Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 20 de 39
Filter
1.
Med Phys ; 2024 Apr 24.
Article in English | MEDLINE | ID: mdl-38656695

ABSTRACT

BACKGROUND: In water calorimetry, absolute dose to water is determined by measuring radiation-induced temperature rises. In conventional water calorimeters, temperature detectors are housed in handmade glass vessels that are filled with high-purity water, thus mitigating radiation-induced exo/endothermic chemical reactions of impurities that would otherwise introduce additional heat gain/loss, known as heat defect. Being hand-crafted, these glass vessels may suffer from imperfections, have shape and design constraints, are often backordered, and can be prohibitively expensive. PURPOSE: The purpose of this work is to determine suitability of 3D-printed plastic vessels that are further coated for use in water calorimetry applications, and to study their stability and characterize their associated heat defect correction factor ( k hd ) ${k_{{\mathrm{hd}}}})$ . This novel vessel production technique would allow for cost-effective rapid construction of vessels that can be produced with high accuracy and designs that are simply not practical with current glass vessel construction techniques. This in turn enables water calorimetry applications in many novel radiation delivery modalities, which may include spherical vessels in GammaKnife ICON water calorimetry as an example. METHODS: Eight vessels were 3D-printed using Accura ClearVue in an SLA 3D-printer. Two vessels were coated with Parylene C and four were coated with Parylene N. The water calorimetry preparation procedures followed for these vessels was identical to that of our traditional glass-vessels (i.e., same cleaning procedures, same high purity water, and same saturation procedures with high purity hydrogen gas). The performance of each vessel was characterized using our in-house built water calorimeter in an Elekta Versa using both 6 MV flattening filter-free (FFF) and 18 MV beams. The stability of the coating as function of time and accumulated dose was evaluated through repeated measurements. k hd ${k_{{\mathrm{hd}}}}\;$ of each vessel was determined through cross-comparisons against an Exradin A1SL ionization chamber with direction calibration link to Canada's primary standard laboratory. RESULTS: k HD ${k_{{\mathrm{HD}}}}\;$ of the two uncoated vessels differed by 2.8% under a 6 MV FFF beam. Vessels coated with Parylenes resulted in a stable and reproducible heat defect for both energies. An overall k hd ${k_{{\mathrm{hd}}}}$ of 1.001 ± 0.010 and 1.005 ± 0.010 were obtained for Parylene N and Parylene C coated vessels respectively. All Parylene coated vessels showed agreement, within the established uncertainties, to the zero-heat defect observed in a hydrogen-saturated glass vessel system. An additional long-term study (17 days) of a Parylene N vessel showed no change in response with accumulated dose and time. Electron microscopy images of a Parylene N coated vessel showed a uniform intact coating after repeated irradiations. CONCLUSIONS: An uncoated 3D-printed vessel is not viable for water calorimetry because it exhibits an unstable vessel-dependent heat defect. However, applying a Parylene coating stabilizes the heat defect, suggesting that coated 3D-printed vessels may be suitable for use in water calorimetry. This method facilitates the creation of intricate vessel shapes, which can be efficiently manufactured using 3D printing.

2.
Med Phys ; 51(4): 2998-3009, 2024 Apr.
Article in English | MEDLINE | ID: mdl-38060696

ABSTRACT

BACKGROUND: The static magnetic field present in magnetic resonance (MR)-guided radiotherapy systems can influence dose deposition and charged particle collection in air-filled ionization chambers. Thus, accurately quantifying the effect of the magnetic field on ionization chamber response is critical for output calibration. Formalisms for reference dosimetry in a magnetic field have been proposed, whereby a magnetic field quality conversion factor kB,Q is defined to account for the combined effects of the magnetic field on the radiation detector. Determination of kB,Q in the literature has focused on Monte Carlo simulation studies, with experimental validation limited to only a few ionization chamber models. PURPOSE: The purpose of this study is to experimentally measure kB,Q for 11 ionization chamber models in two commercially available MR-guided radiotherapy systems: Elekta Unity and ViewRay MRIdian. METHODS: Eleven ionization chamber models were characterized in this study: Exradin A12, A12S, A28, and A26, PTW T31010, T31021, and T31022, and IBA FC23-C, CC25, CC13, and CC08. The experimental method to measure kB,Q utilized cross-calibration against a reference Exradin A1SL chamber. Absorbed dose to water was measured for the reference A1SL chamber positioned parallel to the magnetic field with its centroid placed at the machine isocenter at a depth of 10 cm in water for a 10 × 10 cm2 field size at that depth. Output was subsequently measured with the test chamber at the same point of measurement. kB,Q for the test chamber was computed as the ratio of reference dose to test chamber output, with this procedure repeated for each chamber in each MR-guided radiotherapy system. For the high-field 1.5 T Elekta Unity system, the dependence of kB,Q on the chamber orientation relative to the magnetic field was quantified by rotating the chamber about the machine isocenter. RESULTS: Measured kB,Q values for our test dataset of ionization chamber models ranged from 0.991 to 1.002, and 0.995 to 1.004 for the Elekta Unity and ViewRay MRIdian, respectively, with kB,Q tending to increase as the chamber sensitive volume increased. Measured kB,Q values largely agreed within uncertainty to published Monte Carlo simulation data and available experimental data. kB,Q deviation from unity was minimized for ionization chamber orientation parallel or antiparallel to the magnetic field, with increased deviations observed at perpendicular orientations. Overall (k = 1) uncertainty in the experimental determination of the magnetic field quality conversion factor, kB,Q was 0.71% and 0.72% for the Elekta Unity and ViewRay MRIdian systems, respectively. CONCLUSIONS: For a high-field MR-linac, the characterization of ionization chamber performance as angular orientation varied relative to the magnetic field confirmed that the ideal orientation for output calibration is parallel. For most of these chamber models, this study represents the first experimental characterization of chamber performance in clinical MR-linac beams. This is a critical step toward accurate output calibration for MR-guided radiotherapy systems and the measured kB,Q values will be an important reference data source for forthcoming MR-linac reference dosimetry protocols.


Subject(s)
Radiometry , Radiotherapy, Image-Guided , Relative Biological Effectiveness , Magnetic Fields , Monte Carlo Method , Water
3.
J Appl Clin Med Phys ; 24(11): e14087, 2023 Nov.
Article in English | MEDLINE | ID: mdl-37354202

ABSTRACT

BACKGROUND: Magnetic resonance (MR)-guided radiation therapy provides capabilities to utilize high-resolution and real-time MR imaging before and during treatment, which is critical for adaptive radiotherapy. This emerging modality has been promptly adopted in the clinic settings in advance of adaptations to reference dosimetry formalism that are needed to account for the presence of strong magnetic fields. In particular, the influence of magnetic field on the uncertainty of parameters in the reference dosimetry equation needs to be determined in order to fully characterize the uncertainty budget for reference dosimetry in MR-guided radiation therapy systems. PURPOSE: To identify and quantify key sources of uncertainty in the reference dosimetry of external high energy radiotherapy beams in the presence of a strong magnetic field. METHODS: In the absence of a formalized Task Group report for reference dosimetry in MR-integrated linacs, the currently suggested formalism follows the TG-51 protocol with the addition of a quality conversion factor kBQ accounting for the effects of the magnetic field on ionization chamber response. In this work, we quantify various sources of uncertainty that impact each of the parameters in the formalism, and evaluate their overall contribution to the final dose. Measurements are done in a 1.5 T MR-Linac (Unity, Elekta AB, Stockholm, Sweden) which integrates a 1.5 T Philips MR scanner and a 7 MVFFF linac. The responses of several reference-class small volume ionization chambers (Exradin:A1SL, IBA:CC13, PTW:Semiflex-3D) and Farmer type ionization chambers (Exradin:A19, IBA:FC65-G) were evaluated throughout this process. Long-term reproducibility and stability of beam quality, TPR 10 20 ${\mathrm{TPR}}_{10}^{20}$ , was also measured with an in-house built phantom. RESULTS: Relative to the conventional external high energy linacs, the uncertainty on overall reference dose in MR-linac is more significantly affected by the chamber setup: A translational displacement along y-axis of ± 3 mm results in dose variation of < |0.20| ± 0.02% (k = 1), while rotation of ± 5° in horizontal and vertical parallel planes relative to relative to the direction of magnetic field, did not exceed variation of < |0.44| ± 0.02% for all 5 ionization chambers. We measured a larger dose variation for xy-plane (horizontal) rotations (< |0.44| ± 0.02% (k = 1)) than for yz-plane (vertical) rotations (< ||0.28| ± 0.02% (k = 1)), which we associate with the gradient of kB,Q as a function of chamber orientation with respect to direction of the B0 -field. Uncertainty in Pion (for two depths), Ppol (with various sub-studies including effects of cable length, cable looping in the MRgRT bore, connector type in magnetic environment), and Prp were determined. Combined conversion factor kQ × kB,Q was provided for two reference depths at four cardinal angle orientations. Over a two-year period, beam quality was quite stable with TPR 10 20 ${\mathrm{TPR}}_{10}^{20}$ being 0.669 ± 0.01%. The actual magnitude of TPR 10 20 ${\mathrm{TPR}}_{10}^{20}$ was measured using identical equipment and compared between two different Elekta Unity MR-Linacs with results agreeing to within 0.21%. CONCLUSION: In this work, the uncertainty of a number of parameters influencing reference dosimetry was quantified. The results of this work can be used to identify best practice guidelines for reference dosimetry in the presence of magnetic fields, and to evaluate an uncertainty budget for future reference dosimetry protocols for MR-linac.


Subject(s)
Particle Accelerators , Radiometry , Humans , Uncertainty , Reproducibility of Results , Radiometry/methods , Magnetic Resonance Imaging , Magnetic Resonance Spectroscopy
4.
Med Phys ; 49(11): 6739-6764, 2022 Nov.
Article in English | MEDLINE | ID: mdl-36000424

ABSTRACT

Practical guidelines that are not explicit in the TG-51 protocol and its Addendum for photon beam dosimetry are presented for the implementation of the TG-51 protocol for reference dosimetry of external high-energy photon and electron beams. These guidelines pertain to: (i) measurement of depth-ionization curves required to obtain beam quality specifiers for the selection of beam quality conversion factors, (ii) considerations for the dosimetry system and specifications of a reference-class ionization chamber, (iii) commissioning a dosimetry system and frequency of measurements, (iv) positioning/aligning the water tank and ionization chamber for depth ionization and reference dose measurements, (v) requirements for ancillary equipment needed to measure charge (triaxial cables and electrometers) and to correct for environmental conditions, and (vi) translation from dose at the reference depth to that at the depth required by the treatment planning system. Procedures are identified to achieve the most accurate results (errors up to 8% have been observed) and, where applicable, a commonly used simplified procedure is described and the impact on reference dosimetry measurements is discussed so that the medical physicist can be informed on where to allocate resources.

5.
Med Phys ; 49(11): 7071-7084, 2022 Nov.
Article in English | MEDLINE | ID: mdl-35842918

ABSTRACT

PURPOSE: Target localization, for stereotactic radiosurgery (SRS) treatment with Gamma Knife, has become increasingly reliant on the co-registration between the planning MRI and the stereotactic cone-beam computed tomography (CBCT). Validating image registration between modalities would be particularly beneficial when considering the emergence of novel functional and metabolic MRI pulse sequences for target delineation. This study aimed to develop a phantom-based methodology to quantitatively compare the co-registration accuracy of the standard clinical imaging protocol to a representative MRI sequence that was likely to fail co-registration. The comparative methodology presented in this study may serve as a useful tool to evaluate the clinical translatability of novel MRI sequences. METHODS: A realistic human skull phantom with fiducial marker columns was designed and manufactured to fit into a typical MRI head coil and the Gamma Knife patient positioning system. A series of "optimized" 3D MRI sequences-T1 -weighted Dixon, T1 -weighted fast field echo (FFE), and T2 -weighted fluid-attenuated inversion recovery (FLAIR)-were acquired and co-registered to the CBCT. The same sequences were "compromised" by reconstructing without geometric distortion correction and re-collecting with lower signal-to-noise-ratio (SNR) to simulate a novel MRI sequence with poor co-registration accuracy. Image similarity metrics-structural similarity (SSIM) index, mean squared error (MSE), and peak SNR (PSNR)-were used to quantitatively compare the co-registration of the optimized and compromised MR images. RESULTS: The ground truth fiducial positions were compared to positions measured from each optimized image volume revealing a maximum median geometric uncertainty of 0.39 mm (LR), 0.92 mm (AP), and 0.13 mm (SI) between the CT and CBCT, 0.60 mm (LR), 0.36 mm (AP), and 0.07 mm (SI) between the CT and T1 -weighted Dixon, 0.42 mm (LR), 0.23 mm (AP), and 0.08 mm (SI) between the CT and T1 -weighted FFE, and 0.45 mm (LR), 0.19 mm (AP), and 1.04 mm (SI) between the CT and T2 -weighted FLAIR. Qualitatively, pairs of optimized and compromised image slices were compared using a fusion image where separable colors were used to differentiate between images. Quantitatively, MSE was the most predictive and SSIM the second most predictive metric for evaluating co-registration similarity. A clinically relevant threshold of MSE, SSIM, and/or PSNR may be defined beyond which point an MRI sequence should be rejected for target delineation based on its dissimilarity to an optimized sequence co-registration. All dissimilarity thresholds calculated using correlation coefficients with in-plane geometric uncertainty would need to be defined on a sequence-by-sequence basis and validated with patient data. CONCLUSION: This study utilized a realistic skull phantom and image similarity metrics to develop a methodology capable of quantitatively assessing whether a modern research-based MRI sequence can be co-registered to the Gamma Knife CBCT with equal or less than equal accuracy when compared to a clinically accepted protocol.


Subject(s)
Radiosurgery , Humans , Skull
6.
Med Phys ; 49(8): 5483-5490, 2022 Aug.
Article in English | MEDLINE | ID: mdl-35536047

ABSTRACT

PURPOSE: To optimize the design, develop and test a prototype ionization chamber for accurate daily output constancy measurements in solid phantoms in clinical magnetic resonance-guided radiation therapy (MRgRT) radiotherapy beams. Up to 4% variations in response using commercial ionization chambers have been previously reported; the prototype ionization chamber developed here aims to minimize these variations. METHODS: Monte Carlo simulations with the EGSnrc code system are used to optimize an ionization chamber design by increasing the thickness of a brass (high-density, nonferromagnetic, easy-to-machine) wall until results consistent with no air gap are produced for simulations with a 1.5 T and 0.35 T magnetic field, with a 0.2 mm air gap and varying the placement of the chamber model within the air gap. Based on the results of these simulations, prototype ionization chambers are manufactured and tested in conventional linac beams and in a 7 MV Elekta Unity MR-linac. The chambers are rotated about their axes, both parallel and perpendicular to the 1.5 T magnetic field, through 360º in a plastic phantom with measurements made at each cardinal angle. This reveals any variation in chamber response by varying the thickness of the air gap between the chamber and the phantom. RESULTS: Monte Carlo simulations demonstrate that the optimal thickness of the chamber wall to mitigate the effect of an asymmetric air gap between the chamber and the plastic phantom is 1.1 mm of brass. With this thickness, the differences between simulations with and without an air gap and with asymmetric placement of the chamber within the air gap are less than 0.2%. A prototype chamber constructed with a 1.1 mm brass wall thickness exhibits less than 0.3% variation in response when rotated about its axis in the plastic phantom in a beam from an MR-linac, independent of whether its axis is parallel or perpendicular to the magnetic field. CONCLUSION: The optimized ionization chamber design and validated prototype for accurate MR-linac daily output constancy measurements allows utilization of conventional phantoms and procedures in MRgRT systems. This can minimize disruption to clinical workflow for MR-linac quality assurance measurements.


Subject(s)
Radiotherapy, Image-Guided , Magnetic Fields , Magnetic Resonance Spectroscopy , Monte Carlo Method , Particle Accelerators , Phantoms, Imaging , Plastics , Radiometry/methods , Radiotherapy, Image-Guided/methods
7.
Med Phys ; 48(11): 7476-7492, 2021 Nov.
Article in English | MEDLINE | ID: mdl-34549805

ABSTRACT

PURPOSE: To characterize and build a cylindrically layered graphite calorimeter the size of a thimble ionization chamber for absolute dosimetry of small fields. This detector has been designed in a familiar probe format to facilitate integration into the clinical workflow. The feasibility of operating this absorbed dose calorimeter in quasi-adiabatic mode is assessed for high-energy accelerator-based photon beams. METHODS: This detector, herein referred to as Aerrow MK7, is a miniaturized version of a previously validated aerogel-insulated graphite calorimeter known as Aerrow. The new model was designed and developed using numerical methods. Medium conversion factors from graphite to water, small-field output correction factors, and layer perturbation factors for this dosimeter were calculated using the EGSnrc Monte Carlo code system. A range of commercially available aerogel densities were studied for the insulating layers, and an optimal density was selected by minimizing the small-field output correction factors. Heat exchange within the detector was simulated using a five-body compartmental heat transfer model. In quasi-adiabatic mode, the sensitive volume (a 3 mm diameter cylindrical graphite core) experiences a temperature rise during irradiation on the order of 1.3 mK·Gy-1 . The absorbed dose is obtained by calculating the product of this temperature rise with the specific heat capacity of the graphite. The detector was irradiated with 6 MV ( % dd ( 10 ) x  = 63.5%) and 10 MV ( % dd ( 10 ) x  = 71.1%) flattening filter-free (FFF) photon beams for two field sizes, characterized by S clin dimensions of 2.16 and 11.0 cm. The dose readings were compared against a calibrated Exradin A1SL ionization chamber. All dose values are reported at d max in water. RESULTS: The field output correction factors for this dosimeter design were computed for field sizes ranging from S clin  = 0.54 to 11.0 cm. For all aerogel densities studied, these correction factors did not exceed 1.5%. The relative dose difference between the two dosimeters ranged between 0.3% and 0.7% for all beams and field sizes. The smallest field size experimentally investigated, S clin  = 2.16 cm, which was irradiated with the 10 MV FFF beam, produced readings of 84.4 cGy (±1.3%) in the calorimeter and 84.5 cGy (±1.3%) in the ionization chamber. CONCLUSION: The median relative difference in absorbed dose values between a calibrated A1SL ionization chamber and the proposed novel graphite calorimeter was 0.6%. This preliminary experimental validation demonstrates that Aerrow MK7 is capable of accurate and reproducible absorbed dose measurements in quasi-adiabatic mode.


Subject(s)
Graphite , Feasibility Studies , Monte Carlo Method , Photons , Radiometry
9.
J Neurooncol ; 152(1): 173-182, 2021 Mar.
Article in English | MEDLINE | ID: mdl-33453002

ABSTRACT

PURPOSE: The concept of a radioresistant (RR) phenotype has been challenged with use of stereotactic body radiotherapy (SBRT). We compared outcomes following SBRT to RR spinal metastases to a radiosensitive cohort. METHODS: Renal cell, melanoma, sarcoma, gastro-intestinal, and thyroid spinal metastases were identified as RR and prostate cancer (PCA) as radiosensitive. The primary endpoint was MRI-based local failure (LF). Secondary endpoints included overall survival (OS) and vertebral compression fracture (VCF). RESULTS: From a prospectively maintained database of 1394 spinal segments in 605 patients treated with spine SBRT, 173 patients/395 RR spinal segments were compared to 94 patients/185 PCA segments. Most received 24-28 Gy in 2 fractions (68.9%) and median follow-up was 15.5 months (range, 1.4-84.2 months). 1- and 2-year LF rates were 19.2% and 22.4% for RR metastases, respectively, which were significantly greater (p < 0.001) than PCA (3.2% and 8.4%, respectively). Epidural disease (HR: 2.47, 95% CI 1.65-3.71, p < 0.001) and RR histology (HR: 2.41, 95% CI 1.45-3.99, p < 0.001) predicted for greater LF. Median OS was 17.4 and 61.0 months for RR and PCA cohorts, respectively. Lung/liver metastases, polymetastatic disease and epidural disease predicted for worse OS. 2-year VCF rates were ~ 13% in both cohorts. Coverage of the CTV V90 (clinical target volume receiving 90% of prescription dose) by ≥ 87% (HR: 2.32, 95% CI 1.29-4.18, p = 0.005), no prior spine radiotherapy (HR: 1.96, 95% CI 1.09-3.55, p = 0.025), and a greater Spinal Instability Neoplasia Score (p = 0.013) predicted for VCF. CONCLUSIONS: Higher rates of LF were observed after spine SBRT in RR metastases. Optimization strategies include dose escalation and aggressive management of epidural disease.


Subject(s)
Radiation Tolerance/radiation effects , Radiosurgery/methods , Spinal Neoplasms/radiotherapy , Adolescent , Adult , Aged , Aged, 80 and over , Cohort Studies , Female , Humans , Male , Middle Aged , Spinal Neoplasms/secondary , Treatment Outcome , Young Adult
10.
Med Phys ; 47(12): 6458-6469, 2020 Dec.
Article in English | MEDLINE | ID: mdl-32970325

ABSTRACT

PURPOSE: To use a portable 4°C cooled MR-compatible water calorimeter to measure absorbed dose in a magnetic resonance-guided radiation therapy (MRgRT) system. Furthermore, to use the calorimetric dose results and direct cross-calibration to experimentally measure the combined beam quality and magnetic field correction factor ( k Q mag ) of a clinically used reference-class ionization chamber placed under the same radiation field. METHODS: An Elekta Unity MR-linac (7 MV FFF, B = 1.5 T) was used in this study. Measurements were taken using the in-house designed and built water calorimeter. Following preparation and cooling of the system, the MR-compatible calorimeter was positioned using a combination of MR and EPID imaging and the dose to water was measured by monitoring the radiation-induced temperature change. Immediately after the calorimetric measurements, an A1SL ionization chamber was placed inside the calorimeter for direct cross-calibration. The results allowed for a direct and absolute experimental measurement of k Q mag for this chamber and comparison against existing Monte Carlo values. RESULTS: The calorimeter was successfully positioned using imaging in under an hour. The 1-hour setup time is from the time the calorimeter leaves storage to the first calorimetric measurement. Absorbed dose was successfully measured with a relative combined standard uncertainty of 0.71 % (k = 1). Through a cross-calibration, the k Q mag for an Exradin A1SL ionization chamber, set up perpendicular to the incident photon beam and opposite to the direction of the Lorentz force, was directly determined in water in absolute terms to be 0.977 ± 0.010. The currently published k Q mag results, obtained via Monte Carlo calculations, agree with experimental measurements in this work within combined uncertainties. CONCLUSIONS: A novel design of an MR-compatible water calorimeter was successfully used to measure absorbed dose in an MR-linac and determine an experimental value of k Q mag for a clinically used ionization chamber.


Subject(s)
Radiometry , Water , Calorimetry , Magnetic Fields , Particle Accelerators
11.
Med Phys ; 47(10): 5312-5323, 2020 Oct.
Article in English | MEDLINE | ID: mdl-32786081

ABSTRACT

PURPOSE: The purpose of this study is to design a water calorimeter with three goals in mind: (a) To be fully magnetic resonance (MR)-compatible; (b) To be imaged using kV cone beam computed tomography (CBCT), MV portal imaging or MRI for accurate positioning; (c) To accommodate both vertical and horizontal beam incidence, as well as volumetric deliveries or Gamma Knife®. Following this, the calorimeter performance will be measured using an accelerator-based high-energy photon beam. METHODS: A portable 4°C cooled stagnant water calorimeter was built using MR-compatible materials. The walls consist of layers of acrylic plastic, aerogel-based material acting as thermal insulation, as well as tubing for coolant to flow to keep the calorimeter temperature stable at 4°C. The lid contains additional pathways for coolant to flow through as well as two hydraulically driven stirrers. The water calorimeter was positioned in an Elekta Versa using kV CBCT imaging as well as orthogonal MV image pairs. Absolute absorbed dose to water was then determined under a 6 MV flattening filter-free (FFF) beam. This was compared against reference dosimetry results that were measured under identical conditions with an Exradin A1SL ionization chamber with a calibration coefficient directly traceable to the National Research Council Canada. RESULTS: The dose to water determined with the calorimeter (n = 30) agreed with the A1SL ionization chamber reference dose measurements (n = 15) to within 0.25%. The uncertainty associated with the water calorimeter absorbed dose measurement was estimated to be 0.54% (k = 1). CONCLUSIONS: An MR-compatible water calorimeter was successfully built and absolute absorbed dose to water under a conventional 6 MV FFF beam was determined successfully as a first-stage validation of the system.


Subject(s)
Radiometry , Water , Calibration , Calorimetry , Canada , Magnetic Resonance Spectroscopy , Photons
12.
Phys Med Biol ; 65(6): 065003, 2020 03 11.
Article in English | MEDLINE | ID: mdl-31914427

ABSTRACT

Currently, the American Association of Physicists in Medicine (AAPM) TG-21 is the conventional protocol currently used for the calibration of the Leksell Gamma Knife® (LGK) (despite the publication of the AAPM TG-51 protocol). However, this protocol is based on the air-kerma standards requiring an elaborate conversion process resulting in an increase in the possibility of errors in the clinic. The International Atomic Energy Agency (IAEA) Technical Reports Series (TRS)-483 Code of Practice provides new recommendations on the dosimetry of small static fields and correction factor data for the calibration of the LGK unit. The purpose of this study is to experimentally validate previously calculated [Formula: see text] factors for the calibration of the LGK Perfexion/Icon unit in the context of the TRS-483 protocol. An experimental comparison between three protocols (TG-51, TG-21 and TRS-483 with the aforementioned correction factors) for calibration of the LGK unit is provided. Dose-rate measurements were performed on a LGK Icon unit using three ionization chambers and three phantoms with different orientations of the chambers with respect to the LGK unit. The dose rate was determined following the three calibration protocols. The standard deviation on the mean dose rate over all phantoms and chambers in different orientations determined using TG-51, TG-21 and TRS-483 protocols were 0.9%, 0.5% and 0.4%, respectively. The mean dose rate calculated using TG-51 protocol was 1.6% and 1.2% lower comparing to the TG-21 and TRS-483 protocols respectively. Applying the [Formula: see text] values calculated in Mirzakhanian et al (2018) to the measured dose rates in LGK unit for all chambers and phantoms resulted in dose rates that are consistent to within 0.4%. The TRS-483 protocol improves the consistency of the results especially when the chamber was positioned in different orientations with respect to the LGK (from 1.6% when using TG-51 or TG-21 protocols to 0.2% when using TRS-483 protocol) since the other protocols do not correct for the different chamber orientations.


Subject(s)
International Agencies , Nuclear Energy , Radiosurgery , Calibration , Humans , Phantoms, Imaging , Photons , Radiometry/methods , Reproducibility of Results , Water
13.
Med Phys ; 47(4): 1940-1948, 2020 Apr.
Article in English | MEDLINE | ID: mdl-31955432

ABSTRACT

PURPOSE: To measure ionization chamber dose response as a function of the angle between magnetic field direction and ionization chamber orientation in magnetic resonance-guided radiation therapy (MRgRT) system, and to evaluate angular dependence of magnetic field correction factor for reference dosimetry. METHODS: Measurements were performed on an Elekta MR-linac that integrates a 1.5-T Philips MRI and a 7-MV FFF photon beam accelerator. The response of four reference class chambers (Exradin-A19, A1SL, IBA FC65-G, and CC13, paired with a PTW UE electrometer) was studied. An in-house built MR-compatible water tank and an accompanying cylindrical insert that allowed chamber rotation around the cylinder's axis was used. The EPID onboard imaging was used to center chamber at the MR-linac isocenter (143.5 cm, SAD), as well as to verify position at each datapoint. RESULTS: A clear angular dependence of dose response for all chambers has been measured. The most significant effect of magnetic field on relative chamber response in the presence of magnetic field was observed in the orientation when chamber axis is perpendicular to the direction of magnetic field with the tip pointing in the same direction as Lorentz force. This effect is more pronounced for larger volume chambers; the maximum relative variation in the chamber response (between the setup described above and the one where chamber and magnetic field are parallel) is a 5.3% and 4.6% increase for A19 and FC65-G, respectively, and only 2.0% and 1.9% for smaller volume A1SL and CC13 chamber, respectively. We measured the absolute magnitude of the magnetic field correction factor k Q mag for the Exradin-A19, A1SL, IBA FC65-G, and CC13 to be 0.938 ± 1.13%, 0.968 ± 0.99%, 0.950 ± 1.13%, and 0.975 ± 1.13%, respectively. The values are for perpendicular orientation of the chamber relative to magnetic field and parallel to the Lorentz force. CONCLUSIONS: Experimental measurements carried out in this study have verified the optimal orientation of ionization chamber in terms of minimizing effect of magnetic field on the chamber dose response. This study provides a detailed high-resolution measurement of absolute k Q mag values for four reference class chambers as a function of the angle between ionization chamber's central axis and the direction of strong magnetic field over a full 360° rotation. The experimental results of this study can further be used for optimization of the actual sensitive volume of the chamber (and analysis of dead volume) in future Monte Carlo chamber simulations in the presence of strong magnetic fields. In addition, it will provide some necessary data for future reference dosimetry protocols for MR-linac.


Subject(s)
Magnetic Fields , Magnetic Resonance Imaging/instrumentation , Particle Accelerators , Radiometry/instrumentation
14.
Biomed Phys Eng Express ; 7(1)2020 12 21.
Article in English | MEDLINE | ID: mdl-35037902

ABSTRACT

Objective: MR-linac machines are being developed for image-guided radiation therapy but the magnetic field of such machines could affect dose distributions. The purpose of this work was to evaluate the effect of a magnetic field on linac beam dosimetric parameters including penumbra for circular cones used in radiosurgery.Methods: Monte Carlo simulation was conducted for a linac machine with circular cones at 6 MV beam. A homogenous magnetic field of 1.5 T was applied transversely and parallel to the radiation beam. Percentage depth dose (PDD) and beam profiles in a water phantom with and without the magnetic field were calculated.Results: The results have shown that when the magnetic field is applied transversely, the PDDs in the water phantom differ in the buildup region and distant part of PDD curves. The beam profiles at three different depths are all significantly different from those without the magnetic field. The penumbra is greater when a magnetic field has been applied.Conclusion: Linear accelerator-based SRT and SRS use small circular cones. The beam penumbra for these cones can change in the presence of a magnetic field. The perturbation of dose distribution has been also observed in a patient plan due to the presence of a magnetic field. The results of this study show that dose distributions in the presence of a magnetic field must be considered for MR-guided radiotherapy treatments.


Subject(s)
Radiosurgery , Humans , Magnetic Fields , Particle Accelerators , Radiosurgery/methods , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted/methods
15.
Biomed Phys Eng Express ; 6(1): 015021, 2020 01 13.
Article in English | MEDLINE | ID: mdl-33438609

ABSTRACT

PURPOSE: The purpose of this study was to examine RBE variation as a function of distance from the radioactive source, and the potential impact of this variation on a realistic prostate brachytherapy treatment plan. METHODS: Three brachytherapy sources (125I, 192Ir, and 169Yb) were modelled in Geant4 Monte Carlo code, and the resulting electron energy spectrum in water in 3D space around these sources was scored (voxel size of 2 mm3). With this energy spectrum, microdosimetric techniques were used to calculate the maximum RBE, RBEM, as a function of distance from the source. RBEM of 125I relative to 192Ir was calculated in order to validate simulations against literature; all other RBEM calculations were done by normalizing electron fluence at various distances to the source position. In order to examine the impact of RBEM variation in treatment planning, a realistic 192Ir prostate plan was re-evaluated in terms of RBE instead of absorbed dose. RESULTS: The RBEM of 125I, 192Ir, and 169Yb at 8 cm away from the source was 0.994 (+/-0.002), 1.030 (+/-0.003), and 1.066 (+/-0.008), respectively. RBEM in the HDR prostate treatment plan exhibited several hot (+3.6% in RBEM) spots. CONCLUSIONS: The large increase RBEM observed in 169Yb has not yet been described in the literature. Despite the presence of radiobiological hotspots in the HDR treatment, these variations are likely nominal and clinically insignificant.


Subject(s)
Brachytherapy/methods , Iodine Radioisotopes/therapeutic use , Iridium Radioisotopes/therapeutic use , Prostatic Neoplasms/radiotherapy , Radiotherapy Planning, Computer-Assisted/methods , Relative Biological Effectiveness , Ytterbium/therapeutic use , Humans , Male , Monte Carlo Method , Prostatic Neoplasms/pathology , Radiotherapy Dosage
16.
Med Phys ; 47(3): 1291-1304, 2020 Mar.
Article in English | MEDLINE | ID: mdl-31834640

ABSTRACT

PURPOSE: In this work, the fabrication, operation, and evaluation of a probe-format graphite calorimeter - herein referred to as Aerrow - as an absolute clinical dosimeter of high-energy photon beams while in the presence of a B = 1.5 T magnetic field is described. Comparable to a cylindrical ionization chamber (IC) in terms of utility and usability, Aerrow has been developed for the purpose of accurately measuring absorbed dose to water in the clinic with a minimum disruption to the existing clinical workflow. To our knowledge, this is the first reported application of graphite calorimetry to magnetic resonance imaging (MRI)-guided radiotherapy. METHODS: Based on a previously numerically optimized and experimentally validated design, an Aerrow prototype capable of isothermal operation was constructed in-house. Graphite-to-water dose conversions as well as magnetic field perturbation factors were calculated using Monte Carlo, while heat transfer and mass impurity corrections and uncertainties were assessed analytically. Reference dose measurements were performed in the absence and presence of a B = 1.5 T magnetic field using Aerrow in the 7 MV FFF photon beam of an Elekta MRI-linac and were directly compared to the results obtained using two calibrated reference-class IC types. The feasibility of performing solid phantom-based dosimetry with Aerrow and the possible influence of clearance gaps is also investigated by performing reference-type dosimetry measurements for multiple rotational positions of the detector and comparing the results to those obtained in water. RESULTS: In the absence of the B-field, as well as in the parallel orientation while in the presence of the B-field, the absorbed dose to water measured using Aerrow was found to agree within combined uncertainties with those derived from TG-51 using calibrated reference-class ICs. Statistically significant differences on the order of (2-4)%, however, were observed when measuring absorbed dose to water using the ICs in the perpendicular orientation in the presence of the B-field. Aerrow had a peak-to-peak response of about 0.5% when rotated within the solid phantom regardless of whether the B-field was present or not. CONCLUSIONS: This work describes the successful use of Aerrow as a straightforward means of measuring absolute dose to water for large high-energy photon fields in the presence of a 1.5 T B-field to a greater accuracy than currently achievable with ICs. The detector-phantom air gap does not appear to significantly influence the response of Aerrow in absolute terms, nor does it contribute to its rotational dependence. This work suggests that the accurate use of solid phantoms for absolute point dose measurement is possible with Aerrow.


Subject(s)
Calorimetry/instrumentation , Magnetic Resonance Imaging , Particle Accelerators , Phantoms, Imaging , Photons/therapeutic use , Radiotherapy, Image-Guided/instrumentation , Water , Hot Temperature , Radiometry , Uncertainty
17.
Int J Radiat Oncol Biol Phys ; 105(2): 394-399, 2019 10 01.
Article in English | MEDLINE | ID: mdl-31283978

ABSTRACT

PURPOSE: To compare normal tissue dosimetry between hippocampal-avoidance whole brain radiation therapy (HA-WBRT) and stereotactic radiosurgery (SRS) in patients with 10 to 30 brain metastases, and to describe a novel SRS strategy we term Spatially Partitioned Adaptive RadiosurgEry (SPARE). METHODS AND MATERIALS: A retrospective review identified SRS treatment plans with >10 brain metastases located >5 mm from the hippocampi. Our Gamma Knife Icon (GKI) SPARE (GKI-Spr) technique treats multiple metastases with single-fraction SRS partitioned over consecutive days while limiting the total treatment time to ≤60 minutes per day. Hippocampal and normal brain dosimetry were compared among GKI-Spr, single-fraction single-day GKI (GKI-Sfr), and 30 Gy in 10 fractions HA-WBRT. Dose metrics were converted to equivalent dose in 2 Gy fractions. RESULTS: Ten cases were analyzed. Compared with HA-WBRT, GKI-Spr significantly reduced the median equivalent dose in 2 Gy fractions hippocampal maximum point dose, mean dose, and dose to 40% of the hippocampi (D40%) by 86%, 93%, and 93%, respectively, and similarly for GKI-Sfr by 81%, 92%, and 91%, respectively. The normal brain median mean dose was reduced by 95% with GKI-Spr and 94% with GKI-Sfr. Compared with GKI-Sfr, GKI-Spr further reduced all normal brain and hippocampal dose metrics (P ≤ .014). CONCLUSIONS: GKI yields superior hippocampal and normal brain dosimetry compared with HA-WBRT, and GKI-Spr results in further dosimetric advantages.


Subject(s)
Brain Neoplasms/radiotherapy , Brain Neoplasms/secondary , Cranial Irradiation/methods , Hippocampus/radiation effects , Organ Sparing Treatments/methods , Organs at Risk/radiation effects , Radiosurgery/methods , Algorithms , Brain/radiation effects , Humans , Radiotherapy Dosage , Retrospective Studies
18.
Med Phys ; 46(9): 4215-4223, 2019 Sep.
Article in English | MEDLINE | ID: mdl-31264229

ABSTRACT

PURPOSE: Through the addition of high-Z dopants, the sensitivity of plastic scintillators to low-energy radiation can be increased. This study quantifies this change in sensitivity as a function of dopant concentration. METHODS: Measurements were conducted using four different lead-doped scintillators (0%, 1%, 1.5%, and 5% Pb) in high-energy electrons (6 to 15 MeV) and low-energy photon (100 to 300 kVp) radiation fields. High-energy and low-energy irradiations were done using a clinical linear accelerator and an orthovoltage unit, respectively. Light emitted by the scintillator was quantified using a photosensor module. The experimental setup was replicated in Geant4.10.3 Monte Carlo and scintillator parameters (Quenching parameter: kB and the light yield: L0 ) were varied until agreement between measured and simulated results was reached. Monoenergetic electrons were used to simulate the high-energy electron beam while a spectrum generated using SpekCalc® software was used in the low-energy simulations. Light produced by the scintillator was quantified using a flux scorer sensitive only to photons in the visible wavelength range. In order to compare measured and simulated results, the light produced by the scintillator was normalized to the absorbed dose-to-water at the point of measurement. RESULTS: At high lead dopant concentrations, the scintillator's sensitivity to the 100 kVp beam increased by 474% relative to the 15 MeV electron beam; the scintillator's kB parameter increased from 0.126 to 0.27 mm/MeV. A model quantifying the change in kB and L0 as a function of Zeff was derived; presenting a modified Birks' Law for metal-doped plastic scintillators. CONCLUSION: The impact of high-Z doping on plastic scintillator response was quantified; this can allow for the controlled induction of energy dependence in plastic scintillator detectors.


Subject(s)
Lead , Plastics , Scintillation Counting/instrumentation , Monte Carlo Method
19.
Med Phys ; 46(4): 1874-1882, 2019 Apr.
Article in English | MEDLINE | ID: mdl-30729543

ABSTRACT

PURPOSE: With the introduction of a novel graphite probe calorimeter, called the Aerrow, various thermal insulating materials are being explored to further improve the device. Silica-based aerogels are proving to be an optimal material due to their low densities, small thermal conductivities, rigidity, and machinability. The aim of this work is to determine how various silica aerogel densities affect the Aerrow's performance. METHODS: Performance concerns three areas: heat transfer from the core, the Aerrow's beam quality dependence, and the effects of an applied magnetic field on its measurement of absorbed dose to water. A numerical heat transfer study was done to determine heat transfer time constants. The EGSnrc radiation transport toolkit was used to determine absorbed dose conversion factors which are used to quantify the Aerrow's beam quality dependence. Dose conversion factors for Cobalt-60 and two clinical photon beams (6 and 10 MV) were determined. Magnetic field perturbation factors are used to characterize the Aerrow's performance under an applied magnetic field. EGSnrc with the magnetic field transport algorithm was used to determine these perturbations for a 1.5 T MR-linac. Several aerogel densities (0.01-0.55 g  cm - 3 ) were examined for each performance area. RESULTS: Heat transfer time constants were found to vary from 52 ± 2 to 117.4 ± 0.4 s. The time constants decreased with increasing aerogel density. The Aerrow's beam quality dependence varied between 0.5% and 1%, decreasing with increasing aerogel density. Beam quality dependence was determined in the range of 60 Co to 10 MV (58.4%  ≤  % d d ( 10 ) x  ≤ 73.5%). Under an applied magnetic field, perturbations were smallest when the Aerrow was parallel to the field. Perturbations varied more so when the Aerrow was perpendicular to the magnetic field and increased with increasing aerogel density. In all cases, perturbations were less than 0.6% from unity with a relative uncertainty of 0.1%. CONCLUSION: Silica-based aerogels demonstrate an improved performance over thermal insulation used in previous iterations of the Aerrow. With it, the Aerrow has shown to be robust in several areas. If heat transfer can be properly corrected for in the dose determination and the parallel orientation is used under a magnetic field, then the high density aerogel is possibly more preferable.


Subject(s)
Algorithms , Calorimetry/instrumentation , Graphite/chemistry , Phantoms, Imaging , Radiometry/instrumentation , Silicon Dioxide/chemistry , Calorimetry/methods , Hot Temperature , Humans , Magnetic Fields , Monte Carlo Method , Uncertainty , Water
20.
Neurosurgery ; 85(1): E101-E108, 2019 07 01.
Article in English | MEDLINE | ID: mdl-30032234

ABSTRACT

BACKGROUND: On a new dedicated radiosurgery unit enabling frameless treatments, a cone-beam computed tomography (CBCT) can be used for stereotactic definition. Since magnetic resonance imaging (MRI) is used to delineate target, reproducible MRI-to-CBCT coregistration is vital for accurate target localization. OBJECTIVE: To evaluate reproducibility of image coregistration in patient images. METHODS: Three types of coregistration (source-to-target) were analyzed: (1) MRI-to-CT; (2) MRI-to-CBCT; and (3) CT-to-CBCT. For each patient (n = 15), each coregistration type was independently performed 5 to 30 times (total: 465 coregistrations). Each coregistration yielded a transformation matrix, which was subsequently applied to transform every point in the source image to stereotactic coordinates. Two metrics were measured: (1) target registration error (TRE): mean distance between the registered position of each target point and the average registration position of that point; (2) compound registration error (CRE): mean spatial difference between stereotactic coordinates using (A) MRI-to-CT-to-CBCT and (B) MRI-to-CBCT. RESULTS: The median (range) of TRE was 0.11 mm (0.06-0.22 mm), 0.17 mm (0.10-0.36 mm), and 0.12 mm (0.08-0.21 mm) for MRI-to-CT, MRI-to-CBCT, and CT-to-CBCT, respectively. The TRE for MRI-to-CBCT was statistically higher than the other 2 methods (P < .01). The median (range) of CRE was 0.44 mm (0.22-0.59 mm). The maximum point CRE between patients ranged from 0.37-1.15 mm when considering all MRI points, but reduced to 0.31-0.90 mm within the central 16 cm. The CRE varied across the image volume, and typically was minimized near the center. CONCLUSION: The variation in image coregistration is within 0.2 mm, indicating a high degree of reproducibility. The CRE varies throughout the head but is submillimeter in the central 16 cm region.


Subject(s)
Cone-Beam Computed Tomography/methods , Magnetic Resonance Imaging/methods , Neuronavigation/methods , Radiosurgery/methods , Humans , Imaging, Three-Dimensional/methods , Male , Reproducibility of Results
SELECTION OF CITATIONS
SEARCH DETAIL
...