1.
J Virol Methods ; 183(1): 8-13, 2012 Jul.
Article in English | MEDLINE | ID: mdl-22425698

ABSTRACT

This report describes the development and pre-clinical testing of a new, random-access RNA sample preparation system (TruTip) for nasopharyngeal samples. The system is based on a monolithic, porous nucleic acid binding matrix embedded within an aerosol-resistant pipette tip and can be operated with single or multi-channel pipettors. Equivalent extraction efficiencies were obtained between automated QIAcube and manual TruTip methods at 10(6) gene copies influenza A per mL nasopharyngeal aspirate. Influenza A and B amended into nasopharyngeal swabs (in viral transport medium) were detected by real-time RT-PCR at approximately 745 and 370 gene copies per extraction, respectively. RNA extraction efficiency in nasopharyngeal swabs was also comparable to that obtained on an automated QIAcube instrument over a range of input concentrations; the correlation between threshold cycles (or nucleic acid recovery) for TruTip and QIAcube-purified RNA was R(2)>0.99. Preclinical testing of TruTip on blinded nasopharyngeal swab samples resulted in 98% detection accuracy relative to a clinically validated easyMAG extraction method. The physical properties of the TruTip binding matrix and ability to customize its shape and dimensions likewise make it amenable to automation and/or fluidic integration.


Subject(s)
Influenza A virus/genetics , Influenza B virus/genetics , Nasopharynx/virology , RNA, Viral/isolation & purification , Specimen Handling/methods , Virology/methods , Animals , Humans
SELECTION OF CITATIONS
SEARCH DETAIL
...