J Arthroplasty ; 37(6): 1130-1135, 2022 06.
Article in English | MEDLINE | ID: mdl-35131388

ABSTRACT

BACKGROUND: A uniquely designed, non-heat-treated moderately cross-linked acetabular polyethylene liner used in total hip arthroplasty (THA) demonstrated excessive wear during routine follow-up, prompting an evaluation of the linear wear rate. METHODS: All THAs were performed by the senior author. The study group included 38 THAs using the uniquely designed polyethylene in question, compared to a control group of 21 THAs using another moderately cross-linked polyethylene with good 10-year outcomes. Two-dimensional linear head penetration wear measurements were obtained using the Martell Hip Analysis Suite, and retrieval analysis was performed on two liners. RESULTS: The study group had a significantly higher average penetration rate of 0.089 mm/y than the control group average rate of 0.047 mm/y (P = .04). Forty-five percent of the study group had a wear rate above the osteolysis threshold (0.1 mm/y), compared to 24% in the control group. Macroscopic analysis of two retrieved liners validated the radiographic findings. CONCLUSION: The data suggest unexpectedly higher wear rates for a moderately cross-linked polyethylene design, with nearly half of the study group at risk for osteolysis. Further registry or database analyses of this particular moderately cross-linked polyethylene are warranted.


Subject(s)
Arthroplasty, Replacement, Hip , Hip Prosthesis , Osteolysis , Follow-Up Studies , Humans , Polyethylene , Prosthesis Design , Prosthesis Failure
SELECTION OF CITATIONS
SEARCH DETAIL