2.
Pacing Clin Electrophysiol ; 47(3): 433-436, 2024 03.
Article in English | MEDLINE | ID: mdl-37221929

ABSTRACT

An 88-year-old Japanese woman underwent DDD pacemaker (MicroPort KORA 250 DR, V lead: VEGA R52) implantation for complete atrioventricular block. A 12-lead electrocardiogram for a routine examination showed atrial pacing within the intrinsic P wave, followed by inhibition of ventricular pacing. Pacemaker interrogation revealed no abnormalities in the basic parameters; however, ventricular pacing was inhibited by far-field sensing of intrinsic atrial waves before atrial events; type II far-field P-wave sensing. As a result, unusual atrial pacing occurred due to the pause suppression algorithm, which is the one of the functions that prevent atrial fibrillation development.


Subject(s)
Atrioventricular Block , Pacemaker, Artificial , Female , Humans , Aged, 80 and over , Atrioventricular Block/diagnosis , Atrioventricular Block/therapy , Cardiac Pacing, Artificial , Heart Atria , Heart Ventricles
3.
J Arrhythm ; 39(6): 956-959, 2023 Dec.
Article in English | MEDLINE | ID: mdl-38045463

ABSTRACT

We presented a detailed time course of RFCA-associated PV calcification process. Because RFCA-associated PV calcification may progress over time even with a single ablation, long-term caution should be paid in cases of especially extensive ablation and/or patients with stiff LA syndrome.

4.
J Cardiol Cases ; 27(5): 212-214, 2023 May.
Article in English | MEDLINE | ID: mdl-37180215

ABSTRACT

Belt-type ambulatory electrocardiograph (EV-201) is an arrhythmia diagnostic device that can record an electrocardiogram (ECG) for a maximum of 2 weeks. Here, we report the novel utility of EV-201 in detecting arrhythmias in two professional athletes. Treadmill exercise test and Holter ECG failed to detect arrhythmia because of insufficient exercise, electrocardiogram noise. However, by wearing EV-201 only during a marathon run, supraventricular tachycardia onset and termination were successfully detected. Throughout, both athletes were diagnosed with fast-slow atrioventricular nodal re-entrant tachycardia. Therefore, EV-201 enables long-term belt-type recording, thereby being useful in detecting tachyarrhythmias that occur infrequently and during strenuous exercises. Learning objective: Diagnosis of arrhythmias during high-intensity exercise in athletes by conventional electrocardiography is sometimes difficult due to inducibility and frequency of arrythmias or motion artifacts. The primary finding of this report is that EV-201 is useful in diagnosing such arrhythmias. The secondary finding is that fast-slow atrioventricular nodal re-entrant tachycardia is a common occurrence in arrhythmias among athletes.

5.
J Arrhythm ; 38(6): 1049-1055, 2022 Dec.
Article in English | MEDLINE | ID: mdl-36524042

ABSTRACT

Background: The effect of cardiac rehabilitation (CR) on patients undergoing device implantation (DI) for arrhythmias has been reported; however, the implementation status of these patients has not been clarified. This study aimed to verify the implementation status of CR for patients with heart disease who have undergone DI using real-world data. Methods: This was an observational study using a nationwide administrative database associated with the diagnosis procedure combination (DPC) system in Japan (2014-2018). Subjects were patients with heart disease (70 667 cases) who underwent DI during the above scheduled hospitalization period. The overall rate of CR and the background factors of the subjects were verified. Results: The CR rate for patients with heart disease who underwent DI during hospitalization was 23%, and the CR rate for patients with comorbid heart failure who underwent DI was only 32%. It was confirmed that progressing age was associated with a higher CR implementation rate. The lower the Barthel index score at the time of admission, the higher the CR implementation rate. Conclusions: CR was performed for only one-quarter of all the patients during admission for DI and just one-third of the patients for DI with heart failure. Most of these patients were elderly and had a decreased ability to perform activities of daily living. The DPC data are subject to various limitations, and further research is necessary.

6.
BMC Health Serv Res ; 22(1): 1229, 2022 Oct 04.
Article in English | MEDLINE | ID: mdl-36192749

ABSTRACT

BACKGROUND: The number of employees with physical diseases is increasing, and there is a need for support to help them return and continue to work. To provide effective support, it is important to identify barriers and facilitators for individuals in returning and continuing to work. Previous studies have reported barriers and facilitators for specific diseases. However, few reports have dealt with these issues across various diseases. To identify a range of barriers and facilitators that may apply to different physical diseases, we conducted a qualitative analysis by interviewing patients with diverse characteristics being treated for diseases. METHODS: We conducted semi-structured interviews based on the criteria for qualitative research. We investigated three disease groups to obtain details of barriers and facilitators: impairments that were visible to other people (mainly stroke); impairments invisible to others (mainly heart disease); and impairments that changed over time (mainly cancer). Interview transcripts were analyzed and the results reported using systematic text condensation. RESULTS: We extracted 769 meaning units from 28 patient interviews. We categorized barriers and facilitators that were generalizable to various diseases into three themes (personal factors, workplace factors, and inter-sectoral collaboration and social resources) and 10 sub-themes (work ability, psychological impacts, health literacy, social status, family background, workplace structure, workplace system, workplace support, inter-sectoral collaboration, and social resources). CONCLUSIONS: This study identified 10 sub-themes that can be applied for workers with physical diseases; those sub-themes may be used as a basis for communicating with those individuals about returning and continuing to work. Our results suggest that various barriers and facilitators for workers with physical diseases should be understood and addressed at medical institutions, workplaces, and support sites.


Subject(s)
Stroke , Workplace , Disabled Persons , Humans , Qualitative Research , Return to Work
8.
J Cardiol Cases ; 26(3): 208-211, 2022 Sep.
Article in English | MEDLINE | ID: mdl-36091607

ABSTRACT

A three-dimensional (3D) mapping system is safe and effective for catheter ablation of incisional atrial tachycardia (IAT). Prolongation of the PR interval is occasionally observed after this procedure. Although a first-degree atrioventricular block is typically benign, an excessively prolonged PR interval can indicate a worse prognosis. Currently, a method to predict the PR interval after ablation therapy for IAT is lacking. We report the case of a 70-year-old woman with paroxysmal atrial tachycardia, in which our maneuver using a 3D mapping system and the electrophysiological findings enabled us to preoperatively predict the post-ablation PR interval. We believe that this method is useful for determining treatment strategies for IAT. Learning objective: Three-dimensional (3D) mapping systems can clearly visualize macro-reentrant circuits and enable the creation of precise ablation lines. When creating ablation lines for incisional atrial tachycardia, attention should be paid to the prolongation of the PR interval, because an excessively prolonged PR interval may indicate a worse prognosis. Herein, we have presented a method for predicting the post-ablation PR interval using a 3D mapping system and electrophysiological findings.

9.
J Arrhythm ; 38(3): 482-485, 2022 Jun.
Article in English | MEDLINE | ID: mdl-35785396

ABSTRACT

We reported a case of early S-ICD shock due to the carryover of previously charged energy. Depending on the interval with the preceding event, an early shock may occur for the subsequent event. Especially, in cases where non-sustained VTs occur frequently, the indication for S-ICD surgery should be considered carefully.

10.
Int Heart J ; 63(3): 498-503, 2022.
Article in English | MEDLINE | ID: mdl-35650151

ABSTRACT

In terms of the pulmonary vein (PV), atrial fibrillation (AF) patients have a shorter effective refractory period (ERP) than those without AF and a large dispersion of the ERP. Although the frequency of AF from the superior vena cava (SVC) was the highest among non-PV foci, the characteristics of the ERP in the SVC (SVC-ERP) were unclear. The purpose of this study was to elucidate the relationship between SVC-ERP and the inducibility of AF after PV isolation (PVI).Consecutive 28 patients who underwent PVI were included. After successful PVI, the SVC-ERP was measured at three positions in SVC. Rapid electrical stimuli were delivered at the shortest SVC-ERP to induce AF. Patients in whom AF was induced were assigned to the SVC-induced group (SIG), and the remaining patients were the non-SVC-induced group (non-SIG). The size of the SVC sleeve was evaluated via three-dimensional electroanatomic mapping.The SIG had a significantly shorter average SVC-ERP (236.0 ± 25.2 versus 294.8 ± 36.8 ms, P < 0.001), whereas SVC-ERP dispersion was not significantly different (30.0 ± 25.4 versus 33.3 ± 20.1 ms, P = 0.56). Although the longer SVC diameter was significantly longer in the SIG (27.4 ± 4.3 versus 22.9 ± 4.6 mm, P = 0.03), the SVC-ERP was significantly associated with pacing inducibility of AF after adjustment for the longer SVC diameter (odds ratio: 0.96 [1 ms increments], P = 0.01).The SIG had a shorter SVC-ERP, whereas the dispersion was not significantly different between the two groups. The SVC-ERP can be one of the mechanisms of arrhythmogenicity for AF originating from the SVC.


Subject(s)
Atrial Fibrillation , Catheter Ablation , Pulmonary Veins , Atrial Fibrillation/surgery , Catheter Ablation/adverse effects , Catheter Ablation/methods , Humans , Odds Ratio , Pulmonary Veins/surgery , Vena Cava, Superior/surgery
11.
J Cardiol ; 80(4): 332-338, 2022 10.
Article in English | MEDLINE | ID: mdl-35637122

ABSTRACT

BACKGROUND: The prognosis associated with right ventricular (RV) free-wall pacing is worse than that of septal pacing. Identification of the pacing site using a 12­lead electrocardiogram (ECG) is controversial and may be influenced by ECG changes within the same septal or free-wall area. The relationship between the diagnostic capabilities of ECG and pacing sites has previously been qualitatively evaluated. However, in this study, this relationship was analyzed quantitatively, and accurate evaluation of the pacing site was determined using computed tomography (CT). METHODS: Of 779 consecutive outpatients, 65 who underwent pacemaker implantation and thoracic CT were prospectively included and classified into the following groups according to the lead tip position: free-wall, septal, or septum/free-wall boundary (hinge) group. The hinge was used as an anatomical marker, and the distance from the hinge to the lead tip was measured. Under RV pacing, a 12­lead ECG was obtained. ECG findings were evaluated using three criteria (including lead I, II, and aVL and precordial leads V5 and V6) previously reported to be useful in differentiating pacing sites. RESULTS: The lead tips were anchored at the free-wall in 10 patients, the septal wall in 19 patients, and the hinge in 32 patients. Paced QRS duration correlated with the distance from the hinge to the lead tip for the free-wall and septum (r = 0.47 and - 0.68, respectively). Estimation of the lead tip implantation site using the ECG algorithm was useful; however, the algorithm's accuracy decreased around the hinge. CONCLUSIONS: ECG is useful in differentiating pacing sites but is less accurate around the hinge, which may be the reason that the identification of the RV pacing site using ECG remains controversial.


Subject(s)
Pacemaker, Artificial , Ventricular Septum , Algorithms , Cardiac Pacing, Artificial/methods , Electrocardiography/methods , Heart Ventricles/diagnostic imaging , Humans , Tomography, X-Ray Computed/methods
12.
Pacing Clin Electrophysiol ; 45(3): 297-301, 2022 Mar.
Article in English | MEDLINE | ID: mdl-35172014

ABSTRACT

BACKGROUND: The displacement of cardiac implantable electronic devices (CIEDs) toward the caudal side during standing after CIED implantation could cause lead dislodgement. This study investigated the relationship between supine pocket position and standing CIEDs' displacement distance after the implantation. METHODS: After CIED surgeries performed at 2 hospitals between 2012 and 2020, 134 patients underwent postoperative chest x-rays in the supine and standing positions during hospitalization. To measure the displacement distance of CIEDs from the supine to the standing position, we identified the first thoracic vertebrae (Th1) in the supine position using the first rib as an index, drew a horizontal line at the lower edge of the Th1, and calculated the distance from that point to the upper edge of the CIED. The difference between measures for the two positions was compared. At the position of the pocket in the thorax in the supine position, the ratio of the distance between the thorax and the device is defined as the device thorax ratio (DTR). We examined the relationship between DTR and CIED displacement distance. RESULTS: In this study, we included 134 patients (53% men; median age, 79 years, body mass index, 22.3 ± 3.4; pacemaker 93%, left implantation 96%). We found that the more lateral the position of the CIED pocket, the more the CIED fell when standing (confidence interval = 0.34-0.60, P < .001). CONCLUSIONS: The farther the CIED was implanted outside the thorax in the supine position, the more significantly the CIED was displaced in the standing position.


Subject(s)
Defibrillators, Implantable , Pacemaker, Artificial , Aged , Female , Humans , Male , Retrospective Studies , Standing Position , Thorax
13.
Pacing Clin Electrophysiol ; 45(3): 435-438, 2022 Mar.
Article in English | MEDLINE | ID: mdl-34793604

ABSTRACT

A 74-year-old man experienced complete atrioventricular (AV) block 2 days after catheter ablation for right atrial (RA) macroreentrant tachycardia. We performed DDD pacemaker implantation with atrial septal pacing because other sites of pacing threshold were not acceptable. The maximum left ventricular outflow tract velocity time integral was 15.8 cm with sensed AV delay (40 ms) and 15.0 cm with paced AV delay (220 ms); however, this exceeded the pacemaker's maximum difference of 100 ms. We herein report the case of a large discrepancy in optimal AV delay intervals between sensed and paced atrial events, requiring consideration of proper pacemaker settings.


Subject(s)
Atrioventricular Block , Pacemaker, Artificial , Aged , Atrioventricular Block/therapy , Cardiac Pacing, Artificial , Heart Atria , Humans , Male
14.
Pacing Clin Electrophysiol ; 45(3): 431-434, 2022 Mar.
Article in English | MEDLINE | ID: mdl-34752651

ABSTRACT

An 84-year-old woman with type B Wolff-Parkinson-White (WPW) with Ebstein anomaly was admitted with heart failure. She had rapid wide QRS tachycardia due to accessory pathway (AP) conduction associated with atrial fibrillation (AF). Since transesophageal echocardiography before catheter ablation showed a left atrial thrombus, ablation was performed using a 3D mapping system under AF. After marking the functional tricuspid anulus with intra-cardiac echocardiography, 3D intra-cardiac electrogram visualization (ripple map) during AF enabled clear identification of location of the AP. After ablation, there was no complication of cerebral infarction, and the heart failure improved.


Subject(s)
Accessory Atrioventricular Bundle , Atrial Fibrillation , Catheter Ablation , Ebstein Anomaly , Heart Failure , Wolff-Parkinson-White Syndrome , Accessory Atrioventricular Bundle/complications , Accessory Atrioventricular Bundle/surgery , Aged, 80 and over , Catheter Ablation/adverse effects , Ebstein Anomaly/complications , Ebstein Anomaly/surgery , Electrocardiography , Female , Heart Failure/complications , Humans , Wolff-Parkinson-White Syndrome/complications , Wolff-Parkinson-White Syndrome/surgery
15.
J Cardiol Cases ; 24(6): 303-306, 2021 Dec.
Article in English | MEDLINE | ID: mdl-34917216

ABSTRACT

An 83-year-old man with no structural heart disease underwent pulmonary vein isolation (PVI) for symptomatic paroxysmal atrial fibrillation (AF). The PVI was successfully performed by cryoballoon ablation with a single transseptal puncture. A 12Fr deflectable sheath and an 8.5Fr long sheath crossed the interatrial septum via the same puncture site. Five months after PVI, the patient was readmitted because of heart failure and recurrence of AF. The echocardiogram showed a large (10.7 × 5.8 mm) iatrogenic atrial septal defect (IASD) at the previous puncture site. Both right-to-left and left-to-right shunts were observed during systole and diastole, respectively. Despite the initiation of optimal medical therapy for heart failure, symptoms were not completely controlled and IASD remained 11 months after PVI. Eventually, he received multiple additional PVI for recurrence of AF and percutaneous transcatheter closure (13 mm disc for 10.9 × 8.9 mm- IASD), then heart failure was controlled with the improvement of the right atrial and ventricular size. Although the induction of heart failure by IASD after PVI is rare, it should be noted as one of the causes, especially after cryoballoon ablation with simultaneous mapping catheter insertion via a single transseptal puncture. .

16.
BMJ Open ; 11(12): e047932, 2021 12 06.
Article in English | MEDLINE | ID: mdl-34872993

ABSTRACT

OBJECTIVES: The prevention and improvement of the prognosis of out-of-hospital cardiac arrests (OHCAs) are important issues especially with respect to their social and economic significance in working populations. The age distribution of the working population in Japan is expected to change continually due to its ageing society and extension of retirement; however, few reports have examined the long-term condition of OHCA in the working population, defined by age. The aim of this study was to determine the incidence of OHCAs and the survival rates after 1 month, among the Japanese working population, defined by age, considering the changing age distribution. DESIGN AND SETTING: We analysed the All-Japan Utstein registry, a prospective, nationwide, population-based, observational registry (2005-2016). PARTICIPANTS: From the registry, 212 961 patients with OHCA from the Japanese working population (defined aged 20-69 years), with only cardiogenic aetiology participated in this study. These patients were further divided into four groups according to the type of citizen bystander (family, friends, work-colleagues and passers-by). PRIMARY AND SECONDARY OUTCOME MEASURES: The main outcomes were 1-month survival with favourable neurological outcomes. RESULTS: The incidence of OHCAs, in any age group, was almost constant during the 12-year period. The work-colleagues had the best prognosis despite having significantly longer times to initial defibrillations compared with the passers-by (13 vs 12 min, respectively, p<0.001) that was associated independently with 1-month survival with favourable neurological outcomes (adjusted OR: 0.94 (1 min increments), p<0.001). CONCLUSIONS: In the 12-year period, the incidence of OHCAs in any age group remained almost constant, whereas the prognosis improved each year. Reducing the time to initial defibrillation may further improve the prognosis of OHCAs with a work-colleague bystander.


Subject(s)
Cardiopulmonary Resuscitation , Emergency Medical Services , Out-of-Hospital Cardiac Arrest , Adult , Aged , Cardiopulmonary Resuscitation/methods , Cohort Studies , Humans , Incidence , Japan/epidemiology , Middle Aged , Prospective Studies , Registries , Survival Rate , Young Adult
17.
J Arrhythm ; 37(4): 1052-1060, 2021 Aug.
Article in English | MEDLINE | ID: mdl-34386132

ABSTRACT

BACKGROUND: An implantable cardioverter defibrillator (ICD) is the most reliable therapeutic device for preventing sudden cardiac death in patients with sustained ventricular tachycardia (VT). Regarding its effectiveness, targeted VT is defined based on the tachyarrhythmia cycle length. However, variations in RR interval variability of VTs may occur. Few studies have reported on VT characteristics and effects of ICD therapy according to the RR interval variability. We aimed to identify the clinical characteristics of VTs and ICD therapy effects according to the RR interval variability. METHODS: We analyzed 821 VT episodes in 69 patients with ICDs or cardiac resynchronization therapy defibrillators. VTs were classified as irregular when the difference between two successive beats was >20 ms in at least one of 10 RR intervals; otherwise, they were classified as regular. We evaluated successful termination using anti-tachycardia pacing (ATP)/shock therapy, spontaneous termination, and acceleration between regular and irregular VTs. The RR interval variability reproducibility rates were evaluated. RESULTS: Regular VT was significantly more successfully terminated than irregular VT by ATP. No significant difference was found in shock therapy or VT acceleration between the regular and irregular VTs. Spontaneous termination occurred significantly more often in irregular than in regular VT cases. The reproducibility rates of RR interval variability in each episode and in all episodes were 89% and 73%, respectively. CONCLUSIONS: ATP therapy showed greater effectiveness for regular than for irregular VT. Spontaneous termination was more common in irregular than in regular VT. RR interval variability of VTs seems to be reproducible.

18.
J Cardiovasc Electrophysiol ; 32(11): 3103-3106, 2021 11.
Article in English | MEDLINE | ID: mdl-34460986

ABSTRACT

Adenosine can hyperpolarize the atrial action potential, which helps rapidly re-establish the membrane potential in ablated sites and unmask "dormant conduction." It has been reported that pharmacological agents, including adenosine, were unable to revive traumatized tissues. We present the first case of the catheter-induced mechanical block ("bump" phenomenon) that was unmasked with adenosine administration in the working myocardium of the superior vena cava. This result may be because, unlike before, we could determine the force of contact between the tip of the ablation catheter and the myocardial tissue. This case suggests the clinical usefulness of adenosine for unmasking bumped sites.


Subject(s)
Atrial Fibrillation , Catheter Ablation , Pulmonary Veins , Adenosine , Atrial Fibrillation/diagnosis , Atrial Fibrillation/surgery , Catheter Ablation/adverse effects , Catheters , Humans , Pulmonary Veins/surgery , Treatment Outcome , Vena Cava, Superior
19.
Physiol Rep ; 8(17): e14558, 2020 09.
Article in English | MEDLINE | ID: mdl-32914562

ABSTRACT

We generated a transgenic rat line that expresses oxytocin (OXT)-monomeric red fluorescent protein 1 (mRFP1) fusion gene to visualize the dynamics of OXT. In this transgenic rat line, hypothalamic OXT can be assessed under diverse physiological and pathophysiological conditions by semiquantitative fluorometry of mRFP1 fluorescence intensity as a surrogate marker for endogenous OXT. Using this transgenic rat line, we identified the changes in hypothalamic OXT synthesis under various physiological conditions. However, few reports have directly examined hypothalamic OXT synthesis under hyperosmolality or hypovolemia. In this study, hypothalamic OXT synthesis was investigated using the transgenic rat line after acute osmotic challenge and acute hypovolemia induced by intraperitoneal (i.p.) administration of 3% hypertonic saline (HTN) and polyethylene glycol (PEG), respectively. The mRFP1 fluorescence intensity in the paraventricular (PVN) and supraoptic nuclei (SON) was significantly increased after i.p. administration of HTN and PEG, along with robust Fos-like immunoreactivity (co-expression). Fos expression showed neuronal activation in the brain regions that are associated with the hypothalamus and/or are involved in maintaining water and electrolyte homeostasis in HTN- and PEG-treated rats. OXT and mRFP1 gene expressions were dramatically increased after HTN and PEG administration. The plasma OXT level was extremely increased after HTN and PEG administration. Acute osmotic challenge and acute hypovolemia induced upregulation of hypothalamic OXT in the PVN and SON. These results suggest that not only endogenous arginine vasopressin (AVP) but also endogenous OXT has a key role in maintaining body fluid homeostasis to cope with hyperosmolality and hypovolemia.


Subject(s)
Hypothalamus/metabolism , Hypovolemia/metabolism , Osmotic Pressure , Oxytocin/genetics , Animals , Hypovolemia/physiopathology , Luminescent Proteins/genetics , Luminescent Proteins/metabolism , Male , Osmoregulation , Oxytocin/metabolism , Rats , Transgenes , Up-Regulation , Red Fluorescent Protein
20.
J Cardiovasc Electrophysiol ; 31(10): 2765-2769, 2020 10.
Article in English | MEDLINE | ID: mdl-32757423

ABSTRACT

An 82-year-old woman received pacemaker implantation for sick sinus syndrome. Two days after the implantation, electrocardiography showed 2:1 atrial pacing failure, followed by a bradycardia-dependent increase in the atrial pacing threshold during a pacemaker examination. However, transient 1:1 atrial pacing capture recovered by adenosine triphosphate (ATP) administration, which was performed to evaluate the bradycardia-dependent pacing failure mechanism. We considered this phenomenon to be caused by Phase 4 depolarization and avoided replacing this atrial lead. Three weeks later, the atrial pacing threshold had improved. We report the potential role of Phase 4 depolarization in a bradycardia-dependent increase in pacing threshold by using ATP.


Subject(s)
Atrial Fibrillation , Pacemaker, Artificial , Adenosine , Aged, 80 and over , Atrial Fibrillation/diagnosis , Atrial Fibrillation/therapy , Bradycardia/chemically induced , Bradycardia/diagnosis , Bradycardia/therapy , Cardiac Pacing, Artificial/adverse effects , Electrocardiography , Female , Humans , Pacemaker, Artificial/adverse effects , Sick Sinus Syndrome/diagnosis , Sick Sinus Syndrome/therapy