Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters











Database
Language
Publication year range
1.
Article in English | MEDLINE | ID: mdl-37001202

ABSTRACT

This short communication introduced a simple and sensitive LC-MS/MS method for therapeutic drug monitoring of digoxin in children with the lower limit of quantitation of 0.2 ng/mL based on 30 µL of plasma. The plasma sample was pretreated by one-step protein precipitation. Then the chromatographic separation was performed on a short C-18 column with a total run time of 2.4 min. The detection was achieved through multiple reaction monitoring using positive ionization mode on a triple quadrupole mass spectrometer. The linear range of digoxin in human plasma was among 0.2-6.4 ng/mL. The intra-day and inter-day accuracies of digoxin ranged from -6.0 % to 10.1 % and imprecisions were less than 8.8 %. The extraction recovery rate of digoxin in plasma samples was above 90 %. Matrix factor normalized by internal standard was within acceptance criteria. This method was fully verified and applied to determine the plasma digoxin concentrations of 43 pediatric patients. It is approved appropriate and practical for the therapeutic drug monitoring of digoxin in routine clinical laboratory practice, especially for children.


Subject(s)
Digoxin , Drug Monitoring , Humans , Child , Chromatography, Liquid/methods , Digoxin/chemistry , Drug Monitoring/methods , Tandem Mass Spectrometry/methods , Reproducibility of Results