Article in English | MEDLINE | ID: mdl-25734939

ABSTRACT

A Reversed Phase-High Performance Liquid Chromatography/Diode Array Detection method was developed and validated for paracetamol quantification in cell culture fluid from an in vitro Blood Brain Barrier model. The chromatographic method and sample preparation were developed using only aqueous solvents. The column was a XTerra RP18 150 × 4.6mm, 3.5 µm with a guard column XTerra RP18 20 × 4.6 mm, 3.5 µm at 35 °C and the mobile phase was composed by 100% formate buffer 20 mM at pH 4 and flow rate was set at 1 mL/min. The detection was at 242 nm. The sample was injected at 10 µL. Validation was performed using the accuracy profile approach. The analytical procedure was validated with the acceptance limits at ± 10% over a range of concentration from 1 to 58 mg L(-1). The procedure was then used in routine to determine paracetamol concentration in a brain blood barrier in vitro model. Application of the Unither paracetamol formulation in Blood Brain Barrier model allowed the determination and comparison of the transcellular passage of paracetamol at 37 °C and 4 °C, that excludes paracellular or non specific leakage.


Subject(s)
Acetaminophen/analysis , Acetaminophen/pharmacokinetics , Blood-Brain Barrier/metabolism , Chromatography, High Pressure Liquid/methods , Chromatography, Reverse-Phase/methods , Blood-Brain Barrier/cytology , Cell Line , Drug Stability , Humans , Reproducibility of Results , Sensitivity and Specificity
SELECTION OF CITATIONS