1.
Lab Chip ; 24(4): 955-965, 2024 02 13.
Article in English | MEDLINE | ID: mdl-38275173

ABSTRACT

Microfluidic artificial lungs (µALs) are a new class of membrane oxygenators. Compared to traditional hollow-fiber oxygenators, µALs closely mimic the alveolar microenvironment due to their size-scale and promise improved gas exchange efficiency, hemocompatibility, biomimetic blood flow networks, and physiologically relevant blood vessel pressures and shear stresses. Clinical translation of µALs has been stalled by restrictive microfabrication techniques that limit potential artificial lung geometries, overall device size, and throughput. To address these limitations, a high-resolution Asiga MAX X27 UV digital light processing (DLP) 3D printer and custom photopolymerizable polydimethylsiloxane (PDMS) resin were used to rapidly manufacture small-scale µALs via vat photopolymerization (VPP). Devices were designed in SOLIDWORKS with 500 blood channels and 252 gas channels, where gas and blood flow channels were oriented orthogonally and separated by membranes on the top and bottom, permitting two-sided gas exchange. Successful devices were post-processed to remove uncured resin from microchannels and assembled with external tubing in preparation for gas exchange performance testing with ovine whole blood. 3D printed channel dimensions were 172 µm-tall × 320 µm-wide, with 62 µm-thick membranes and 124 µm-wide support columns. Measured outlet blood oxygen saturation (SO2) agreed with theoretical models and rated flow of the device was 1 mL min-1. Blood side pressure drop was 1.58 mmHg at rated flow. This work presents the highest density of 3D printed microchannels in a single device, one of the highest CO2 transfer efficiencies of any artificial lung to date, and a promising approach to translate µALs one step closer to the clinic.


Subject(s)
Microfluidics , Pulmonary Gas Exchange , Sheep , Animals , Pulmonary Gas Exchange/physiology , Biomimetics , Lung/physiology , Printing, Three-Dimensional
2.
J Pediatr Surg ; 59(1): 103-108, 2024 Jan.
Article in English | MEDLINE | ID: mdl-37858393

ABSTRACT

BACKGROUND: Children with end-stage lung disease are commonly managed with extracorporeal life support (ECLS) as a bridge to lung transplantation. A pumpless artificial lung (MLung) is a portable alternative to ECLS and it allows for ambulation. Both ECLS and pumpless artificial lungs require systemic anticoagulation which is associated with hemorrhagic complications. We tested the MLung with a novel Nitric Oxide (NO) Surface Anticoagulation (NOSA) system, to provide local anticoagulation for 72 h of support in a pediatric-size ovine model. METHODS: Four mini sheep underwent thoracotomy and cannulation of the pulmonary artery (inflow) and left atrium (outflow), recovered and were monitored for 72hr. The circuit tubing and connectors were coated with the combination of an NO donor (diazeniumdiolated dibutylhexanediamine; DBHD-N2O2) and argatroban. The animals were connected to the MLung and 100 ppm of NO was added to the sweep gas. Systemic hemodynamics, blood chemistry, blood gases, and methemoglobin were collected. RESULTS: Mean device flow was 836 ± 121 mL/min. Device outlet saturation was 97 ± 4%. Pressure drop across the lung was 3.5 ± 1.5 mmHg and resistance was 4.3 ± 1.7 mmHg/L/min. Activated clotting time averaged 170 ± 45s. Methemoglobin was 2.9 ± 0.8%. Platelets declined from 590 ± 101 at baseline to 160 ± 90 at 72 h. NO flux (x10-10 mol/min/cm2) of the NOSA circuit averaged 2.8 ± 0.6 (before study) and 1.9 ± 0.1 (72 h) and across the MLung 18 ± 3 NO flux was delivered. CONCLUSION: The MLung is a more portable form of ECLS that demonstrates effective gas exchange for 72 h without hemodynamic changes. Additionally, the NOSA system successfully maintained local anticoagulation without evidence of systemic effects.


Subject(s)
Extracorporeal Membrane Oxygenation , Nitric Oxide , Animals , Humans , Sheep , Child , Methemoglobin , Lung , Hemodynamics , Anticoagulants/pharmacology , Anticoagulants/therapeutic use
3.
Micromachines (Basel) ; 14(4)2023 Mar 30.
Article in English | MEDLINE | ID: mdl-37421006

ABSTRACT

Microfluidics is a rapidly advancing technology with expansive applications but has been restricted by slow, laborious fabrication techniques for polydimethylsiloxane (PDMS)-based devices. Currently, 3D printing promises to address this challenge with high-resolution commercial systems but is limited by a lack of material advances in generating high-fidelity parts with micron-scale features. To overcome this limitation, a low-viscosity, photopolymerizable PDMS resin was formulated with a methacrylate-PDMS copolymer, methacrylate-PDMS telechelic polymer, photoabsorber, Sudan I, photosensitizer, 2-isopropylthioxanthone, and a photoinitiator, 2,4,6-trimethyl benzoyl diphenylphosphine oxide. The performance of this resin was validated on a digital light processing (DLP) 3D printer, an Asiga MAX X27 UV. Resin resolution, part fidelity, mechanical properties, gas permeability, optical transparency, and biocompatibility were investigated. This resin produced resolved, unobstructed channels as small as 38.4 (±5.0) µm tall and membranes as thin as 30.9 (±0.5) µm. The printed material had an elongation at break of 58.6% ± 18.8%, Young's modulus of 0.30 ± 0.04 MPa, and was highly permeable to O2 (596 Barrers) and CO2 (3071 Barrers). Following the ethanol extraction of the unreacted components, this material demonstrated optical clarity and transparency (>80% transmission) and viability as a substrate for in vitro tissue culture. This paper presents a high-resolution, PDMS 3D-printing resin for the facile fabrication of microfluidic and biomedical devices.

4.
ASAIO J ; 69(7): e301-e307, 2023 07 01.
Article in English | MEDLINE | ID: mdl-37146595

ABSTRACT

Portable artificial lung (AL) systems are under development, but there are few technologies available that adjust the carbon dioxide (CO 2 ) removal in response to changes in patient metabolic needs. Our work describes the second generation of a CO 2 -based portable servoregulation system that automatically adjusts CO 2 removal in ALs. Four adult sheep (68 ± 14.3 kg) were used to test the servoregulator. The servoregulator controlled air sweep flow through the lung to meet a target exhaust gas CO 2 (tEGCO 2 ) level in normocapnic and hypercapnic (arterial partial pressure of CO 2 [PaCO 2 ] >60 mm Hg) conditions at varying flow rates (0.5-1.5 L/min) and at tEGCO 2 levels of 10, 20, and 40 mm Hg. In hypercapnic sheep, average post-AL blood partial pressure of CO 2 (pCO 2 ) values were 22.4 ± 3.6 mm Hg for tEGCO 2 of 10 mm Hg, 28.0 ± 4.1 mm Hg for tEGCO 2 of 20 mm Hg and 40.6 ± 4.8 mm Hg for tEGCO 2 of 40 mm Hg. The controller successfully and automatically adjusted the sweep gas flow to rapidly (<10 minutes) meet the tEGCO 2 level when challenged with changes in inlet blood flow or target EGCO 2 levels for all animals. These in vivo data demonstrate an important step toward portable ALs that can automatically modulate CO 2 removal and allow for substantial changes in patient activity or disease status in ambulatory applications.


Subject(s)
Extracorporeal Membrane Oxygenation , Hemodynamics , Animals , Sheep , Carbon Dioxide , Hypercapnia , Lung/metabolism
5.
Bioengineering (Basel) ; 9(10)2022 Oct 21.
Article in English | MEDLINE | ID: mdl-36290561

ABSTRACT

Artificial lung (AL) systems provide respiratory support to patients with severe lung disease, but none can adapt to the changing respiratory needs of the patients. Precisely, none can automatically adjust carbon dioxide (CO2) removal from the blood in response to changes in patient activity or disease status. Because of this, all current systems limit patient comfort, activity level, and rehabilitation. A portable servoregulation controller that automatically modulates CO2 removal in ALs to meet the real-time metabolic demands of the patient is described. The controller is based on a proportional-integral-derivative (PID) based closed-loop feedback control system that modulates sweep gas (air) flow through the AL to maintain a target exhaust gas CO2 partial pressure (target EGCO2 or tEGCO2). The presented work advances previous research by (1) using gas-side sensing that avoids complications and clotting associated with blood-based sensors, (2) incorporating all components into a portable, battery-powered package, and (3) integrating smart moisture removal from the AL to enable long term operation. The performance of the controller was tested in vitro for ∼12 h with anti-coagulated bovine blood and 5 days with distilled water. In tests with blood, the sweep gas flow was automatically adjusted by the controller rapidly (<2 min) meeting the specified tEGCO2 level when confronted with changes in inlet blood partial pressure of CO2 (pCO2) levels at various AL blood flows. Overall, the CO2 removal from the AL showed a strong correlation with blood flow rate and blood pCO2 levels. The controller successfully operated continuously for 5 days when tested with water. This study demonstrates an important step toward ambulatory AL systems that automatically modulate CO2 removal as required by lung disease patients, thereby allowing for physiotherapy, comfort, and activity.

6.
Micromachines (Basel) ; 13(6)2022 May 25.
Article in English | MEDLINE | ID: mdl-35744436

ABSTRACT

Microfluidic artificial lungs (µALs) are being investigated for their ability to closely mimic the size scale and cellular environment of natural lungs. Researchers have developed µALs with small artificial capillary diameters (10-50 µm; to increase gas exchange efficiency) and with large capillary diameters (~100 µm; to simplify design and construction). However, no study has directly investigated the impact of capillary height on µAL properties. Here, we use Murray's law and the Hagen-Poiseuille equation to design single-layer, small-scale µALs with capillary heights between 10 and 100 µm. Each µAL contained two blood channel types: capillaries for gas exchange; and distribution channels for delivering blood to/from capillaries. Three designs with capillary heights of 30, 60, and 100 µm were chosen for further modeling, implementation and testing with blood. Flow simulations were used to validate and ensure equal pressures. Designs were fabricated using soft lithography. Gas exchange and pressure drop were tested using whole bovine blood. All three designs exhibited similar pressure drops and gas exchange; however, the µAL with 60 µm tall capillaries had a significantly higher wall shear rate (although physiologic), smaller priming volume and smaller total blood contacting surface area than the 30 and 100 µm designs. Future µAL designs may need to consider the impact of capillary height when optimizing performance.

7.
ASAIO J ; 68(5): 698-706, 2022 05 01.
Article in English | MEDLINE | ID: mdl-34380953

ABSTRACT

A laptop-driven, benchtop control system that automatically adjusts carbon dioxide (CO2) removal in artificial lungs (ALs) is described. The proportional-integral-derivative (PID) feedback controller modulates pump-driven air sweep gas flow through an AL to achieve a desired exhaust gas CO2 partial pressure (EGCO2). When EGCO2 increases, the servoregulator automatically and rapidly increases sweep flow to remove more CO2. If EGCO2 decreases, the sweep flow decreases to reduce CO2 removal. System operation was tested for 6 hours in vitro using bovine blood and in vivo in three proof-of-concept sheep experiments. In all studies, the controller automatically adjusted the sweep gas flow to rapidly (<1 minute) meet the specified EGCO2 level when challenged with changes in inlet blood or target EGCO2 levels. CO2 removal increased or decreased as a function of arterial pCO2 (PaCO2). Such a system may serve as a controller in wearable AL systems that allow for large changes in patient activity or disease status.


Subject(s)
Extracorporeal Membrane Oxygenation , Wearable Electronic Devices , Animals , Blood Gas Analysis , Carbon Dioxide , Cattle , Humans , Lung/surgery , Respiration, Artificial , Sheep
8.
Micromachines (Basel) ; 12(10)2021 Oct 18.
Article in English | MEDLINE | ID: mdl-34683317

ABSTRACT

The rapid expansion of microfluidic applications in the last decade has been curtailed by slow, laborious microfabrication techniques. Recently, microfluidics has been explored with additive manufacturing (AM), as it has gained legitimacy for producing end-use products and 3D printers have improved resolution capabilities. While AM satisfies many shortcomings with current microfabrication techniques, there still lacks a suitable replacement for the most used material in microfluidic devices, poly(dimethylsiloxane) (PDMS). Formulation of a gas-permeable, high-resolution PDMS resin was developed using a methacrylate-PDMS copolymer and the novel combination of a photoabsorber, Sudan I, and photosensitizer, 2-Isopropylthioxanthone. Resin characterization and 3D printing were performed using a commercially available DLP-SLA system. A previously developed math model, mechanical testing, optical transmission, and gas-permeability testing were performed to validate the optimized resin formula. The resulting resin has Young's modulus of 11.5 MPa, a 12% elongation at break, and optical transmission of >75% for wavelengths between 500 and 800 nm after polymerization, and is capable of creating channels as small as 60 µm in height and membranes as thin as 20 µm. The potential of AM is just being realized as a fabrication technique for microfluidics as developments in material science and 3D printing technologies continue to push the resolution capabilities of these systems.

9.
Analyst ; 145(23): 7582-7594, 2020 Nov 23.
Article in English | MEDLINE | ID: mdl-32966357

ABSTRACT

We describe a microfabricated passive preconcentrator (µPP) intended for integration into gas chromatographic microsystems (µGC) for analyzing volatile/semi-volatile organic compounds (S/VOC). Devices (8 × 8 mm) were made from a silicon-on-insulator top layer and a glass bottom layer. The top layer has 237 apertures (47 × 47 µm) distributed around the periphery of a circular region (5.2 mm o.d.) through which ambient vapors diffuse at predictable rates. Two internal annular cavities offset from the apertures are packed with ∼800 µg each of commercial carbon adsorbents. Thin-film heaters thermally desorb captured vapors, which are drawn by a pump through a central exit port to a micro injector for analysis with a bench scale GC. The 15 test compounds spanned a vapor pressure range of 0.033 to 1.1 kPa. Effective (diffusional) µPP sampling rates ranged from 0.16 to 0.78 mL min-1 for short-duration exposures to ∼mg m-3 vapor concentrations. Observed and modeled sampling rates generally agreed within 15%. Sampling rates for two representative compounds declined by ≤30% between 0.25 and 24 h of continuous exposure. For one of these, the sampling rate declined by only 8% over a ∼2300-fold concentration range (0.25 h samples). Desorption (transfer) efficiencies were >95% for most compounds (250-275 °C, 60 s, 5 mL min-1). Sampling rates for mixtures matched those for the individual compounds. Dissipating no energy while sampling, additional advantages of this novel device include short- or long-term sampling, high capacity and transfer efficiency for a diverse set of S/VOCs, low transfer flow rate, and a robust fabrication process.

10.
Acta Biomater ; 112: 190-201, 2020 08.
Article in English | MEDLINE | ID: mdl-32434076

ABSTRACT

Microfluidic artificial lungs (µALs) have the potential to improve the treatment and quality of life for patients with acute or chronic lung injury. In order to realize the full potential of this technology (including as a destination therapy), the biocompatibility of these devices needs to be improved to produce long-lasting devices that are safe for patient use with minimal or no systemic anticoagulation. Many studies exist which probe coagulation and thrombosis on polydimethyl siloxane (PDMS) surfaces, and many strategies have been explored to improve surface biocompatibility. As the field of µALs is young, there are few studies which investigate biocompatibility of functioning µALs; and even fewer which were performed in vivo. Here, we use both in vitro and in vivo models to investigate two strategies to improve µAL biocompatibility: 1) a hydrophilic surface coating (polyethylene glycol, PEG) to prevent surface fouling, and 2) the addition of nitric oxide (NO) to the sweep gas to inhibit platelet activation locally within the µAL. In this study, we challenge µALs with clottable blood or platelet-rich plasma (PRP) and monitor the resistance to blood flow over time. Device lifetime (the amount of time the µAL remains patent and unobstructed by clot) is used as the primary indicator of biocompatibility. This study is the first study to: 1) investigate the effect of NO release on biocompatibility in a microfluidic network; 2) combine a hydrophilic PEG coating with NO release to improve blood compatibility; and 3) perform extended in vivo biocompatibility testing of a µAL. We found that µALs challenged in vitro with PRP remained patent significantly longer when the sweep gas contained NO than without NO. In the in vivo rabbit model, neither approach alone (PEG coating nor NO sweep gas) significantly improved biocompatibility compared to controls (though with larger sample size significance may become apparent); while the combination of a PEG coating with NO sweep gas resulted in significant improvement of device lifetime. STATEMENT OF SIGNIFICANCE: The development of microfluidic artificial lungs (µALs) can potentially have a massive impact on the treatment of patients with acute and chronic lung impairments. Before these devices can be deployed clinically, the biocompatibility of µALs must be improved and more comprehensively understood. This work explores two strategies for improving biocompatibility, a hydrophilic surface coating (polyethylene glycol) for general surface passivation and the addition of nitric oxide (NO) to the sweep gas to quell platelet and leukocyte activation. These two strategies are investigated separately and as a combined device treatment. Devices are challenged with clottable blood using in vitro testing and in vivo testing in rabbits. This is the first study to our knowledge that allows statistical comparisons of biocompatible µALs in animals, a key step towards eventual clinical use.


Subject(s)
Microfluidics , Quality of Life , Animals , Blood Platelets , Humans , Hydrophobic and Hydrophilic Interactions , Lung , Rabbits
11.
ASAIO J ; 66(9): 1054-1062, 2020.
Article in English | MEDLINE | ID: mdl-32149748

ABSTRACT

Microchannel artificial lungs may provide highly efficient, long-term respiratory support, but a robust predictive oxygen transfer (VO2) model is needed to better design them. To meet this need, we first investigated the predictive accuracy of Mikic, Benn, and Drinker's advancing front (AF) oxygen transfer theory by applying it to previous microchannel lung studies. Here, the model that included membrane resistance showed no bias toward overprediction or underprediction of VO2 (median error: -1.13%, interquartile range: [-26.9%, 19.2%]) and matched closely with existing theory. Next, this theory was expanded into a general model for investigating a family of designs. The overall model suggests that, for VO2 = 100 ml/min, fraction of delivered oxygen (FDO2) = 40%, wall shear stress ((Equation is included in full-text article.)) = 30 dyn/cm, and blood channel height = 20-50 µm, a compact design can be achieved with priming volume ((Equation is included in full-text article.)) = 5.8-32 ml; however, manifolding may be challenging to satisfy the rigorous total width ((Equation is included in full-text article.)) requirement ((Equation is included in full-text article.)= 76-475 m). In comparison, 100-200 µm heights would yield larger dimensions ((Equation is included in full-text article.)122-478 ml) but simpler manifolding ((Equation is included in full-text article.)4.75-19.0 m). The device size can be further adjusted by varying FDO2, (Equation is included in full-text article.), or VO2. This model may thus serve as a simple yet useful tool to better design microchannel artificial lungs.


Subject(s)
Artificial Organs , Equipment Design , Lung , Models, Biological
12.
ASAIO J ; 66(4): 423-432, 2020 04.
Article in English | MEDLINE | ID: mdl-31192843

ABSTRACT

Children with end-stage lung failure awaiting lung transplant would benefit from improvements in artificial lung technology allowing for wearable pulmonary support as a bridge-to-transplant therapy. In this work, we designed, fabricated, and tested the Pediatric MLung-a dual-inlet hollow fiber artificial lung based on concentric gating, which has a rated flow of 1 L/min, and a pressure drop of 25 mm Hg at rated flow. This device and future iterations of the current design are designed to relieve pulmonary arterial hypertension, provide pulmonary support, reduce ventilator-associated injury, and allow for more effective therapy of patients with end-stage lung disease, including bridge-to-transplant treatment.


Subject(s)
Artificial Organs , Respiratory Insufficiency/therapy , Child , Equipment Design , Humans , Lung Transplantation
13.
IEEE Trans Biomed Eng ; 66(4): 1082-1093, 2019 04.
Article in English | MEDLINE | ID: mdl-30139043

ABSTRACT

OBJECTIVE: Microfluidic artificial lungs (µALs) are being researched for future clinical use due to the potential for increased gas exchange efficiency, small blood contacting surface area, small priming volume, and biomimetic blood flow paths. However, a current roadblock to clinical use is the need to stack hundreds to thousands of these small-scale µALs in parallel to reach clinically relevant blood flows. The need for so many layers not only increases the complexity and projected cost to manufacture a µAL, but also could result in devices which are cumbersome, and, therefore, not suitable for portable artificial lung systems. METHODS: Here, we describe the design analysis and optimization of a single-layer µAL that simultaneously calculates rated blood flow, blood contacting surface area, blood volume, pressure drop, and shear stress as a function of blood channel height using previously developed closed-form mathematical equations. A µAL designed using this procedure is then implemented and tested. RESULTS: The resulting device exhibits a rated flow of 17 mL/min and reduces the number of layers required for clinically relevant µAL devices by a factor of up to 32X compared to previous work. CONCLUSION: This procedure could significantly reduce manufacturing complexity as well as eliminate a barrier to the clinical application of these promising devices. SIGNIFICANCE: The described method results in the highest rated flow for any single-layer µAL to date.


Subject(s)
Lab-On-A-Chip Devices , Lung/blood supply , Lung/physiology , Microfluidic Analytical Techniques/instrumentation , Models, Biological , Animals , Artificial Organs , Biomedical Engineering/methods , Dimethylpolysiloxanes/chemistry , Equipment Design , Humans , Nylons/chemistry , Regional Blood Flow/physiology
14.
Annu Int Conf IEEE Eng Med Biol Soc ; 2018: 2989-2992, 2018 Jul.
Article in English | MEDLINE | ID: mdl-30441026

ABSTRACT

Real-time monitoring of arteriovenous graft blood flow would provide early warning of graft failure to permit interventions such as angioplasty or graft replacement to avoid catastrophic failure. We have developed a new type of flexible pulsation sensor (FPS) consisting of a 3D printed elastic cuff wrapped around a graft and thus not in contact with blood. The FPS uses multi-walled carbon nanotubes (MWCNTs) dispersed in polydimethylsiloxane (PDMS) as a piezoresistive sensor layer, which is embedded within structural thixotropic PDMS. These materials were specifically developed to enable sensor additive manufacturing via 3D Bio-plotting, and the resulting strain sensor is more compliant and has a wider maximum strain range than graft materials. Here, we analyze the strain transduction mechanics on a vascular graft and describe the memristive properties of MWCNT-PDMS composites, which may be mitigated using AC biasing. In vitro testing of the FPS on a vascular graft phantom showed a robust, linear sensor output to pulsatile flows (170-650 mL/min) and pressures (62-175 mmHg). The FPS showed an RMS error when measuring pressure and flow of 7.7 mmHg and 29.3 mL/min, with a mean measurement error of 6.5% (pressure) and 8.0% (flow).


Subject(s)
Printing, Three-Dimensional , Dimethylpolysiloxanes , Nanotubes, Carbon , Pressure
15.
Langmuir ; 34(1): 492-502, 2018 01 09.
Article in English | MEDLINE | ID: mdl-29231737

ABSTRACT

Blood-material interactions are crucial to the lifetime, safety, and overall success of blood contacting devices. Hydrophilic polymer coatings have been employed to improve device lifetime by shielding blood contacting materials from the natural foreign body response, primarily the intrinsic pathway of the coagulation cascade. These coatings have the ability to repel proteins, cells, bacteria, and other micro-organisms. Coatings are desired to have long-term stability, so that the nonthrombogenic and nonfouling effects gained are long lasting. Unfortunately, there exist limited studies which investigate their stability under dynamic flow conditions as encountered in a physiological setting. In addition, direct comparisons between multiple coatings are lacking in the literature. In this study, we investigate the stability of polyethylene glycol (PEG), zwitterionic sulfobetaine silane (SBSi), and zwitterionic polyethylene glycol sulfobetaine silane (PEG-SBSi) grafted by a room temperature, sequential flow chemistry process on polydimethylsiloxane (PDMS) over time under ambient, static fluid (no flow), and physiologically relevant flow conditions and compare the results to uncoated PDMS controls. PEG, SBSi, and PEG-SBSi coatings maintained contact angles below 20° for up to 35 days under ambient conditions. SBSi and PEG-SBSi showed increased stability and hydrophilicity after 7 days under static conditions. They also retained contact angles ≤40° for all shear rates after 7 days under flow, demonstrating their potential for long-term stability. The effectiveness of the coatings to resist platelet adhesion was also studied under physiological flow conditions. PEG showed a 69% reduction in adhered platelets, PEG-SBSi a significant 80% reduction, and SBSi a significant 96% reduction compared to uncoated control samples, demonstrating their potential applicability for blood contacting applications. In addition, the presented coatings and their stability under shear may be of interest in other applications including marine coatings, lab on a chip devices, and contact lenses, where it is desirable to reduce surface fouling due to proteins, cells, and other organisms.


Subject(s)
Dimethylpolysiloxanes/chemistry , Lab-On-A-Chip Devices , Polyethylene Glycols/chemistry , Adsorption , Hydrophobic and Hydrophilic Interactions , Platelet Adhesiveness/drug effects , Silanes/chemistry , Surface Properties
16.
ASAIO J ; 63(5): 637-643, 2017.
Article in English | MEDLINE | ID: mdl-28665829

ABSTRACT

Current hollow fiber membrane lungs feature a predominantly straight blood path length across the fiber bundle, resulting in limited O2 transfer efficiency because of the diffusion boundary layer effect. Using computational fluid dynamics and optical flow visualization methods, a hollow fiber membrane lung was designed comprising unique concentric circular blood flow paths connected by gates. The prototype lung, comprising a fiber surface area of 0.28 m, has a rated flow of 2 L/min, and the oxygenation efficiency is 357 ml/min/m. The CO2 clearance of the lung is 200 ml/min at the rated blood flow. Given its high gas transfer efficiency, as well as its compact size, low priming volume, and propensity for minimal thrombogenicity, this lung design has the potential to be used in a range of acute and chronic respiratory support applications, including providing total respiratory support for infants and small children and CO2 clearance in adults.


Subject(s)
Oxygenators, Membrane , Adult , Carbon Dioxide/blood , Child , Equipment Design , Humans , Lung/physiology , Oxygen/blood
17.
Int J Polym Mater ; 65(15): 769-778, 2016.
Article in English | MEDLINE | ID: mdl-27493297

ABSTRACT

Due to the role of nitric oxide (NO) in regulating a variety of biological functions in humans, numerous studies on different NO releasing/generating materials have been published over the past two decades. Although NO has been demonstrated to be a strong antimicrobial and potent antithrombotic agent, NO-releasing (NOrel) polymers have not reached the clinical setting. While increasing the concentration of the NO donor in the polymer is a common method to prolong the NO-release, this should not be at the cost of mechanical strength or biocompatibility of the original material. In this work, it was shown that the incorporation of S-nitroso-penicillamine (SNAP), an NO donor molecule, into Elast-eon E2As (a copolymer of mixed soft segments of polydimethylsiloxane and poly(hexamethylene oxide)), does not adversely impact the physical and biological attributes of the base polymer. Incorporating 10 wt % of SNAP into E2As reduces the ultimate tensile strength by only 20%. The inclusion of SNAP did not significantly affect the surface chemistry or roughness of E2As polymer. Ultraviolet radiation, ethylene oxide, and hydrogen peroxide vapor sterilization techniques retained approximately 90% of the active SNAP content, where sterilization of these materials did not affect the NO-release profile over an 18 day period. Furthermore, these NOrel materials were shown to be biocompatible with the host tissues as observed through hemocompatibility and cytotoxicity analysis. In addition, the stability of SNAP in E2As was studied under a variety of storage conditions, as they pertain to translational potential of these materials. SNAP-incorporated E2As stored at room temperature for over 6 months retained 87% of its initial SNAP content. Stored and fresh films exhibited similar NO release kinetics over an 18 day period. Combined, the results from this study suggest that SNAP-doped E2As polymer is suitable for commercial biomedical applications due to the reported physical and biological characteristics that are important for commercial and clinical success.

18.
ASAIO J ; 62(4): 470-6, 2016.
Article in English | MEDLINE | ID: mdl-27164040

ABSTRACT

Although total body perfusion with extracorporeal life support (ECLS) can be maintained for weeks, individual organ perfusion beyond 12 hours has yet to be achieved clinically. Normothermic ex situ heart perfusion (ESHP) offers the potential for prolonged cardiac preservation. We developed an ESHP system to study the effect of perfusate variables on organ preservation, with the ultimate goal of extending organ perfusion for ≥24 hours. Forty porcine hearts were perfused for a target of 12 hours. Hearts that maintained electromechanical activity and had a <3× increase in vascular resistance were considered successful preservations. Perfusion variables, metabolic byproducts, and histopathology were monitored and sampled to identify factors associated with preservation failure. Twenty-two of 40 hearts were successfully preserved at 12 hours. Successful 12 hour experiments demonstrated lower potassium (4.3 ± 0.8 vs. 5.0 ± 1.2 mmol/L; p = 0.018) and lactate (3.5 ± 2.8 vs. 4.5 ± 2.9 mmol/L; p = 0.139) levels, and histopathology revealed less tissue damage (p = 0.003) and less weight gain (p = 0.072). Results of these early experiments suggest prolonged ESHP is feasible, and that elevated lactate and potassium levels are associated with organ failure. Further studies are necessary to identify the ideal perfusate for normothermic ESHP.


Subject(s)
Heart Transplantation , Organ Preservation/methods , Perfusion/methods , Animals , Swine , Time Factors
19.
Lab Chip ; 16(7): 1274-7, 2016 Apr 07.
Article in English | MEDLINE | ID: mdl-26957040

ABSTRACT

This response explores and discusses the critiques of Wagner et al. in their "Comment on 'The promise of microfluidic artificial lungs' by Joseph A. Potkay, Lab Chip, 2014, 14, 4122-4138".


Subject(s)
Artificial Organs , Lung/physiology , Microfluidic Analytical Techniques , Humans
20.
Proc IEEE Sens ; 20162016.
Article in English | MEDLINE | ID: mdl-31435454

ABSTRACT

Graft wall pulsation amplitude sensing can provide a measure of functional status, e.g. in hemodialysis access grafts. Current implantable graft monitoring sensors require graft modification and direct bloodstream contact. We propose a new class of piezoresistive flexible pulsation sensors which can be wrapped around the graft to measure wall movement. These sensors must be highly flexible to prevent graft constriction; typical strain sensors are too rigid and the strain sensing range is too limited for this application. We describe a novel additive manufacturing (AM) method for printing polydimethylsiloxane (PDMS) with an internal porous structure, such that material compliance may be tuned anisotropically for a given sensor geometry. Prototype flexible pulsation sensors (FPS) consisting of structured PDMS with an embedded conductive PDMS sensor layer were fabricated and tested. Initial tests demonstrated reliable sensor response to 1-Hz cyclic elongation of 20%, and a sensor gauge factor of 1.0.