCT02145065.


Subject(s)
Angioplasty, Balloon , Cardiovascular Agents , Peripheral Arterial Disease , Angioplasty, Balloon/adverse effects , Cardiovascular Agents/adverse effects , Coated Materials, Biocompatible , Femoral Artery/diagnostic imaging , Humans , Paclitaxel/adverse effects , Peripheral Arterial Disease/diagnostic imaging , Peripheral Arterial Disease/etiology , Peripheral Arterial Disease/therapy , Popliteal Artery/diagnostic imaging , Prospective Studies , Treatment Outcome , Vascular Patency
6.
Circ Cardiovasc Interv ; 13(6): e008737, 2020 06.
Article in English | MEDLINE | ID: mdl-32466676

ABSTRACT

BACKGROUND: Numerous randomized clinical trials have demonstrated the superiority of thin-strut biodegradable polymer second-generation drug-eluting stent to the first-generation drug-eluting stent and the noninferiority to the thin-strut second-generation permanent polymer drug-eluting stent. Data on long-term clinical outcomes with a novel ultrathin drug-eluting stent, to date, are limited. METHODS: The DESSOLVE III trial (Multicenter Randomized Study of the MiStent Sirolimus Eluting Absorbable Polymer Stent System for Revascularization of Coronary Arteries; n=1398) is a prospective, multicenter, single-blinded, all-comers, randomized controlled trial (NCT02385279), allocating in a 1:1 ratio to either ultrathin-strut biodegradable polymer MiStent sirolimus-eluting stent or to thin-strut permanent polymer Xience everolimus-eluting stent. The primary end point was device-oriented composite end point, defined as the composite of cardiac death, target vessel myocardial infarction, or clinically indicated target lesion revascularization. The secondary end point was patient-oriented composite end point, defined as the composite of all-cause mortality, any myocardial infarction, or any revascularization. RESULTS: At 3 years, follow-up data were available in 1381 patients (98.8%). The primary end point of device-oriented composite end point occurred in 10.5% for MiStent sirolimus-eluting stent and in 11.5% for Xience everolimus-eluting stent (P=0.55). Rates of cardiac death (3.9% versus 3.8%; P=0.88), target vessel myocardial infarction (3.2% versus 2.5%; P=0.43), and clinically indicated target lesion revascularization (5.2% versus 6.5%; P=0.30) did not differ significantly between the 2 devices. The rate of definite or probable stent thrombosis was infrequent and similar between the 2 arms (1.2% versus 1.5%; P=0.64). The 90-day landmark analysis showed no significant difference in device-oriented composite end point between the 2 groups after polymer degradation of MiStent. The risk of patient-oriented composite end point was comparable between the 2 groups (22.7% versus 22.9%; P=0.34). CONCLUSIONS: In the DESSOLVE III trial, early safety and efficacy with MiStent sirolimus-eluting bioabsorbable polymer-coated stent are confirmed at a longer term follow-up when compared with Xience everolimus-eluting permanent polymer-coated stent in a large all-comers population. Registration: URL: https://www.clinicaltrials.gov; Unique identifier: NCT02385279.


Subject(s)
Absorbable Implants , Cardiovascular Agents/administration & dosage , Drug-Eluting Stents , Everolimus/administration & dosage , Myocardial Ischemia/therapy , Percutaneous Coronary Intervention/instrumentation , Polymers/chemistry , Aged , Cardiovascular Agents/adverse effects , Europe , Everolimus/adverse effects , Female , Humans , Male , Middle Aged , Myocardial Ischemia/mortality , Percutaneous Coronary Intervention/adverse effects , Percutaneous Coronary Intervention/mortality , Prospective Studies , Prosthesis Design , Risk Factors , Single-Blind Method , Time Factors , Treatment Outcome
7.
Cardiol J ; 27(5): 600-607, 2020.
Article in English | MEDLINE | ID: mdl-30394507

ABSTRACT

BACKGROUND: The prevalence of diabetes has increased significantly in well-developed countries during the last decade and it continues to grow. Diabetes increases the risk of restenosis in patients treated percutaneously for peripheral artery disease. The present study sought to compare outcomes of atherectomy treatment in diabetic (DM) vs. non-diabetic (nDM) patients suffering from peripheral artery disease. METHOD: Between 2008 and 2012, 204 revascularization atherectomy procedures were performed on arteries of the lower extremities. The endpoints included target lesion revascularization (TLR), amputation and death. The type of atherectomy (excisional-soft plaque, orbital-calcified plaque, with active aspiration - with a thrombus) was left to operator discretion. RESULTS: This study contains 132 DM (66% male, age 68 ± 11.2 years) and 72 nDM (63% male, age 75 ± 11.3 years) subjects. DM were younger but had a higher prevalence of coronary artery disease (DM: 91% vs. nDM: 62%, p < 0.0001) and end-stage renal disease (DM: 22% vs. nDM: 2.5%, p < 0.0001). There were no differences in critical limb ischemia between the groups (DM: 21% vs. nDM: 12%, p = = 0.13). Mean time of follow-up was 384 and 411 days in DM and nDM, respectively (p = 0.43). There were no significant differences in TLR (DM: 15.2% vs. nDM: 22.2%, p = 0.249), amputations (DM: 3.0% vs. nDM: 1.5%, p = NS) or death rates (DM: 2.2% vs. nDM: 2.7%, p = NS). Kaplan-Mayer analysis showed no significant differences between the groups in the time to TLR, amputation or death. CONCLUSIONS: Plaque modification with adjusted atherectomy appears to have similar outcomes in diabetic as well as in non-diabetic patients. Nonetheless, a randomized study would be warranted to confirm the findings of the current study.


Subject(s)
Diabetes Mellitus , Peripheral Arterial Disease , Aged , Aged, 80 and over , Amputation, Surgical , Atherectomy , Coronary Artery Disease , Female , Humans , Male , Middle Aged , Retrospective Studies , Risk Factors , Treatment Outcome
8.
Postepy Kardiol Interwencyjnej ; 15(1): 81-90, 2019.
Article in English | MEDLINE | ID: mdl-31043989

ABSTRACT

INTRODUCTION: Preclinical, vascular response studies are limited due to lack of underlying disease. The available cholesterol-diet-based and genetic atherosclerotic models are not satisfactory due to long breeding, unpredictable lesion formation, low plaque volume and degree of stenosis. AIM: To evaluate the vascular response to local, intramural delivery of human, highly atherogenic lipids into healthy domestic swine (DS) coronary arteries. MATERIAL AND METHODS: A total of 24 coronary artery segments of 10 DS were enrolled. Following balloon injury (plain old balloon angioplasty - POBA), segments were assigned to local delivery of 2 ml of human LDL from apheresis (400 mg/dl, n = 9), 0.9% NaCl (control, n = 7) or to POBA alone. The solutions were infused with a modified, triple micro-needle catheter into the vessel wall. After 28 days, optical coherence tomography (OCT), virtual histology IVUS (VH-IVUS) and near-infra-red spectroscopy (NIRS) were performed. Following euthanasia, vessel segments were harvested for pathological evaluation. RESULTS: At 28 days the % area stenosis in OCT was highest in the LDL group (23.6 ±13 vs. 10.8 ±7 vs. 8.1 ±7%; p = 0.02). The presence of necrotic core (LDL: 55.5%, control: 37.5% and POBA: 42.8%; p = 0.77) and dense calcium (LDL: 33.3%, control: 28.5%, POBA: 37.5%; p = 0.94) in VH-IVUS were comparable between groups. The lipid core burden index in NIRS was negative in all cases. In pathology, the injury was comparable between groups (LDL: 1.6 ±0.4, control: 1.7 ±0.8, POBA: 1.7; p = 0.8) and specimens showed no signs of necrotic or lipid core. The tissue consisted of smooth muscle cells (SMC)/proteoglycan-rich lesions and inflammatory cells. CONCLUSIONS: Local delivery of saturated human LDL into the coronary artery wall was feasible and resulted in a higher degree of stenosis caused by intimal thickening. A discrepancy between histopathological findings and virtual histology intravascular ultrasound (VH-IVUS) was also noted.

9.
Lancet ; 391(10119): 431-440, 2018 02 03.
Article in English | MEDLINE | ID: mdl-29203070

ABSTRACT

BACKGROUND: MiStent is a drug-eluting stent with a fully absorbable polymer coating containing and embedding a microcrystalline form of sirolimus into the vessel wall. It was developed to overcome the limitation of current durable polymer drug-eluting stents eluting amorphous sirolimus. The clinical effect of MiStent sirolimus-eluting stent compared with a durable polymer drug-eluting stents has not been investigated in a large randomised trial in an all-comer population. METHODS: We did a randomised, single-blind, multicentre, phase 3 study (DESSOLVE III) at 20 hospitals in Germany, France, Netherlands, and Poland. Eligible participants were any patients aged at least 18 years who underwent percutaneous coronary intervention in a lesion and had a reference vessel diameter of 2·50-3·75 mm. We randomly assigned patients (1:1) to implantation of either a sirolimus-eluting bioresorbable polymer stent (MiStent) or an everolimus-eluting durable polymer stent (Xience). Randomisation was done by local investigators via web-based software with random blocks according to centre. The primary endpoint was a non-inferiority comparison of a device-oriented composite endpoint (DOCE)-cardiac death, target-vessel myocardial infarction, or clinically indicated target lesion revascularisation-between the groups at 12 months after the procedure assessed by intention-to-treat. A margin of 4·0% was defined for non-inferiority of the MiStent group compared with the Xience group. All participants were included in the safety analyses. This trial is registered with ClinicalTrials.gov, number NCT02385279. FINDINGS: Between March 20, and Dec 3, 2015, we randomly assigned 1398 patients with 2030 lesions; 703 patients with 1037 lesions were assigned to MiStent, of whom 697 received the index procedure, and 695 patients with 993 lesions were asssigned to Xience, of whom 690 received the index procedure. At 12 months, the primary endpoint had occurred in 40 patients (5·8%) in the sirolimus-eluting stent group and in 45 patients (6·5%) in the everolimus-eluting stent group (absolute difference -0·8% [95% CI -3·3 to 1·8], pnon-inferiority=0·0001). Procedural complications occurred in 12 patients (1·7%) in the sirolimus-eluting stent group and ten patients (1·4%) in the everolimus-eluting stent group; no clinical adverse events could be attributed to these dislodgements through a minimum of 12 months of follow-up. The rate of stent thrombosis, a safety indicator, did not differ between groups and was low in both treatment groups. INTERPRETATION: The sirolimus-eluting bioabsorbable polymer stent was non-inferior to the everolimus-eluting durable polymer stent for a device-oriented composite clinical endpoint at 12 months in an all-comer population. MiStent seems a reasonable alternative to other stents in clinical practice. FUNDING: The European Cardiovascular Research Institute, Micell Technologies (Durham, NC, USA), and Stentys (Paris, France).


Subject(s)
Acute Coronary Syndrome/drug therapy , Coronary Artery Disease/drug therapy , Drug-Eluting Stents , Everolimus/administration & dosage , Immunosuppressive Agents/administration & dosage , Sirolimus/administration & dosage , Absorbable Implants , Aged , Female , Humans , Male , Myocardial Infarction/etiology , Percutaneous Coronary Intervention , Prospective Studies , Single-Blind Method , Treatment Outcome
10.
Circulation ; 137(2): 166-183, 2018 01 09.
Article in English | MEDLINE | ID: mdl-29030347

ABSTRACT

BACKGROUND: Extracellular matrix (ECM) remodeling contributes to in-stent restenosis and thrombosis. Despite its important clinical implications, little is known about ECM changes post-stent implantation. METHODS: Bare-metal and drug-eluting stents were implanted in pig coronary arteries with an overstretch under optical coherence tomography guidance. Stented segments were harvested 1, 3, 7, 14, and 28 days post-stenting for proteomics analysis of the media and neointima. RESULTS: A total of 151 ECM and ECM-associated proteins were identified by mass spectrometry. After stent implantation, proteins involved in regulating calcification were upregulated in the neointima of drug-eluting stents. The earliest changes in the media were proteins involved in inflammation and thrombosis, followed by changes in regulatory ECM proteins. By day 28, basement membrane proteins were reduced in drug-eluting stents in comparison with bare-metal stents. In contrast, the large aggregating proteoglycan aggrecan was increased. Aggrecanases of the ADAMTS (a disintegrin and metalloproteinase with thrombospondin motifs) family contribute to the catabolism of vascular proteoglycans. An increase in ADAMTS-specific aggrecan fragments was accompanied by a notable shift from ADAMTS1 and ADAMTS5 to ADAMTS4 gene expression after stent implantation. Immunostaining in human stented coronary arteries confirmed the presence of aggrecan and aggrecan fragments, in particular, at the contacts of the stent struts with the artery. Further investigation of aggrecan presence in the human vasculature revealed that aggrecan and aggrecan cleavage were more abundant in human arteries than in human veins. In addition, aggrecan synthesis was induced on grafting a vein into the arterial circulation, suggesting an important role for aggrecan in vascular plasticity. Finally, lack of ADAMTS-5 activity in mice resulted in an accumulation of aggrecan and a dilation of the thoracic aorta, confirming that aggrecanase activity regulates aggrecan abundance in the arterial wall and contributes to vascular remodeling. CONCLUSIONS: Significant differences were identified by proteomics in the ECM of coronary arteries after bare-metal and drug-eluting stent implantation, most notably an upregulation of aggrecan, a major ECM component of cartilaginous tissues that confers resistance to compression. The accumulation of aggrecan coincided with a shift in ADAMTS gene expression. This study provides the first evidence implicating aggrecan and aggrecanases in the vascular injury response after stenting.


Subject(s)
ADAMTS Proteins/metabolism , Aggrecans , Coronary Vessels/surgery , Endopeptidases/metabolism , Extracellular Matrix/enzymology , Percutaneous Coronary Intervention/instrumentation , Proteomics/methods , Stents , Vascular Remodeling , ADAMTS Proteins/genetics , ADAMTS5 Protein/genetics , ADAMTS5 Protein/metabolism , Animals , Chromatography, High Pressure Liquid , Coronary Vessels/enzymology , Coronary Vessels/physiopathology , Drug-Eluting Stents , Endopeptidases/genetics , Female , Humans , Male , Metals , Mice, Knockout , Models, Animal , Neointima , Prosthesis Design , Signal Transduction , Sus scrofa , Tandem Mass Spectrometry , Time Factors
11.
Expert Rev Med Devices ; 14(4): 279-284, 2017 Apr.
Article in English | MEDLINE | ID: mdl-28394644

ABSTRACT

INTRODUCTION: Coronary bifurcation lesions pose therapeutic challenges during percutaneous coronary interventions. Dedicated bifurcation stents are one of the treatment options available. Areas covered: This paper will analyze the results of the BiOSS (Bifurcation Optimization Stent System) Clinical Program with a special focus on the cobalt-chromium BiOSS LIM C stent which has recently been introduced into the market (CE 2016). The stent is characterized by a strut thickness of 70 µm and sirolimus concentration of 1.4 µg/mm2. Expert commentary: The first BiOSS stent was a bare metal one, but shortly after a paclitaxel-eluting version was introduced - BiOSS Expert stent (CE 2010). After acceptable results of BiOSS Expert stent were obtained in registries as well as in the randomized clinical trial POLBOS I, a way for improvement was to change paclitaxel into the - olimus drug. Sirolimus has been chosen and the BiOSS LIM stent was developed (CE 2012). BiOSS LIM is the dedicated bifurcation stent with stainless steel platform that releases sirolimus (1.4 µg/mm2) from the surface of a biodegradable coating comprised of a copolymer of lactic and glycolic acids. Nevertheless, the idea to reduce late lumen loss and as a result TLR rate by using thinner struts appeared. Therefore, presently first-in-man registry with BiOSS LIM C is ongoing.


Subject(s)
Chromium , Cobalt , Coronary Disease/therapy , Drug-Eluting Stents , Coronary Disease/pathology , Equipment Design , Humans , Sirolimus/administration & dosage , Treatment Outcome
12.
Cardiovasc Ther ; 35(4)2017 Aug.
Article in English | MEDLINE | ID: mdl-28423237

ABSTRACT

AIMS: The optimal treatment strategy for coronary bifurcation lesions is still unknown. The aim of the study was to assess applicability of the new cobalt-chromium version of the sirolimus-eluting dedicated bifurcation BiOSS® LIM C stent in comparison with regular sirolimus-eluting Orsiro®  stent in a porcine coronary model. METHODS: A total of 13 BiOSS® LIM C stents and 6 Orsiro® stents were implanted in normal nonatherosclerotic porcine straight coronary arteries of six animals using 1.2:1.0 stent-to-artery ratio. Stent geometry and morphology were evaluated by Faxitron imaging. Vascular response was assessed by quantitative coronary angiography (QCA), optical coherence tomography (OCT), and histological analyses. RESULTS: OCT performed at 28 days confirmed that all stents were patent with no signs of thrombus. In morphometric analysis, no differences between groups regarding stent diameter (P=.141), neointima area (P=.247), % area stenosis (P=.293), or % diameter stenosis (P=.069) were observed. Also, no significant differences were noted between groups regarding their histopathology scores. The injury and inflammation scores were low (mean grade<1) in all groups. CONCLUSIONS: The novel BiOSS® LIM C stent demonstrates good short-term vascular effects in a porcine coronary bifurcation model which are comparable with Orsiro® stents.


Subject(s)
Coronary Vessels/physiology , Drug-Eluting Stents , Angiography , Animals , Chromium Alloys , Coronary Angiography , Coronary Vessels/drug effects , Heart/diagnostic imaging , Immunosuppressive Agents/administration & dosage , Immunosuppressive Agents/therapeutic use , Materials Testing , Myocardium/cytology , Sirolimus/administration & dosage , Sirolimus/therapeutic use , Swine , Tomography, Optical Coherence
13.
Circ J ; 81(3): 376-382, 2017 Feb 24.
Article in English | MEDLINE | ID: mdl-28090010

ABSTRACT

BACKGROUND: The impact of endovascular revascularization of the lower extremity arteries with atherectomy (AT) compared with percutaneous transluminal angioplasty (PTA) is still unclear. Therefore, the aim of the study was to compare long-term outcomes after percutaneous PTA and AT in patients requiring endovascular revascularization.Methods and Results:This was a single-center, retrospective registry of obstructive and symptomatic PAD patients who underwent endovascular revascularization. PTA was performed in 215 patients, and AT in 204 (Silver Hawk, EV3, n=125; CSI 360°, n=66; Pathway Medical Technologies, n=13). There were no significant between-group differences in baseline characteristics except for increased CAD, dialysis and CLI prevalence in the PTA group. Following propensity score analysis 131 well-matched pairs were included in analysis. Bail-out stenting was more frequent in the reference group (PTA, 6.1% vs. AT, 0%; P=0.004). At 6- and 12-month follow-up there were no differences in TLR between the groups (PTA, 8.3% vs. AT, 5.3%; P=0.47; and PTA, 16.7% vs. AT, 13.7%; P=0.73, respectively). The difference was in favor of AT at 24-month follow-up (PTA, 29.0% vs. AT, 16.7%; P=0.05). No difference was observed in amputation rate (PTA, 0.7% vs AT, 1.5%; P=0.62). On Kaplan-Meier analysis there were no significant differences between groups in time to TLR, amputation or death. CONCLUSIONS: AT was associated with lower risk of TLR, and this should be confirmed in randomized controlled trials.


Subject(s)
Angioplasty, Balloon , Atherectomy , Registries , Aged , Aged, 80 and over , Disease-Free Survival , Female , Follow-Up Studies , Humans , Lower Extremity/surgery , Male , Middle Aged , Retrospective Studies , Survival Rate
14.
JACC Cardiovasc Interv ; 9(4): 318-327, 2016 Feb 22.
Article in English | MEDLINE | ID: mdl-26892080

ABSTRACT

OBJECTIVES: This study has reported 10-year clinical follow-up of patients enrolled in the prospective, randomized LE MANS (Left Main Stenting) trial. BACKGROUND: The very long-term outcome after left main stenting in comparison with surgical revascularization remains unknown. METHODS: In this prospective, multicenter trial, we randomly assigned 105 patients with unprotected left main coronary artery stenosis with low and medium complexity of coexisting coronary artery disease according to SYNTAX (Synergy Between Percutaneous Coronary Intervention With Taxus and Cardiac Surgery) score to percutaneous coronary intervention (PCI) with stenting (n = 52) or coronary artery bypass grafting (CABG) (n = 53). Drug-eluting stents were implanted in 35%, whereas arterial grafts to the left anterior descending artery were utilized in 81%. Currently, the mean long-term follow-up was collected at 9.8 ± 1.0 years. Follow up for all-cause mortality is complete, whereas the incidence of major adverse cardiovascular and cerebral events (MACCE) was reported from 90% of patients. Ambulatory follow-up was completed in 46 (43.9%) patients. RESULTS: At 10 years, there was a trend toward higher ejection fraction in stenting when compared with surgery (54.9 ± 8.3% vs. 49.8 ± 10.3%; p = 0.07). The mortality (21.6% vs. 30.2%; p = 0.41) and MACCE (51.1% vs. 64.4%; p = 0.28) were statistically not different between groups; however, numerically the difference was in favor of stenting. Similarly, there was no difference in the occurrence of myocardial infarction (8.7 vs. 10.4%; p = 0.62), stroke (4.3 vs. 6.3%; p = 0.68), and repeated revascularization rates (26.1% vs. 31.3%; p = 0.64). The probability of very long-term survival up to 14 years was comparable between PCI and CABG (74.2% vs. 67.5%; p = 0.34; hazard ratio: 1.45, 95% confidence interval: 0.67 to 3.13); however, there was a trend toward higher MACCE-free survival in the PCI group (34.7% vs. 22.1%; p = 0.06; hazard ratio: 1.71, 95% confidence interval: 0.97 to 2.99). CONCLUSIONS: In patients with unprotected left main coronary artery stenosis with low and medium complexity of coexisting coronary artery disease, stenting offers numerically, but statistically nonsignificant, favorable long-term outcome up to 10 years in terms of safety and efficacy outcome measures, therefore, constitutes an alternative therapy for CABG.


Subject(s)
Coronary Artery Bypass , Coronary Artery Disease/therapy , Coronary Stenosis/therapy , Percutaneous Coronary Intervention/instrumentation , Stents , Aged , Cardiovascular Agents/therapeutic use , Coronary Angiography , Coronary Artery Bypass/adverse effects , Coronary Artery Bypass/mortality , Coronary Artery Disease/diagnostic imaging , Coronary Artery Disease/mortality , Coronary Artery Disease/physiopathology , Coronary Stenosis/diagnostic imaging , Coronary Stenosis/mortality , Coronary Stenosis/physiopathology , Disease-Free Survival , Drug-Eluting Stents , Female , Humans , Male , Middle Aged , Myocardial Infarction/etiology , Percutaneous Coronary Intervention/adverse effects , Percutaneous Coronary Intervention/mortality , Risk Factors , Stroke/etiology , Stroke Volume , Time Factors , Treatment Outcome , Ventricular Function, Left
15.
J Interv Cardiol ; 28(4): 348-57, 2015 Aug.
Article in English | MEDLINE | ID: mdl-26224390

ABSTRACT

BACKGROUND: There is limited comparative clinical data regarding the safety and efficacy profile of paclitaxel delivery via balloon versus stent-polymer matrix. In this study, we aimed to compare the clinical and angiographic results of two different methods of paclitaxel delivery among patients undergoing percutaneous coronary intervention (PCI) for single de novo coronary lesions. METHODS: A total of 202 patients undergoing PCI due to symptomatic heart disease and at least one significant coronary artery lesion were prospectively enrolled in a multicenter non-inferiority trial. Eligible patients were randomized to a revascularization with either a paclitaxel eluting stent (PES = Coroflex Please, B.Braun) or a bare metal stent (BMS) followed by a paclitaxel coated balloon (PCB) dilation (BMS = Coroflex + PCB = Sequent Please, B.Braun). Clinical follow-up was obtained at 9 months in all patients, whereas angiographic in a subset of 94 (46.5%) patients. RESULTS: The baseline characteristics were well balanced between groups. At 9 months, the primary endpoint of in-stent late lumen loss in BMS + PCB was comparable and non-inferior to PES (0.21 ± 0.5 vs. 0.30 ± 0.7 mm, respectively. P(non-inf) < 0.05). At 9 months, the incidence of MACE (7.0 vs. 6.9%, HR = 1, 95%CI: 0.3-2.8; P = 0.99), comprising the occurrence of myocardial infarction (4.9 vs. 3.0%, HR = 1.62, 95%CI: 0.4-6.5; P = 0.32), target lesion revascularization (6.9 vs. 5.0%, HR = 1.42, 95%CI: 0.4-4.4; p = 0.54) and stent thrombosis (4.9 vs. 3.0%, HR = 2.01, 95%CI: 0.5-7.4; P = 0.74) was comparable between BMS + PCB and PES, respectively. In the BMS + PCB group, thrombosis tended to occur within 30 days (3.9 vs. 1.0%; P = 0.38). CONCLUSIONS: Paclitaxel delivery via drug coated balloon or polymer-stent matrix achieved comparable angiographic and clinical results among patients with de novo coronary lesions. BMS + PCB revascularization was associated with a higher rate of stent thrombosis when compared to newer generation drug eluting stents, therefore, should be recommended as a bail-out for PCB alone angioplasty.


Subject(s)
Angioplasty, Balloon, Coronary/instrumentation , Coronary Stenosis/therapy , Paclitaxel/administration & dosage , Stents , Coronary Angiography , Coronary Artery Disease/therapy , Coronary Thrombosis/epidemiology , Female , Humans , Male , Middle Aged , Myocardial Infarction/epidemiology , Prospective Studies