Liquid chromatography (LC) coupled with mass spectrometry (MS) detection is one of the most powerful analytical tools for organic compound analysis. The advantages of using LC/MS methods over HPLC methods include: selectivity, chromatographic integrity, peak assignment, structural information, and rapid method development. In this paper, a new liquid chromatography-tandem mass spectrometry (LC-MS/MS) method has been developed and validated for the determination of bisoprolol in human plasma samples, using metoprolol as internal standard and liquid-liquid extraction procedure. The assay has proven to be sensitive, specific and reproducible, suitable to determine the bisoprolol concentration, following a single oral administration of a 10 mg bisoprolol tablet in 22 healthy volunteers, in the bioequivalence study of Bisoprolol 10 mg coated tablets, produced by Antibiotice S.A. versus Concor 10 mg, produced by Merck.