Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add more filters










Database
Language
Publication year range
2.
Article in English | MEDLINE | ID: mdl-36744654

ABSTRACT

The use of cannulated screws and titanium plates to reinforce the sternal closure or to treat sternal dehiscence after median sternotomy has already been suggested in several articles. The system proposed here has some important advantages over those already described. Moreover, thanks to its characteristics, this system can also be used to treat pathologies affecting the entire rib cage. The system consists of a first threaded cannulated screw that is inserted in the bone or chondral cartilage and accommodates a cap screw that is tightened into the first screw and fixes a plate according to the following scheme: a threaded cannulated screw/plate/cap screw (Brixia system of screws). This system allows the plates to be fixed on the anterior face of the ribs and/or sternum without the need to enlarge dissection of the tissue, thereby lowering the danger of haemorrhage and injury to the thoracic organs. For this reason, it is particularly suitable for treating post-sternotomy sternal dehiscence, but it can be used to reinforce the primary sternal closure (after median or transversal sternotomy) in high-risk patients with sternal dehiscence. Owing to the modular nature of the system, singular components can also be utilized independently.


Subject(s)
Thoracic Surgery , Humans , Titanium/therapeutic use , Surgical Wound Dehiscence/surgery , Sternum/surgery , Sternotomy , Bone Screws , Bone Wires
SELECTION OF CITATIONS
SEARCH DETAIL
...