1.
J Thorac Imaging ; 2024 May 07.
Article in English | MEDLINE | ID: mdl-38712920

ABSTRACT

PURPOSE: We investigated spatial resolution loss away from isocenter for a prototype deep silicon photon-counting detector (PCD) CT scanner and compare with a clinical energy-integrating detector (EID) CT scanner. MATERIALS AND METHODS: We performed three scans on a wire phantom at four positions (isocenter, 6.7, 11.8, and 17.1 cm off isocenter). The acquisition modes were 120 kV EID CT, 120 kV high-definition (HD) EID CT, and 120 kV PCD CT. HD mode used double the projection view angles per rotation as the "regular" EID scan mode. The diameter of the wire was calculated by taking the full width of half max (FWHM) of a profile drawn over the radial and azimuthal directions of the wire. Change in wire diameter appearance was assessed by calculating the ratio of the radial and azimuthal diameter relative to isocenter. t tests were used to make pairwise comparisons of the wire diameter ratio with each acquisition and mean ratios' difference from unity. RESULTS: Deep silicon PCD CT had statistically smaller (P<0.05) changes in diameter ratio for both radial and azimuthal directions compared with both regular and HD EID modes and was not statistically different from unity (P<0.05). Maximum increases in FWMH relative to isocenter were 36%, 12%, and 1% for regular EID, HD EID, and deep silicon PCD, respectively. CONCLUSION: Deep silicon PCD CT exhibits less change in spatial resolution in both the radial and azimuthal directions compared with EID CT.

2.
AJR Am J Roentgenol ; 221(4): 539-547, 2023 10.
Article in English | MEDLINE | ID: mdl-37255042

ABSTRACT

BACKGROUND. Variable beam hardening based on patient size causes variation in CT numbers for energy-integrating detector (EID) CT. Photon-counting detector (PCD) CT more accurately determines effective beam energy, potentially improving CT number reliability. OBJECTIVE. The purpose of the present study was to compare EID CT and deep silicon PCD CT in terms of both the effect of changes in object size on CT number and the overall accuracy of CT numbers. METHODS. A phantom with polyethylene rings of varying sizes (mimicking patient sizes) as well as inserts of different materials was scanned on an EID CT scanner in single-energy (SE) mode (120-kV images) and in rapid-kilovoltage-switching dual-energy (DE) mode (70-keV images) and on a prototype deep silicon PCD CT scanner (70-keV images). ROIs were placed to measure the CT numbers of the materials. Slopes of CT number as a function of object size were computed. Materials' ideal CT number at 70 keV was computed using the National Institute of Standards and Technology XCOM Photon Cross Sections Database. The root mean square error (RMSE) between measured and ideal numbers was calculated across object sizes. RESULTS. Slope (expressed as Hounsfield units per centimeter) was significantly closer to zero (i.e., less variation in CT number as a function of size) for PCD CT than for SE EID CT for air (1.2 vs 2.4 HU/cm), water (-0.3 vs -1.0 HU/cm), iodine (-1.1 vs -4.5 HU/cm), and bone (-2.5 vs -10.1 HU/cm) and for PCD CT than for DE EID CT for air (1.2 vs 2.8 HU/cm), water (-0.3 vs -1.0 HU/cm), polystyrene (-0.2 vs -0.9 HU/cm), iodine (-1.1 vs -1.9 HU/cm), and bone (-2.5 vs -6.2 HU/cm) (p < .05). For all tested materials, PCD CT had the smallest RMSE, indicating CT numbers closest to ideal numbers; specifically, RMSE (expressed as Hounsfield units) for SE EID CT, DE EID CT, and PCD CT was 32, 44, and 17 HU for air; 7, 8, and 3 HU for water; 9, 10, and 4 HU for polystyrene; 31, 37, and 13 HU for iodine; and 69, 81, and 20 HU for bone, respectively. CONCLUSION. For numerous materials, deep silicon PCD CT, in comparison with SE EID CT and DE EID CT, showed lower CT number variability as a function of size and CT numbers closer to ideal numbers. CLINICAL IMPACT. Greater reliability of CT numbers for PCD CT is important given the dependence of diagnostic pathways on CT numbers.


Subject(s)
Iodine , Silicon , Humans , Reproducibility of Results , Polystyrenes , Tomography, X-Ray Computed/methods , Phantoms, Imaging , Photons , Water
3.
J Phys Chem B ; 126(36): 6922-6935, 2022 09 15.
Article in English | MEDLINE | ID: mdl-36067064

ABSTRACT

In an effort to combat rising antimicrobial resistance, our labs have rationally designed cationic, helical, amphipathic antimicrobial peptides (AMPs) as alternatives to traditional antibiotics since AMPs incur bacterial resistance in weeks, rather than days. One highly positively charged AMP, WLBU2 (+13e), (RRWV RRVR RWVR RVVR VVRR WVRR), has been shown to be effective in killing both Gram-negative (G(-)) and Gram-positive (G(+)) bacteria by directly perturbing the bacterial membrane nonspecifically. Previously, we used two equilibrium experimental methods: synchrotron X-ray diffuse scattering (XDS) providing lipid membrane thickness and neutron reflectometry (NR) providing WLBU2 depth of penetration into three lipid model membranes (LMMs). The purpose of the present study is to use the results from the scattering experiments to guide molecular dynamics (MD) simulations to investigate the detailed biophysics of the interactions of WLBU2 with LMMs of Gram-negative outer and inner membranes, and Gram-positive cell membranes, to elucidate the mechanisms of bacterial killing. Instead of coarse-graining, backmapping, or simulating without bias for several microseconds, all-atom (AA) simulations were guided by the experimental results and then equilibrated for ∼0.5 µs. Multiple replicas of the inserted peptide were run to probe stability and reach a combined time of at least 1.2 µs for G(-) and also 2.0 µs for G(+). The simulations with experimental comparisons help rule out certain structures and orientations and propose the most likely set of structures, orientations, and effects on the membrane. The simulations revealed that water, phosphates, and ions enter the hydrocarbon core when WLBU2 is positioned there. For an inserted peptide, the three types of amino acids, arginine, tryptophan, and valine (R, W, V), are arranged with the 13 Rs extending from the hydrocarbon core to the phosphate group, Ws are located at the interface, and Vs are more centrally located. For a surface state, R, W, and V are positioned relative to the bilayer interface as expected from their hydrophobicities, with Rs closest to the phosphate group, Ws close to the interface, and Vs in between. G(-) and G(+) LMMs are thinned ∼1 Å by the addition of WLBU2. Our results suggest a dual anchoring mechanism for WLBU2 both in the headgroup and in the hydrocarbon region that promotes a defect region where water and ions can flow across the slightly thinned bacterial cell membrane.


Subject(s)
Antimicrobial Peptides , Molecular Dynamics Simulation , Antimicrobial Cationic Peptides/chemistry , Bacteria/metabolism , Lipid Bilayers/chemistry , Lipids , Phosphates , Water
4.
Chemistry ; 26(28): 6247-6256, 2020 May 15.
Article in English | MEDLINE | ID: mdl-32166806

ABSTRACT

In the quest for new antibiotics, two novel engineered cationic antimicrobial peptides (eCAPs) have been rationally designed. WLBU2 and D8 (all 8 valines are the d-enantiomer) efficiently kill both Gram-negative and -positive bacteria, but WLBU2 is toxic and D8 nontoxic to eukaryotic cells. We explore protein secondary structure, location of peptides in six lipid model membranes, changes in membrane structure and pore evidence. We suggest that protein secondary structure is not a critical determinant of bactericidal activity, but that membrane thinning and dual location of WLBU2 and D8 in the membrane headgroup and hydrocarbon region may be important. While neither peptide thins the Gram-negative lipopolysaccharide outer membrane model, both locate deep into its hydrocarbon region where they are primed for self-promoted uptake into the periplasm. The partially α-helical secondary structure of WLBU2 in a red blood cell (RBC) membrane model containing 50 % cholesterol, could play a role in destabilizing this RBC membrane model causing pore formation that is not observed with the D8 random coil, which correlates with RBC hemolysis caused by WLBU2 but not by D8.


Subject(s)
Anti-Bacterial Agents/chemistry , Antimicrobial Cationic Peptides/chemistry , Lipopolysaccharides/chemistry , Membrane Lipids/chemistry , Pseudomonas aeruginosa/chemistry , Anti-Bacterial Agents/metabolism , Antimicrobial Cationic Peptides/metabolism , Cell Membrane/metabolism , Hemolysis , Lipopolysaccharides/metabolism , Membrane Lipids/metabolism , Microbial Sensitivity Tests , Protein Structure, Secondary
SELECTION OF CITATIONS
SEARCH DETAIL
...