1.
Front Neurol ; 14: 1255105, 2023.
Article in English | MEDLINE | ID: mdl-38046576

ABSTRACT

Introduction: Patients and technologists commonly describe vertigo, dizziness, and imbalance near high-field magnets, e.g., 7-Tesla (T) magnetic resonance imaging (MRI) scanners. We sought a simple way to alleviate vertigo and dizziness in high-field MRI scanners by applying the understanding of the mechanisms behind magnetic vestibular stimulation and the innate characteristics of vestibular adaptation. Methods: We first created a three-dimensional (3D) control systems model of the direct and indirect vestibulo-ocular reflex (VOR) pathways, including adaptation mechanisms. The goal was to develop a paradigm for human participants undergoing a 7T MRI scan to optimize the speed and acceleration of entry into and exit from the MRI bore to minimize unwanted vertigo. We then applied this paradigm from the model by recording 3D binocular eye movements (horizontal, vertical, and torsion) and the subjective experience of eight normal individuals within a 7T MRI. The independent variables were the duration of entry into and exit from the MRI bore, the time inside the MRI bore, and the magnetic field strength; the dependent variables were nystagmus slow-phase eye velocity (SPV) and the sensation of vertigo. Results: In the model, when the participant was exposed to a linearly increasing magnetic field strength, the per-peak (after entry into the MRI bore) and post-peak (after exiting the MRI bore) responses of nystagmus SPV were reduced with increasing duration of entry and exit, respectively. There was a greater effect on the per-peak response. The entry/exit duration and peak response were inversely related, and the nystagmus was decreased the most with the 5-min duration paradigm (the longest duration modeled). The experimental nystagmus pattern of the eight normal participants matched the model, with increasing entry duration having the strongest effect on the per-peak response of nystagmus SPV. Similarly, all participants described less vertigo with the longer duration entries. Conclusion: Increasing the duration of entry into and exit out of a 7T MRI scanner reduced or eliminated vertigo symptoms and reduced nystagmus peak SPV. Model simulations suggest that central processes of vestibular adaptation account for these effects. Therefore, 2-min entry and 20-s exit durations are a practical solution to mitigate vertigo and other discomforting symptoms associated with undergoing 7T MRI scans. In principle, these findings also apply to different magnet strengths.

2.
OTO Open ; 7(3): e79, 2023.
Article in English | MEDLINE | ID: mdl-37727400

ABSTRACT

7-Tesla (T) magnetic resonance imaging may allow for higher resolution images but may produce greater acoustic noise than 1.5- and 3-T scanners. We sought to characterize the intensity of acoustic noise from 7- versus 3-T scanners. A-weighted sound pressure levels from 5 types of pulse sequences used for brain and inner ear imaging in 3- and 7-T scanners were measured. Time-averaged sound level and maximum sound levels generated for each sequence were compared. Time-averaged sound levels exceeded 95 dB and reached maximums above 105 dB on the majority of 3- and 7-T scans. The mean time-averaged sound level and maximum sound level across pulse sequences were greater in 7- than 3-T (105.6 vs 91.4, P = .01; 114.0 vs. 96.5 dB, P < .01). 7- and 3-T magnetic resonance imaging scanners produce high levels of acoustic noise that exceed acceptable safety limits, emphasizing the need for active and passive noise protection.

3.
J Neurophysiol ; 129(5): 1157-1176, 2023 05 01.
Article in English | MEDLINE | ID: mdl-37018758

ABSTRACT

The otolith end organs inform the brain about gravitational and linear accelerations, driving the otolith-ocular reflex (OOR) to stabilize the eyes during translational motion (e.g., moving forward without rotating) and head tilt with respect to gravity. We previously characterized OOR responses of normal chinchillas to whole body tilt and translation and to prosthetic electrical stimulation targeting the utricle and saccule via electrodes implanted in otherwise normal ears. Here we extend that work to examine OOR responses to tilt and translation stimuli after unilateral intratympanic gentamicin injection and to natural/mechanical and prosthetic/electrical stimulation delivered separately or in combination to animals with bilateral vestibular hypofunction after right ear intratympanic gentamicin injection followed by surgical disruption of the left labyrinth at the time of electrode implantation. Unilateral intratympanic gentamicin injection decreased natural OOR response magnitude to about half of normal, without markedly changing OOR response direction or symmetry. Subsequent surgical disruption of the contralateral labyrinth at the time of electrode implantation surgery further decreased OOR magnitude during natural stimulation, consistent with bimodal-bilateral otolith end organ hypofunction (ototoxic on the right ear, surgical on the left ear). Delivery of pulse frequency- or pulse amplitude-modulated prosthetic/electrical stimulation targeting the left utricle and saccule in phase with whole body tilt and translation motion stimuli yielded responses closer to normal than the deficient OOR responses of those same animals in response to head tilt and translation alone.NEW & NOTEWORTHY Previous studies to expand the scope of prosthetic stimulation of the otolith end organs showed that selective stimulation of the utricle and saccule is possible. This article further defines those possibilities by characterizing a diseased animal model and subsequently studying its responses to electrical stimulation alone and in combination with mechanical motion. We show that we can partially restore responses to tilt and translation in animals with unilateral gentamicin ototoxic injury and contralateral surgical disruption.


Subject(s)
Ototoxicity , Vestibule, Labyrinth , Animals , Reflex, Vestibulo-Ocular/physiology , Otolithic Membrane/physiology , Chinchilla , Gentamicins
4.
PLoS Biol ; 20(9): e3001798, 2022 09.
Article in English | MEDLINE | ID: mdl-36103550

ABSTRACT

Sensory pathways provide complex and multifaceted information to the brain. Recent advances have created new opportunities for applying our understanding of the brain to sensory prothesis development. Yet complex sensor physiology, limited numbers of electrodes, and nonspecific stimulation have proven to be a challenge for many sensory systems. In contrast, the vestibular system is uniquely suited for prosthesis development. Its peripheral anatomy allows site-specific stimulation of 3 separate sensory organs that encode distinct directions of head motion. Accordingly, here, we investigated whether implementing natural encoding strategies improves vestibular prosthesis performance. The eye movements produced by the vestibulo-ocular reflex (VOR), which plays an essential role in maintaining visual stability, were measured to quantify performance. Overall, implementing the natural tuning dynamics of vestibular afferents produced more temporally accurate VOR eye movements. Exploration of the parameter space further revealed that more dynamic tunings were not beneficial due to saturation and unnatural phase advances. Trends were comparable for stimulation encoding virtual versus physical head rotations, with gains enhanced in the latter case. Finally, using computational methods, we found that the same simple model explained the eye movements evoked by sinusoidal and transient stimulation and that a stimulation efficacy substantially less than 100% could account for our results. Taken together, our results establish that prosthesis encodings that incorporate naturalistic afferent dynamics and account for activation efficacy are well suited for restoration of gaze stability. More generally, these results emphasize the benefits of leveraging the brain's endogenous coding strategies in prosthesis development to improve functional outcomes.


Subject(s)
Artificial Limbs , Vestibule, Labyrinth , Animals , Eye Movements , Macaca mulatta , Reflex, Vestibulo-Ocular/physiology , Vestibule, Labyrinth/physiology
5.
IEEE Trans Instrum Meas ; 70: 1-9, 2021.
Article in English | MEDLINE | ID: mdl-33776080

ABSTRACT

OBJECTIVE: Vestibular and oculomotor research often requires measurement of 3-dimensional (3D) eye orientation and movement with high spatial and temporal precision and accuracy. We describe the design, implementation, validation and use of a new magnetic coil system optimized for recording 3D eye movements using small scleral coils in animals. METHODS: Like older systems, the system design uses off-the-shelf components to drive three mutually orthogonal alternating magnetic fields at different frequencies. The scleral coil voltage induced by those fields is decomposed into 3 signals, each related to the coil's orientation relative to the axis of one field component. Unlike older systems based on analog demodulation and filtering, this system uses a field-programmable gate array (FPGA) to oversample each induced scleral coil voltage (at 25 Msamples/s), demodulate in the digital domain, and average over 25 ksamples per data point to generate 1 ksamples/s output in real time. RESULTS: Noise floor is <0.036° peak-to-peak and linearity error is < 0.1° during 345° rotations in all three dimensions. CONCLUSION AND SIGNIFICANCE: This FPGA-based design, which is both reprogrammable and freely available upon request, delivers sufficient performance to record eye movements at high spatial and temporal precision and accuracy using coils small enough for use with small animals.

6.
Sci Rep ; 9(1): 18924, 2019 12 12.
Article in English | MEDLINE | ID: mdl-31831760

ABSTRACT

Recent studies have shown that ionic direct current (iDC) can modulate the vestibular system in-vivo, with potential benefits over conventional pulsed stimulation. In this study, the effects of iDC stimulation on vestibular nerve fiber firing rate was investigated using loose-patch nerve fiber recordings in the acutely excised mouse crista ampullaris of the semicircular canals. Cathodic and anodic iDC steps instantaneously reduced and increased afferent spike rate, with the polarity of this effect dependent on the position of the stimulating electrode. A sustained constant anodic or cathodic current resulted in an adaptation to the stimulus and a return to spontaneous spike rate. Post-adaptation spike rate responses to iDC steps were similar to pre-adaptation controls. At high intensities spike rate response sensitivities were modified by the presence of an adaptation step. Benefits previously observed in behavioral responses to iDC steps delivered after sustained current may be due to post-adaptation changes in afferent sensitivity. These results contribute to an understanding of peripheral spike rate relationships for iDC vestibular stimulation and validate an ex-vivo model for future investigation of cellular mechanisms. In conjunction with previous in-vivo studies, these data help to characterize iDC stimulation as a potential therapy to restore vestibular function after bilateral vestibulopathy.


Subject(s)
Action Potentials/physiology , Adaptation, Physiological , Models, Neurological , Vestibular Nerve/physiology , Vestibule, Labyrinth/physiology , Animals , Female , Ion Transport/physiology , Male , Mice , Mice, Transgenic
7.
JCI Insight ; 4(22)2019 11 14.
Article in English | MEDLINE | ID: mdl-31723056

ABSTRACT

BACKGROUNDBilateral loss of vestibular (inner ear inertial) sensation causes chronically blurred vision during head movement, postural instability, and increased fall risk. Individuals who fail to compensate despite rehabilitation therapy have no adequate treatment options. Analogous to hearing restoration via cochlear implants, prosthetic electrical stimulation of vestibular nerve branches to encode head motion has garnered interest as a potential treatment, but prior studies in humans have not included continuous long-term stimulation or 3D binocular vestibulo-ocular reflex (VOR) oculography, without which one cannot determine whether an implant selectively stimulates the implanted ear's 3 semicircular canals.METHODSWe report binocular 3D VOR responses of 4 human subjects with ototoxic bilateral vestibular loss unilaterally implanted with a Labyrinth Devices Multichannel Vestibular Implant System vestibular implant, which provides continuous, long-term, motion-modulated prosthetic stimulation via electrodes in 3 semicircular canals.RESULTSInitiation of prosthetic stimulation evoked nystagmus that decayed within 30 minutes. Stimulation targeting 1 canal produced 3D VOR responses approximately aligned with that canal's anatomic axis. Targeting multiple canals yielded responses aligned with a vector sum of individual responses. Over 350-812 days of continuous 24 h/d use, modulated electrical stimulation produced stable VOR responses that grew with stimulus intensity and aligned approximately with any specified 3D head rotation axis.CONCLUSIONThese results demonstrate that a vestibular implant can selectively, continuously, and chronically provide artificial sensory input to all 3 implanted semicircular canals in individuals disabled by bilateral vestibular loss, driving reflexive VOR eye movements that approximately align in 3D with the head motion axis encoded by the implant.TRIAL REGISTRATIONClinicalTrials.gov: NCT02725463.FUNDINGNIH/National Institute on Deafness and Other Communication Disorders: R01DC013536 and 2T32DC000023; Labyrinth Devices, LLC; and Med-El GmbH.


Subject(s)
Bilateral Vestibulopathy , Electric Stimulation/instrumentation , Neural Prostheses , Reflex, Vestibulo-Ocular/physiology , Vestibule, Labyrinth , Bilateral Vestibulopathy/physiopathology , Bilateral Vestibulopathy/surgery , Humans , Ototoxicity/physiopathology , Ototoxicity/surgery , Prosthesis Design , Vestibule, Labyrinth/physiopathology , Vestibule, Labyrinth/surgery
8.
J Neurophysiol ; 121(6): 2013-2019, 2019 06 01.
Article in English | MEDLINE | ID: mdl-30969883

ABSTRACT

For many years, people working near strong static magnetic fields of magnetic resonance imaging (MRI) machines have reported dizziness and sensations of vertigo. The discovery a decade ago that a sustained nystagmus can be observed in all humans with an intact labyrinth inside MRI machines led to a possible mechanism: a Lorentz force occurring in the labyrinth from the interactions of normal inner ear ionic currents and the strong static magnetic fields of the MRI machine. Inside an MRI, the Lorentz force acts to induce a constant deflection of the semicircular canal cupula of the superior and lateral semicircular canals. This inner ear stimulation creates a sensation of rotation, and a constant horizontal/torsional nystagmus that can only be observed when visual fixation is removed. Over time, the brain adapts to both the perception of rotation and the nystagmus, with the perception usually diminishing over a few minutes, and the nystagmus persisting at a reduced level for hours. This observation has led to discoveries about how the central vestibular mechanisms adapt to a constant vestibular asymmetry and is a useful model of set-point adaptation or how homeostasis is maintained in response to changes in the internal milieu or the external environment. We review what is known about the effects of stimulation of the vestibular system with high-strength magnetic fields and how the understanding of the mechanism has been refined since it was first proposed. We suggest future ways that magnetic vestibular stimulation might be used to understand vestibular disease and how it might be treated.


Subject(s)
Adaptation, Physiological/physiology , Magnetic Fields/adverse effects , Nystagmus, Physiologic/physiology , Vertigo/physiopathology , Vestibule, Labyrinth/physiology , Humans
9.
Otol Neurotol ; 39(10): e1150-e1159, 2018 12.
Article in English | MEDLINE | ID: mdl-30444848

ABSTRACT

HYPOTHESIS: Magnetic vestibular stimulation (MVS) elicits nystagmus in C57BL/6J mice but not head tilt mice lacking Nox3, which is required for normal otoconial development. BACKGROUND: Humans have vertigo and nystagmus in strong magnetic fields within magnetic resonance imaging machines. The hypothesized mechanism is a Lorentz force driven by electrical current entering the utricular neuroepithelium, acting indirectly on crista hair cells via endolymph movement deflecting cupulae. We tested an alternate hypothesized mechanism: Lorentz action directly on crista hair cell stereocilia, driven by their currents independent of the utricle. METHODS: Before MVS, vestibulo-ocular reflex responses of eight C57BL/6J mice and six head tilt mice were measured during whole-body sinusoidal rotations and tilts using video-oculography. Mice were then placed within a 4.7 Tesla magnetic field with the horizontal semicircular canals approximately Earth-horizontal for ≥1 minute in several head orientations, while eye movements were recorded via infrared video in darkness. RESULTS: Outside the magnet, both C57BL/6J and head tilt mice had intact horizontal vestibulo-ocular reflex, but only C57BL/6J mice exhibited static counter-roll responses to tilt (normal utiruclo-ocular reflex). When placed in the magnet nose-first, C57BL/6J mice had left-beating nystagmus, lasting a median of 32.8 seconds. When tail-first, nystagmus was right-beating and similar duration (median 28.0 s, p > 0.05). In contrast, head tilt mice lacked magnetic field-induced nystagmus (p < 0.001). CONCLUSIONS: C57BL/6J mice generate nystagmus in response to MVS, while mice deficient in Nox3 do not. This suggests 1) a normal utricle is necessary, and 2) functioning semicircular canals are insufficient, to generate MVS-induced nystagmus in mice.


Subject(s)
NADPH Oxidases/metabolism , Nystagmus, Pathologic/physiopathology , Saccule and Utricle/physiology , Animals , Humans , Magnetic Fields , Magnetic Resonance Imaging , Mice , Mice, Inbred C57BL , Reflex, Vestibulo-Ocular/physiology
10.
Front Neurol ; 9: 1197, 2018.
Article in English | MEDLINE | ID: mdl-30723456

ABSTRACT

Background: Strong static magnetic fields such as those in an MRI machine can induce sensations of self-motion and nystagmus. The proposed mechanism is a Lorentz force resulting from the interaction between strong static magnetic fields and ionic currents in the inner ear endolymph that causes displacement of the semicircular canal cupulae. Nystagmus persists throughout an individual's exposure to the magnetic field, though its slow-phase velocity partially declines due to adaptation. After leaving the magnetic field an after effect occurs in which the nystagmus and sensations of rotation reverse direction, reflecting the adaptation that occurred while inside the MRI. However, the effects of visual fixation and of head shaking on this early type of vestibular adaptation are unknown. Methods: Three-dimensional infrared video-oculography was performed in six individuals just before, during (5, 20, or 60 min) and after (4, 15, or 20 min) lying supine inside a 7T MRI scanner. Trials began by entering the magnetic field in darkness followed 60 s later, either by light with visual fixation and head still, or by continuous yaw head rotations (2 Hz) in either darkness or light with visual fixation. Subjects were always placed in darkness 10 or 30 s before exiting the bore. In control conditions subjects remained in the dark with the head still for the entire duration. Results: In darkness with head still all subjects developed horizontal nystagmus inside the magnetic field, with slow-phase velocity partially decreasing over time. An after effect followed on exiting the magnet, with nystagmus in the opposite direction. Nystagmus was suppressed during visual fixation; however, after resuming darkness just before exiting the magnet, nystagmus returned with velocity close to the control condition and with a comparable after effect. Similar after effects occurred with continuous yaw head rotations while in the scanner whether in darkness or light. Conclusions: Visual fixation and sustained head shaking either in the dark or with fixation inside a strong static magnetic field have minimal impact on the short-term mechanisms that attempt to null unwanted spontaneous nystagmus when the head is still, so called VOR set-point adaptation. This contrasts with the critical influence of vision and slippage of images on the retina on the dynamic (gain and direction) components of VOR adaptation.

11.
Front Neurol ; 8: 122, 2017.
Article in English | MEDLINE | ID: mdl-28424657

ABSTRACT

Humans often experience dizziness and vertigo around strong static magnetic fields such as those present in an MRI scanner. Recent evidence supports the idea that this effect is the result of inner ear vestibular stimulation and that the mechanism is a magnetohydrodynamic force (Lorentz force) that is generated by the interactions between normal ionic currents in the inner ear endolymph and the strong static magnetic field of MRI machines. While in the MRI, the Lorentz force displaces the cupula of the lateral and anterior semicircular canals, as if the head was rotating with a constant acceleration. If a human subject's eye movements are recorded when they are in darkness in an MRI machine (i.e., without fixation), there is a persistent nystagmus that diminishes but does not completely disappear over time. When the person exits the magnetic field, there is a transient aftereffect (nystagmus beating in the opposite direction) that reflects adaptation that occurred in the MRI. This magnetic vestibular stimulation (MVS) is a useful technique for exploring set-point adaptation, the process by which the brain adapts to a change in its environment, which in this case is vestibular imbalance. Here, we review the mechanism of MVS, how MVS produces a unique stimulus to the labyrinth that allows us to explore set-point adaptation, and how this technique might apply to the understanding and treatment of vestibular and other neurological disorders.

12.
J Neurol ; 264(Suppl 1): 7-12, 2017 Oct.
Article in English | MEDLINE | ID: mdl-28271407

ABSTRACT

Human subjects placed in strong magnetic fields such as in an MRI scanner often feel dizzy or vertiginous. Recent studies in humans and animals have shown that these effects arise from stimulation of the labyrinth and are accompanied by nystagmus. Here, we measured the three-dimensional pattern of nystagmus using video eye tracking in five normal human subjects placed in a 7T MRI to infer which semicircular canals are activated by magnetic vestibular stimulation. We found that the nystagmus usually had a torsional as well as a horizontal component. Analysis of the relative velocities of the three eye movement components revealed that the lateral and anterior (superior) canals are the only canals activated, and by a similar amount.


Subject(s)
Eye Movements/physiology , Imaging, Three-Dimensional , Nystagmus, Physiologic/physiology , Semicircular Canals/diagnostic imaging , Semicircular Canals/physiology , Adult , Aged , Female , Humans , Male , Middle Aged
13.
J Neurosci Methods ; 283: 7-14, 2017 May 01.
Article in English | MEDLINE | ID: mdl-28300605

ABSTRACT

BACKGROUND: Small, innate asymmetries between the left and right otolith organs can cause ocular misalignment with symptoms that include double vision and motion sickness. Additionally, ocular misalignment affects nearly 5% of the US population. We have developed a portable, non-invasive technology that uses subjective perception of binocular visual signals to estimate relative binocular alignment. NEW METHOD AND RESULTS: The Vertical Alignment Nulling (VAN) and Torsional Alignment Nulling (TAN) tests ask subjects to view one red and one blue line on a tablet computer while looking through color-matched red and blue filters so that each eye sees only one of the lines. Subjects align the red and blue lines, which are initially vertically offset from one another during VAN or rotated relative to one another during TAN, until they perceive a single continuous line. Ocular misalignments are inferred from actual offsets in the final line positions. During testing, all binocular visual cues are eliminated by employing active-matrix organic light-emitting diode (AMOLED) technology and testing in darkness. VAN and TAN can accurately account for visual offsets induced by prisms, and test-retest reliability is excellent, with resolution better than many current standard clinical tests. COMPARISON WITH EXISTING METHOD(S): VAN and TAN tests are similar to the clinical Lancaster red-green test. However, VAN and TAN employ inexpensive, hand-held hardware that can be self-administered with results that are quickly quantifiable. CONCLUSIONS: VAN and TAN provide simple, sensitive, and quantitative measures of binocular positioning alignment that may be useful for detecting subtle abnormalities in ocular positioning.


Subject(s)
Computers, Handheld , Diagnosis, Computer-Assisted/instrumentation , Photic Stimulation/instrumentation , User-Computer Interface , Vision Disparity/physiology , Vision Tests/instrumentation , Vision, Binocular/physiology , Adult , Equipment Design , Equipment Failure Analysis , Eye Movement Measurements , Female , Humans , Male , Photic Stimulation/methods , Reproducibility of Results , Sensitivity and Specificity , Vision Tests/methods
14.
Curr Biol ; 26(10): 1359-66, 2016 05 23.
Article in English | MEDLINE | ID: mdl-27185559

ABSTRACT

A major focus in neurobiology is how the brain adapts its motor behavior to changes in its internal and external environments [1, 2]. Much is known about adaptively optimizing the amplitude and direction of eye and limb movements, for example, but little is known about another essential form of learning, "set-point" adaptation. Set-point adaptation balances tonic activity so that reciprocally acting, agonist and antagonist muscles have a stable platform from which to launch accurate movements. Here, we use the vestibulo-ocular reflex-a simple behavior that stabilizes the position of the eye while the head is moving-to investigate how tonic activity is adapted toward a new set point to prevent eye drift when the head is still [3, 4]. Set-point adaptation was elicited with magneto-hydrodynamic vestibular stimulation (MVS) by placing normal humans in a 7T MRI for 90 min. MVS is ideal for prolonged labyrinthine activation because it mimics constant head acceleration and induces a sustained nystagmus similar to natural vestibular lesions [5, 6]. The MVS-induced nystagmus diminished slowly but incompletely over multiple timescales. We propose a new adaptation hypothesis, using a cascade of imperfect mathematical integrators, that reproduces the response to MVS (and more natural chair rotations), including the gradual decrease in nystagmus as the set point changes over progressively longer time courses. MVS set-point adaptation is a biological model with applications to basic neurophysiological research into all types of movements [7], functional brain imaging [8], and treatment of vestibular and higher-level attentional disorders by introducing new biases to counteract pathological ones [9].


Subject(s)
Eye Movements , Head Movements , Reflex, Vestibulo-Ocular , Vestibule, Labyrinth/physiology , Adaptation, Physiological , Adult , Aged , Female , Humans , Magnetic Fields , Male , Middle Aged , Physical Stimulation , Young Adult
15.
Med Biol Eng Comput ; 54(2-3): 421-30, 2016 Mar.
Article in English | MEDLINE | ID: mdl-26077529

ABSTRACT

After their introduction by Robinson (IEEE Trans Biomed Eng 10:137-145, 1963), magnetic scleral search coils quickly became an accepted standard for precise eye movement recordings. While other techniques such as video-oculography or electro-oculography may be more suitable for routine applications, search coils still provide the best low-noise and low-drift characteristics paired with the highest temporal and spatial resolution. The problem with search coils is that many research laboratories still have their large and expensive coil systems installed and are acquiring eye movement data with old, analog technology. Typically, the number of recording channels is limited and modifications to an existing search coil system can be difficult. We propose a system that allows to retro-fit an existing analog search coil system to become a digital recording system. The system includes digital data acquisition boards and a reference coil as the hardware part, receiver software, and a new calibration method. The circuit design has been kept simple and robust, and the proposed software calibration allows the calibration of a single coil within a few seconds.


Subject(s)
Eye Movements/physiology , Magnetic Phenomena , Video Recording/economics , Video Recording/instrumentation , Adult , Humans , Signal Processing, Computer-Assisted
16.
J Vis ; 15(14): 11, 2015.
Article in English | MEDLINE | ID: mdl-26587699

ABSTRACT

Torsional eye movements are rotations of the eye around the line of sight. Measuring torsion is essential to understanding how the brain controls eye position and how it creates a veridical perception of object orientation in three dimensions. Torsion is also important for diagnosis of many vestibular, neurological, and ophthalmological disorders. Currently, there are multiple devices and methods that produce reliable measurements of horizontal and vertical eye movements. Measuring torsion, however, noninvasively and reliably has been a longstanding challenge, with previous methods lacking real-time capabilities or suffering from intrusive artifacts. We propose a novel method for measuring eye movements in three dimensions using modern computer vision software (OpenCV) and concepts of iris recognition. To measure torsion, we use template matching of the entire iris and automatically account for occlusion of the iris and pupil by the eyelids. The current setup operates binocularly at 100 Hz with noise <0.1° and is accurate within 20° of gaze to the left, to the right, and up and 10° of gaze down. This new method can be widely applicable and fill a gap in many scientific and clinical disciplines.


Subject(s)
Brain/physiology , Electrooculography/methods , Eye Movement Measurements , Imaging, Three-Dimensional , Ocular Motility Disorders/diagnosis , Visual Perception/physiology , Humans , Ocular Motility Disorders/physiopathology , Orientation , Reproducibility of Results , Sensitivity and Specificity
17.
Ann N Y Acad Sci ; 1343: 69-79, 2015 Apr.
Article in English | MEDLINE | ID: mdl-25735662

ABSTRACT

Individuals working next to strong static magnetic fields occasionally report disorientation and vertigo. With the increasing strength of magnetic fields used for magnetic resonance imaging studies, these reports have become more common. It was recently learned that humans, mice, and zebrafish all demonstrate behaviors consistent with constant peripheral vestibular stimulation while inside a strong, static magnetic field. The proposed mechanism for this effect involves a Lorentz force resulting from the interaction of a strong static magnetic field with naturally occurring ionic currents flowing through the inner ear endolymph into vestibular hair cells. The resulting force within the endolymph is strong enough to displace the lateral semicircular canal cupula, inducing vertigo and the horizontal nystagmus seen in normal mice and in humans. This review explores the evidence for interactions of magnetic fields with the vestibular system.


Subject(s)
Electromagnetic Fields/adverse effects , Vertigo/physiopathology , Vestibule, Labyrinth/physiopathology , Animals , Humans , Magnetic Resonance Imaging/adverse effects , Vertigo/etiology
18.
J Neurosci ; 35(3): 1192-8, 2015 Jan 21.
Article in English | MEDLINE | ID: mdl-25609633

ABSTRACT

Previous experiments have shown that the vestibulo-ocular reflex (VOR) is partially suppressed during large head-free gaze (gaze = eye-in-head + head-in-space) shifts when both the eyes and head are moving actively, on a fixed body, or when the eyes are moving actively and the head passively on a fixed body. We tested, in human subjects, the hypothesis that the VOR is also suppressed during gaze saccades made with en bloc, head and body together, rotations. Subjects made saccades by following a target light. During some trials, the chair rotated so as to move the entire body passively before, during, or after a saccade. The modulation of the VOR was a function of both saccade amplitude and the time of the head perturbation relative to saccade onset. Despite the perturbation, gaze remained accurate. Thus, VOR modulation is similar when gaze changes are programmed for the eyes alone or for the eyes and head moving together. We propose that the brain always programs a change in gaze using feedback based on gaze and head signals, rather than on separate eye and head trajectories.


Subject(s)
Feedback, Physiological/physiology , Reflex, Vestibulo-Ocular/physiology , Saccades/physiology , Adult , Aged , Head , Humans , Male , Middle Aged , Orientation/physiology , Rotation
19.
PLoS One ; 9(3): e92109, 2014.
Article in English | MEDLINE | ID: mdl-24647586

ABSTRACT

Zebrafish (Danio rerio) offer advantages as model animals for studies of inner ear development, genetics and ototoxicity. However, traditional assessment of vestibular function in this species using the vestibulo-ocular reflex requires agar-immobilization of individual fish and specialized video, which are difficult and labor-intensive. We report that using a static magnetic field to directly stimulate the zebrafish labyrinth results in an efficient, quantitative behavioral assay in free-swimming fish. We recently observed that humans have sustained nystagmus in high strength magnetic fields, and we attributed this observation to magnetohydrodynamic forces acting on the labyrinths. Here, fish were individually introduced into the center of a vertical 11.7T magnetic field bore for 2-minute intervals, and their movements were tracked. To assess for heading preference relative to a magnetic field, fish were also placed in a horizontally oriented 4.7T magnet in infrared (IR) light. A sub-population was tested again in the magnet after gentamicin bath to ablate lateral line hair cell function. Free-swimming adult zebrafish exhibited markedly altered swimming behavior while in strong static magnetic fields, independent of vision or lateral line function. Two-thirds of fish showed increased swimming velocity or consistent looping/rolling behavior throughout exposure to a strong, vertically oriented magnetic field. Fish also demonstrated altered swimming behavior in a strong horizontally oriented field, demonstrating in most cases preferred swimming direction with respect to the field. These findings could be adapted for 'high-throughput' investigations of the effects of environmental manipulations as well as for changes that occur during development on vestibular function in zebrafish.


Subject(s)
Magnetic Fields , Swimming/physiology , Vestibule, Labyrinth/physiology , Zebrafish/physiology , Animals , Behavior, Animal/drug effects , Behavior, Animal/radiation effects , Female , Gentamicins/pharmacology , Infrared Rays , Locomotion/drug effects , Locomotion/radiation effects , Male , Optical Phenomena , Orientation/drug effects , Orientation/radiation effects , Vestibule, Labyrinth/drug effects
20.
Front Neurol ; 5: 28, 2014.
Article in English | MEDLINE | ID: mdl-24659983

ABSTRACT

We recently discovered that static magnetic fields from high-strength MRI machines induce nystagmus in all normal humans, and that a magneto-hydrodynamic Lorentz force, derived from ionic currents in the endolymph and pushing on the cupula, best explains this effect. Individuals with no labyrinthine function have no nystagmus. The influence of magnetic vestibular stimulation (MVS) in individuals with unilateral deficits in labyrinthine function is unknown and may provide insight into the mechanism of MVS. These individuals should experience MVS, but with a different pattern of nystagmus consistent with their unilateral deficit in labyrinthine function. We recorded eye movements in the static magnetic field of a 7 T MRI machine in nine individuals with unilateral labyrinthine hypofunction, as determined by head impulse testing and vestibular-evoked myogenic potentials (VEMP). Eye movements were recorded using infrared video-oculography. Static head positions were varied in pitch with the body supine, and slow-phase eye velocity (SPV) was assessed. All subjects exhibited predominantly horizontal nystagmus after entering the magnet head-first, lying supine. The SPV direction reversed when entering feet-first. Pitching chin-to-chest caused subjects to reach a null point for horizontal SPV. Right unilateral vestibular hypofunction (UVH) subjects developed slow-phase-up nystagmus and left UVH subjects, slow-phase-down nystagmus. Vertical and torsional components were consistent with superior semicircular canal excitation or inhibition, respectively, of the intact ear. These findings provide compelling support for the hypothesis that MVS is a result of a Lorentz force and suggest that the function of individual structures within the labyrinth can be assessed with MVS. As a novel method of comfortable and sustained labyrinthine stimulation, MVS can provide new insights into vestibular physiology and pathophysiology.