Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 20 de 31
Filter
1.
J Appl Clin Med Phys ; 25(7): e14311, 2024 Jul.
Article in English | MEDLINE | ID: mdl-38386919

ABSTRACT

MOTIVATION: Online adaptive radiotherapy with Ethos is based on the anatomy determined from daily cone beam computed tomography (CBCT) images. Dose optimization and computation are performed on the density map of a synthetic CT (sCT), a deformable registration of the initial planning CT (pCT) onto the current CBCT. Large density changes as present in the lung region are challenging the system. METHODS: Treatment plans for Ethos were created and delivered for 1, 2, and 3 cm diameter lung lesions in an anthropomorphic phantom, combining different insets in the pCT and during adaptive and non-adaptive treatment sessions. Primary and secondary dose calculations as well as back-projected dose from portal images were evaluated. RESULTS: Density changes due to changed insets were not considered in the sCTs. This resulted in errors in the dose; for example, -15.9% of the mean dose for a plan when changing from a 3 cm inset in the pCT to 1 cm at the time of treatment. Secondary dose calculation is based on the sCT and could therefore not reveal these dose errors. However, dose calculation on the CBCT, either as a recalculation in the treatment planning system or as pre-treatment quality assurance (QA) before the treatment, indicated the differences. EPID in-vivo QA also reported discrepancies between calculated and delivered dose distributions. CONCLUSIONS: An incorrect density distribution in the sCT has an impact on the dose calculation accuracy in the adaptive treatment workflow with the Ethos system. Additional quality checks of the sCT can detect such errors.


Subject(s)
Cone-Beam Computed Tomography , Feasibility Studies , Lung Neoplasms , Phantoms, Imaging , Quality Assurance, Health Care , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted , Radiotherapy, Intensity-Modulated , Humans , Radiotherapy Planning, Computer-Assisted/methods , Lung Neoplasms/radiotherapy , Lung Neoplasms/diagnostic imaging , Cone-Beam Computed Tomography/methods , Quality Assurance, Health Care/standards , Radiotherapy, Intensity-Modulated/methods , Image Processing, Computer-Assisted/methods , Radiotherapy, Image-Guided/methods , Organs at Risk/radiation effects , Algorithms
2.
J Appl Clin Med Phys ; 25(2): e14172, 2024 Feb.
Article in English | MEDLINE | ID: mdl-37793069

ABSTRACT

Cranial stereotactic irradiations require accurate reproduction of the planning CT patient position at the time of treatment, including removal of rotational offsets. A device prototype was evaluated for potential clinical use to correct rotational positional offsets in image-guided radiotherapy workflow. Analysis was carried out with a prototype device "RPS head" by gKteso GmbH, rotatable up to 4° in three dimensions by hand wheels. A software tool accounts for the nonrectangular rotation axes and also indicates translational motions to be performed with the standard couch to correct the initial offset and translational shifts introduced by the rotational motion. The accuracy of angular corrections and positioning of an Alderson RANDO head phantom using the prototype device was evaluated for nine treatment plans for cranial targets. Corrections were obtained from cone beam computed tomography (CBCT) imaging. The phantom position was adjusted and the final position was then verified by another CBCT. The long-term stability of the prototype device was evaluated. Attenuation by the device along its three main axes was assessed. A planning study was performed to evaluate if regions of high-density material can be avoided during plan generation. The device enabled the accurate correction of rotational offsets in a clinical setup with a mean residual angular difference of (0.0 ± 0.1)° and a maximum deviation of 0.2°. Translational offsets were less than 1 mm. The device was stable over a period of 20 min, not changing the head support plate position by more than (0.7 ± 0.6) mm. The device contains high-density material in the adjustment mechanism and slightly higher density in the support structures. These can be avoided during planning generation maintaining comparable plan quality. The head positioning device can be used to correct rotational offsets in a clinical setting.


Subject(s)
Radiosurgery , Radiotherapy, Image-Guided , Humans , Patient Positioning , Radiotherapy, Image-Guided/methods , Rotation , Phantoms, Imaging , Software , Cone-Beam Computed Tomography/methods , Radiotherapy Planning, Computer-Assisted/methods , Radiosurgery/methods
3.
J Appl Clin Med Phys ; 24(12): e14139, 2023 Dec.
Article in English | MEDLINE | ID: mdl-37690124

ABSTRACT

PURPOSE: Secondary dose calculation (SDC) with an independent algorithm is one option to perform plan-specific quality assurance (QA). While measurement-based QA can potentially detect errors in plan delivery, the dose values are typically only compared to calculations on homogeneous phantom geometries instead of patient CT data. We analyzed the sensitivity and specificity of an SDC software by purposely introducing different errors and determined thresholds for optimal decisions. METHODS: Thirty head and neck VMAT plans and 30 modifications of those plans, including errors related to density and beam modelling, were recalculated using RadCalc with a Monte Carlo algorithm. Decision thresholds were obtained by receiver operating characteristics (ROC) analysis. For comparison, measurement-based QA using the ArcCHECK phantom was carried out and evaluated in the same way. RESULTS: Despite optimized decision thresholds, none of the systems was able to reliably detect all errors. ArcCHECK analysis using a 2%/2 mm criterion with a threshold of 91.1% had an area under the curve (AUC) of 0.80. Evaluating differences in recalculated and planned DVH parameter of the target structures in RadCalc with a 2% threshold an AUC of 0.86 was achieved. Out-of-field deviations could be attributed to weaknesses in the beam model. CONCLUSIONS: Secondary dose calculation with RadCalc is an alternative to established measurement-based phantom QA. Different tools catch different errors; therefore, a combination of approaches should be preferred.


Subject(s)
Radiotherapy, Intensity-Modulated , Humans , Radiotherapy Planning, Computer-Assisted , Quality Assurance, Health Care , Software , Sensitivity and Specificity , Radiotherapy Dosage
4.
Med Phys ; 50(12): 8044-8056, 2023 Dec.
Article in English | MEDLINE | ID: mdl-37646469

ABSTRACT

BACKGROUND: Beam data commissioning is a core task of radiotherapy physicists. Despite multiple detectors available, a feasible measurement program compromises between detector properties and time constraints. Therefore, it is important to understand how nonideal measurement data propagates into patient dose calculation. PURPOSE: We simulated the effects of realistic errors, due to beam commissioning with presumably nonoptimal detectors, on the resulting patient dose distributions. Additionally, the detectability of such beam commissioning errors during patient plan quality assurance (QA) was evaluated. METHODS: A clinically used beam model was re-commissioned introducing changes to depth dose curves, output factors, profiles or combinations of those. Seventeen altered beam models with incremental changes of the modelling parameters were created to analyze dose changes on simplified anatomical phantoms. Additionally, fourteen altered models incorporate changes in the order of signal differences reported for typically used detectors. Eighteen treatment plans of different types were recalculated on patient CT data sets using the altered beam models. RESULTS: For the majority of clinical plans, dose distributions in the target volume recalculated on the patient computed tomography data were similar between the original and the modified beam models, yielding global 2%/2 mm gamma pass rates above 98.9%. Larger changes were observed for certain combinations of beam modelling errors and anatomical sites, most extreme for output factor changes in a small target volume plan with a pass rate of 80.6%. Modelling an enlarged penumbra as if measured with a 0.125 cm3 ion chamber had the largest effect on the dose distribution (average pass rate of 96.5%, lowest 85.4%). On different QA phantom geometries, dose distributions between calculations with modified and unmodified models typically changed too little to be detected in actual measurements. CONCLUSION: While the simulated errors during beam modelling had little effect on most plans, in some cases changes were considerable. High-quality penumbra and small field output factor should be a main focus of commissioning measurements. Detecting modelling issues using standard patient QA phantoms is unlikely. Verification of a beam model should be performed especially for plans with high modulation and in different depths or geometries representing the variety of situations expected clinically.


Subject(s)
Radiotherapy Planning, Computer-Assisted , Radiotherapy, Intensity-Modulated , Humans , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted/methods , Computer Simulation , Radiotherapy, Intensity-Modulated/methods , Tomography, X-Ray Computed , Phantoms, Imaging
5.
Phys Med ; 113: 102662, 2023 Sep.
Article in English | MEDLINE | ID: mdl-37572393

ABSTRACT

PURPOSE: Ethos allows online adaption of radiotherapy treatment plans. Dose is calculated on synthetic computed tomographies (sCT), CT-like images generated by deforming planning CTs (pCT) onto daily cone beam CTs (CBCT) acquired during treatment sessions. Errors in sCT density distribution may lead to dose calculation errors. sCT correctness was investigated for bolus-covered surfaces. METHODS: pCTs were recorded of a slab phantom covered with bolus of different thicknesses and with air gaps introduced by spacer rings of variable diameters and heights. Treatment plans were irradiated following the adaptive workflow with different bolus configurations present in the pCT and CBCT. sCT densities were compared to those of the pCT for the same air gap size. Additionally, the neck region of an anthropomorphic phantom was imaged using a plane standard bolus versus an individual bolus adapted to the phantom's outer contour. RESULTS: Varying bolus thickness by 5 mm between pCT and CBCT was reproduced in the sCT within 2 mm accuracy. Different air gaps in pCT and CBCT resulted in highly variable bolus thickness in the sCT with a typical error of 5 mm or more. In extreme cases, air gaps were filled with bolus material density in the sCT or the phantom was unrealistically deformed near changed bolus geometries. Changes in bolus thickness and deformation also occurred in the anthropomorphic phantom. CONCLUSION: sCTs must be critically examined and included in plan-specific quality assurance. The use of tight-fitting air gap-free bolus should be preferred to increase the similarity between sCT and CBCT.


Subject(s)
Cone-Beam Computed Tomography , Radiotherapy, Intensity-Modulated , Cone-Beam Computed Tomography/methods , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted/methods , Tomography, X-Ray Computed , Radiotherapy, Intensity-Modulated/methods , Image Processing, Computer-Assisted/methods
6.
Med Phys ; 50(11): 7177-7191, 2023 Nov.
Article in English | MEDLINE | ID: mdl-37531177

ABSTRACT

BACKGROUND: The response of various detectors in the radiotherapy energy range has been investigated, especially for 6 and 10 MV energies for small fields, and is summarized in TRS-483. However, data for accelerator energies above 10 MV are sparse or unavailable for many detectors, especially for the energy of 18 MV. Small variations in field output factors for the commissioning of a treatment planning system can have a high impact on calculation of dose distributions. PURPOSE: Many studies describe an energy dependence of the response for a large number of detectors. We wanted to close the gap for the 18 MV energy regime and determined field output correction factors for different detectors at 18 MV. METHODS: An ELEKTA Versa HD accelerator at 18 MV was used together with a PTW MP3 water phantom at an SSD of 90 cm. The following detectors were examined: PTW Semiflex 31021, PinPoint 3D 31022, diode 60012, diode 60008 and microDiamond 60019, Sun Nuclear EDGE detector, IBA PFD, SFD, Razor Chamber, Razor Nano Chamber and Razor Diode, Standard Imaging Scintillator Exradin W2 1x3, W2 1x1 and Gafchromic EBT3 film. The dose response was determined at a depth of 10 cm for square fields between 0.5 and 10 cm side length. As reference data a composure of radiochromic film data for small fields ( s ≤ 3 $s\le 3$  cm) and data of all compatible chambers for larger fields ( s ≥ 3 $s\ge 3$  cm) was used. The effective field sizes of small fields were determined from profiles obtained on radiochromic film. The obtained field output correction factors obey the rules of the TRS-483 protocol. RESULTS: The W2 1x1 scintillator and the Razor Chamber showed the smallest deviations from the reference curve. The shielded diodes (diode 60008, EDGE detector) showed the highest over-response at small fields, followed by PFD, microDiamond and the unshielded diodes (diode 60012, SFD). The ionization chambers exhibited the well-known volume effect, that is, strong under-response at small fields of up to 9% for the PinPoint 3D, 7% for the Razor Chamber and up to 30% for the Semiflex detector for the smallest studied field size. The small chambers showed a polarity effect in axial orientation, especially the Razor Nano Chamber. Corrections at 18 MV are generally larger than those provided by TRS-483, continuing the trend of increasing corrections between 6 and 10 MV also at a higher accelerator energy. Only the PinPoint 3D Chamber showed a slightly smaller correction. CONCLUSIONS: Field output correction factors were determined for square field sizes between 0.5 and 10 cm at 18 MV. Most detectors needed a larger correction than at 6 and 10 MV. Thus, the use of correction factors will improve beam data for 18 MV.


Subject(s)
Radiometry , Water , Radiometry/methods , Phantoms, Imaging , Carmustine , Photons/therapeutic use , Monte Carlo Method
7.
J Appl Clin Med Phys ; 24(8): e14001, 2023 Aug.
Article in English | MEDLINE | ID: mdl-37086428

ABSTRACT

PURPOSE: Developed as a plan-specific pre-treatment QA tool, Varian portal dosimetry promises a fast, high-resolution, and integrated QA solution. In this study, the agreement between predicted fluence and measured cumulative portal dose was determined for the first 140 patient plans at our Halcyon linear accelerator. Furthermore, the capability of portal dosimetry to detect incorrect plan delivery was compared to that of a common QA phantom. Finally, tolerance criteria for verification of VMAT plan delivery with Varian portal dosimetry were derived. METHODS: All patient plans and the corresponding verification plans were generated within the Eclipse treatment planning system. Four representative plans of different treatment sites (prostate, prostate with lymphatic drainage, rectum, and head & neck) were intentionally altered to model incorrect plan delivery. Investigated errors included both systematic and random errors. Gamma analysis was conducted on both portal dose (criteria γ2%/2 mm , γ2%/1 mm , and γ1%/1 mm ) and ArcCHECK measurements (criteria γ3%/3 mm , γ3%/2 mm , and γ2%/2 mm ) with a 10% low-dose threshold. Performance assessment of various acceptance criteria for plan-specific treatment QA utilized receiver operating characteristic (ROC) analysis. RESULTS: Predicted and acquired portal dosimetry fluences demonstrated a high agreement evident by average gamma passing rates for the clinical patient plans of 99.90%, 96.64%, and 91.87% for γ2%/2 mm , γ2%/1 mm , and γ1%/1 mm , respectively. The ROC analysis demonstrated a very high capability of detecting erroneous plan delivery for portal dosimetry (area under curve (AUC) > 0.98) and in this regard outperforms QA with the ArcCHECK phantom (AUC ≈ 0.82). With the suggested optimum decision thresholds excellent sensitivity and specificity is simultaneously possible. CONCLUSIONS: Owing to the high achievable spatial resolution, portal dosimetry at the Halcyon can reliably be deployed as plan-specific pre-treatment QA tool to screen for errors. It is recommended to support the fluence integrated portal dosimetry QA by independent phantom-based measurements of a random sample survey of treatment plans.


Subject(s)
Radiotherapy, Intensity-Modulated , Male , Humans , Radiotherapy Planning, Computer-Assisted , Radiometry , Radiotherapy Dosage , Sensitivity and Specificity , Quality Assurance, Health Care
8.
Med Phys ; 49(8): 5537-5550, 2022 Aug.
Article in English | MEDLINE | ID: mdl-35717637

ABSTRACT

PURPOSE: The aim of this work was to test the implementation of small field dosimetry following TRS-483 and to develop quality assurance procedures for the experimental determination of small field output factors (SFOFs). MATERIALS AND METHODS: Twelve different centers provided SFOFs determined with various detectors. Various linac models using the beam qualities 6 MV and 10 MV with flattening filter and without flattening filter were utilized to generate square fields down to a nominal field size of 0.5 cm × 0.5 cm. The detectors were positioned at 10 cm depth in water. Depending on the local situation, the source-to-surface distance was either set to 90 cm or 100 cm. The SFOFs were normalized to the output of the 10 cm × 10 cm field. The spread of SFOFs measured with different detectors was investigated for each individual linac beam quality and field size. Additionally, linac-type specific SFOF curves were determined for each beam quality and the SFOFs determined using individual detectors were compared to these curves. Example uncertainty budgets were established for a solid state detector and a micro ionization chamber. RESULTS: The spread of SFOFs for each linac and field was below 5% for all field sizes. With the exception of one linac-type, the SFOFs of all investigated detectors agreed within 10% with the respective linac-type SFOF curve, indicating a potential inter-detector and inter-linac variability. CONCLUSION: Quality assurance on the SFOF measurements can be done by investigation of the spread of SFOFs measured with multiple detectors and by comparison to linac-type specific SFOFs. A follow-up of a measurement session should be conducted if the spread of SFOFs is larger than 5%, 3%, and 2% for field sizes of 0.5 cm × 0.5 cm, 1 cm × 1 cm, and field sizes larger than 2 cm × 2 cm, respectively. Additionally, deviations of measured SFOFs to the linac-type-curves of more than 7%, 3%, and 2% for field sizes 0.5 cm × 0.5 cm, 1 cm × 1 cm, and field sizes larger than 1 cm × 1 cm, respectively, should be followed up.


Subject(s)
Particle Accelerators , Radiometry , Photons , Uncertainty , Water
9.
J Appl Clin Med Phys ; 23(5): e13577, 2022 May.
Article in English | MEDLINE | ID: mdl-35234345

ABSTRACT

PURPOSE: Any Linac will show geometric imprecisions, including non-ideal alignment of the gantry, collimator and couch axes, and gantry sag or wobble. Their angular dependence can be quantified and resulting changes of the dose distribution predicted (Wack, JACMP 20(5), 2020). We analyzed whether it is feasible to correct geometric shifts during treatment planning. The successful implementation of such a correction procedure was verified by measurements of different stereotactic treatment plans. METHODS: Isocentric shifts were quantified for two Elekta Synergy Agility Linacs using the QualiForMed ISO-CBCT+ module, yielding the shift between kV and MV isocenters, the gantry flex and wobble as well as the positions of couch and collimator rotation axes. Next, the position of each field's isocenter in the Pinnacle treatment planning system was adjusted accordingly using a script. Fifteen stereotactic treatment plans of cerebral metastases (0.34 to 26.53 cm3 ) comprising 9-11 beams were investigated; 54 gantry and couch combinations in total. Unmodified plans and corrected plans were measured using the Sun Nuclear SRS-MapCHECK with the Stereophan phantom and evaluated using gamma analysis. RESULTS: Geometric imprecisions, such as shifts of up to 0.8 mm between kV and MV isocenter, a couch rotation axis 0.9 mm off the kV isocente,r and gantry flex with an amplitude of 1.1 mm, were found. For eight, mostly small PTVs D98 values declined more than 5% by simulating these shifts. The average gamma (2%/2 mm, absolute, global, 20% threshold) was reduced from 0.53 to 0.31 (0.32 to 0.30) for Linac 1 (Linac 2) when including the isocentric corrections. Thus, Linac 1 reached the accuracy level of Linac 2 after correction. CONCLUSION: Correcting for Linac geometric deviations during the planning process is feasible and was dosimetrically validated. The dosimetric impact of the geometric imperfections can vary between Linacs and should be assessed and corrected where necessary.


Subject(s)
Particle Accelerators , Radiosurgery , Humans , Phantoms, Imaging , Radiometry , Radiosurgery/methods , Rotation
10.
J Appl Clin Med Phys ; 22(10): 144-151, 2021 Oct.
Article in English | MEDLINE | ID: mdl-34519437

ABSTRACT

PURPOSE: A signal dependence on dose rate was reported for the ArcCHECK array due to recombination processes within the diodes. The purpose of our work was to quantify the necessary correction and apply them to quality assurance measurements. METHODS: Static 10 × 10 cm2 6-MV fields delivered by a linear accelerator were applied to the detector array while decreasing the average dose rate, that is, the pulse frequency, from 500 to 30 MU/min. An ion chamber was placed inside the ArcCHECK cavity as a reference. Furthermore, the instantaneous dose rate dependence (DRD) was studied. The position of the detector was adjusted to change the dose-per-pulse, varying the distance between the focus and the diode closest to the focus between 69.6 and 359.6 cm. Reference measurements were performed with an ion chamber placed inside a PMMA slab phantom at the same source-to-detector distances ( S D D s ) . Exponential saturation functions were fitted to the data, with different parameters to account for two generations of ArcCHECK detectors (types 2 and 3) and both DRDs. Corrections were applied to 12 volumetric modulated arc therapy plans. RESULTS: The sensitivity decreased by up to 2.8% with a decrease in average dose rate and by 9% with a decrease in instantaneous dose rate. Correcting the average DRD, the mean gamma pass rates (2%/2-mm criterion) of the treatment plans were improved by 5 percentage points (PP) for diode type 3 and 0.4 PP for type 2. Correcting the instantaneous DRD, the improvement was 8.4 PP for type 3 and 0.9 PP for type 2. CONCLUSIONS: The instantaneous DRD was identified as the prevailing effect on the diode sensitivity. We developed and validated a method to correct this behavior. The number of falsely not passed treatment plans could be considerably reduced.


Subject(s)
Radiotherapy, Intensity-Modulated , Silicon , Humans , Particle Accelerators , Quality Assurance, Health Care , Radiometry , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted
11.
Med Phys ; 48(10): e886-e921, 2021 Oct.
Article in English | MEDLINE | ID: mdl-34101836

ABSTRACT

Small-field dosimetry used in advance treatment technologies poses challenges due to loss of lateral charged particle equilibrium (LCPE), occlusion of the primary photon source, and the limited choice of suitable radiation detectors. These challenges greatly influence dosimetric accuracy. Many high-profile radiation incidents have demonstrated a poor understanding of appropriate methodology for small-field dosimetry. These incidents are a cause for concern because the use of small fields in various specialized radiation treatment techniques continues to grow rapidly. Reference and relative dosimetry in small and composite fields are the subject of the International Atomic Energy Agency (IAEA) dosimetry code of practice that has been published as TRS-483 and an AAPM summary publication (IAEA TRS 483; Dosimetry of small static fields used in external beam radiotherapy: An IAEA/AAPM International Code of Practice for reference and relative dose determination, Technical Report Series No. 483; Palmans et al., Med Phys 45(11):e1123, 2018). The charge of AAPM task group 155 (TG-155) is to summarize current knowledge on small-field dosimetry and to provide recommendations of best practices for relative dose determination in small megavoltage photon beams. An overview of the issue of LCPE and the changes in photon beam perturbations with decreasing field size is provided. Recommendations are included on appropriate detector systems and measurement methodologies. Existing published data on dosimetric parameters in small photon fields (e.g., percentage depth dose, tissue phantom ratio/tissue maximum ratio, off-axis ratios, and field output factors) together with the necessary perturbation corrections for various detectors are reviewed. A discussion on errors and an uncertainty analysis in measurements is provided. The design of beam models in treatment planning systems to simulate small fields necessitates special attention on the influence of the primary beam source and collimating devices in the computation of energy fluence and dose. The general requirements for fluence and dose calculation engines suitable for modeling dose in small fields are reviewed. Implementations in commercial treatment planning systems vary widely, and the aims of this report are to provide insight for the medical physicist and guidance to developers of beams models for radiotherapy treatment planning systems.


Subject(s)
Photons , Radiometry , International Agencies , Phantoms, Imaging
12.
J Appl Clin Med Phys ; 21(5): 56-64, 2020 May.
Article in English | MEDLINE | ID: mdl-32196950

ABSTRACT

PURPOSE: To assess the impact of isocenter shifts due to linac gantry and table rotation during cranial stereotactic radiosurgery on D98 , target volume coverage (TVC), conformity (CI), and gradient index (GI). METHODS: Winston-Lutz (WL) checks were performed on two Elekta Synergy linacs. A stereotactic quality assurance (QA) plan was applied to the ArcCHECK phantom to assess the impact of isocenter shift corrections on Gamma pass rates. These corrections included gantry sag, distance of collimator and couch axes to the gantry axis, and distance between cone-beam computed tomography (CBCT) isocenter and treatment beam (MV) isocenter. We applied the shifts via script to the treatment plan in Pinnacle 16.2. In a planning study, isocenter and mechanical rotation axis shifts of 0.25 to 2 mm were applied to stereotactic plans of spherical planning target volumes (PTVs) of various volumes. The shifts determined via WL measurements were applied to 16 patient plans with PTV sizes between 0.22 and 10.4 cm3 . RESULTS: ArcCHECK measurements of a stereotactic treatment showed significant increases in Gamma pass rate for all three measurements (up to 3.8 percentage points) after correction of measured isocenter deviations. For spherical targets of 1 cm3 , CI was most severely affected by increasing the distance of the CBCT isocenter (1.22 to 1.62). Gradient index increased with an isocenter-collimator axis distance of 1.5 mm (3.84 vs 4.62). D98 (normalized to reference) dropped to 0.85 (CBCT), 0.92 (table axis), 0.95 (collimator axis), and 0.98 (gantry sag), with similar but smaller changes for larger targets. Applying measured shifts to patient plans lead to relevant drops in D98 and TVC (7%) for targets below 2 cm3 treated on linac 1. CONCLUSION: Mechanical deviations during gantry, collimator, and table rotation may adversely affect the treatment of small stereotactic lesions. Adjustments of beam isocenters in the treatment planning system (TPS) can be used to both quantify their impact and for prospective correction of treatment plans.


Subject(s)
Particle Accelerators , Radiosurgery , Humans , Phantoms, Imaging , Prospective Studies , Radiotherapy Planning, Computer-Assisted
13.
Med Phys ; 47(3): 1327-1339, 2020 Mar.
Article in English | MEDLINE | ID: mdl-31860128

ABSTRACT

PURPOSE: The model used to calculate dose distributions in a radiotherapy treatment plan relies on the data entered during beam commissioning. The quality of these data heavily depends on the detector choice made, especially in small fields and in the buildup region. Therefore, it is necessary to identify suitable detectors for measurements in the buildup region of small fields. To aid the understanding of a detector's limitations, several factors that influence the detector signal are to be analyzed, for example, the volume effect due to the detector size, the response to electron contamination, the signal dependence on the polarity used, and the effective point of measurement chosen. METHODS: We tested the suitability of different small field detectors for measurements of depth dose curves with a special focus on the surface-near area of dose buildup for fields sized between 10 × 10 and 0.6 × 0.6 cm2 . Depth dose curves were measured with 14 different detectors including plane-parallel chambers, thimble chambers of different types and sizes, shielded and unshielded diodes as well as a diamond detector. Those curves were compared with depth dose curves acquired on Gafchromic film. Additionally, the magnitude of geometric volume corrections was estimated from film profiles in different depths. Furthermore, a lead foil was inserted into the beam to reduce contaminating electrons and to study the resulting changes of the detector response. The role of the effective point of measurement was investigated by quantifying the changes occurring when shifting depth dose curves. Last, measurements for the small ionization chambers taken at opposing biasing voltages were compared to study polarity effects. RESULTS: Depth-dependent correction factors for relative depth dose curves with different detectors were derived. Film, the Farmer chamber FC23, a 0.13 cm3 scanning chamber CC13 and a plane-parallel chamber PPC05 agree very well in fields sized 4 × 4 and 10 × 10 cm2 . For most detectors and in smaller fields, depth dose curves differ from the film. In general, shielded diodes require larger corrections than unshielded diodes. Neither the geometric volume effect nor the electron contamination can account for the detector differences. The biggest uncertainty arises from the positioning of a detector with respect to the water surface and from the choice of the detector's effective point of measurement. Depth dose curves acquired with small ionization chambers differ by over 15% in the buildup region depending on sign of the biasing voltage used. CONCLUSIONS: A scanning chamber or a PPC40 chamber is suitable for fields larger than 4 × 4 cm2 . Below that field size, the microDiamond or small ionization chambers perform best requiring the smallest corrections at depth as well as in the buildup region. Diode response changes considerably between the different types of detectors. The position of the effective point of measurement has a huge effect on the resulting curves, therefore detector specific rather than general shifts of half the inner radius of cylindrical ionization chambers for the effective point of measurement should be used. For small ionization chambers, averaging between both polarities is necessary for data obtained near the surface.


Subject(s)
Radiation Dosage , Radiotherapy Planning, Computer-Assisted/methods , Artifacts , Electrons , Film Dosimetry , Radiotherapy Dosage
14.
Med Phys ; 46(11): 5209-5215, 2019 Nov.
Article in English | MEDLINE | ID: mdl-31461533

ABSTRACT

PURPOSE: The effective point of measurement (EPOM) of cylindrical ionization chambers differs from their geometric center. The exact shift depends on chamber construction details, above all the chamber size, and to some degree on the field-size and beam quality. It generally decreases as the chamber dimensions get smaller. In this work, effective points of measurement in small photon fields of a range of cylindrical chambers of different sizes are investigated, including small chambers that have not been studied previously. METHODS: In this investigation, effective points of measurement for different ionization chambers (Farmer type, scanning chambers, micro-ionization chambers) and solid state detectors were determined by measuring depth-ionization curves in a 6 MV beam in field sizes between 2 × 2 cm2 and 10 × 10 cm2 and comparing those curves with curves measured with plane-parallel chambers. RESULTS: It was possible to average the results to one shift per detector, as the results were sufficiently independent of the studied field sizes. For cylindrical ion chambers, shifts of the EPOM were determined to be between 0.49 and 0.30 times the inner chamber radius from the reference point. CONCLUSIONS: We experimentally confirmed the previously reported decrease of the EPOM shift with decreasing detector size. Highly accurate data for a large range of detectors, including new very small ones, were determined. Thus, small chambers noticeably differ from the 0.5-times to 0.6-times the inner chamber radius recommendations in current dosimetry protocols. The detector-individual EPOMs need to be considered for measurements of depth-dose curves.


Subject(s)
Radiometry/instrumentation , Photons , Radiometry/standards , Reference Standards
15.
Z Med Phys ; 29(1): 22-30, 2019 Feb.
Article in English | MEDLINE | ID: mdl-30049551

ABSTRACT

PURPOSE: The characteristics of radiation detectors have to be assessed for dosimetry in the presence of magnetic fields, i.e. in conditions found in combined machines for magnetic resonance imaging and radiotherapy. While a lot of attention is directed toward correction factors for absolute dosimetry in magnetic fields, relative dose measurements are an equally important task to be performed. There is a need to experimentally analyze detector response differences in the build-up region in the presence of a transverse magnetic field. METHODS: Depth dose curves with different detectors (microDiamond PTW 60019, unshielded diode PTW 60012, ionization chamber PTW Semiflex 31010 and EBT3 film) were acquired for a beam quality of 6MV in an 8×10cm2 field at SSD 110cm with and without a transverse magnetic field of up to 1.1T. For these experiments, an electromagnet was placed in front of a conventional linear accelerator of the type Elekta Precise. The detectors were positioned in a water phantom fitting between the poles of the electromagnet. The beam entered through a 0.3mm thin PMMA foil window, which enabled measurements even close to the surface. Ratios of the response with and without the magnetic field for different detectors were investigated. The film served as a reference. RESULTS: Changes in the depth dose curve near the surface due to the magnetic field were not correctly reproduced by all detectors. EBT3 film and the microDiamond detector agreed up to the surface. The diode showed up to 2% deviation from the film in the build-up region, but it could still be considered within the uncertainties. However, the curves obtained with the ionization chamber showed up to 6% deviation from the film and even completely different trends in the surface-near region. At depths larger than 2cm, there were no noticeable differences between the different detectors for relative depth dose curves. CONCLUSIONS: At the descending part of the depth dose, the tested detectors did not show artifacts within the magnetic field. However, air-filled ionization chambers cannot be recommended for relative dosimetry in magnetic fields near the surface. Diamond detectors might be a suitable alternative and future investigations should concentrate on the performance of such detectors.


Subject(s)
Magnetic Fields , Radiometry/instrumentation , Radiometry/methods , Artifacts , Particle Accelerators , Phantoms, Imaging , Photons
16.
Phys Med Biol ; 64(1): 015010, 2018 12 21.
Article in English | MEDLINE | ID: mdl-30524066

ABSTRACT

The 3D-dose module (3DVH) of the ArcCHECK-phantom reconstructs the dose distribution in the phantom volume and transfers it to the patient geometry. Our aim was to evaluate the 3DVH-reconstructed dose systematically building up from simple to complex cases. Therefore, the influence of different field sizes without and with blocking the isocenter was tested. The dose distributions of different radiation techniques, error-free and error-induced VMAT-plans were verified by measuring with films and other detectors in the phantom. It was checked how the inclusion of the dose measured separately in the ArcCHECK-isocenter affects the reconstruction. Thus it was also investigated which detector should be used for the dosimetry in the isocenter. Without including the isocentrically measured dose, the reconstruction for the smallest field (2 × 2 cm²) was 5% (6 MV) and 3.7% (10 MV) higher than measured with an ionization chamber. With increasing field size, the deviation decreased. For fields with blocked isocenters, the reconstructed dose was between -10.6% and -24% lower than determined with a microDiamond. Measurements with the Semiflex of the spinal plan resulted in higher doses than calculated by the treatment planning system (TPS) and measured with the film and the other detectors. Through the inclusion of the isocentric dose in the reconstruction its accordance with the film increased mostly. With exception of an error-induced head and neck plan, the induced errors in the reconstructed dose volume histogram became visible, but were underestimated. With the 3DVH-algorithm not every induced-error was detected. The 3DVH underestimated the dose in blocked areas. To protect organs at risk (OAR), these are often blocked. Consequently, there is a risk that a clinical decision is based on a 3DVH that underestimated the dose for the OAR. We recommend including the isocentric dose in the reconstruction. The detector used for the isocentric measurements should be carefully chosen.


Subject(s)
Radiotherapy Planning, Computer-Assisted/methods , Algorithms , Humans , Phantoms, Imaging , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted/standards
17.
Med Phys ; 45(12): 5622-5630, 2018 Dec.
Article in English | MEDLINE | ID: mdl-30291724

ABSTRACT

PURPOSE: The integral quality monitor (IQM) transmission detector is a wedge-shaped large area ionization chamber that reports a position-weighted dose area product for each control point of an IMRT or VMAT plan. In this study, the accuracy of the signal prediction is verified for the Synergy Agility MLC. Tolerance criteria for VMAT plan verification with the IQM were obtained from the observed sensitivity for the detection of incorrectly delivered plans. METHODS: The predicted IQM signal was compared to the measured signal recorded for a set of 30 VMAT plans for each beam quality of 6 and 10 MV. The system's capability to detect incorrectly delivered plans was tested by measuring altered plans containing small, random deviations. In addition, the observed deviations were related to measurements performed with a second QA phantom. RESULTS: The cumulative IQM signal per arc deviated from the respective calculation on average by -0.48% (6 MV) and +0.21% (10 MV) with a standard deviation of 1.08% in both cases, suggesting a 2% warning and 3% action threshold as plan acceptance criteria. This choice was confirmed by the optimum threshold of 2.5% obtained via receiver operating characteristic (ROC) analysis. Reproducibility of individual control points in multiply measured plans was low (on average 7% for 1SD) and thus, segment-by-segment comparison was impractical. A suitable criterion to resolve the angular distribution of the plan was identified by binning three to five control points as a running average. While the correlation between IQM signal deviations and gamma passing rates obtained with the ArcCHECK phantom was low for clinical plans, it was apparent for erroneous plans. Binning led to even higher sensitivity to errors. CONCLUSIONS: The IQM was able to detect induced errors at least as reliable as the standard phantom and showed the potential to be used in pretreatment plan verification to ensure the correct plan transfer and delivery. However, there is no direct correlation between the IQM signal deviation and DVH metrics, so the IQM should be primarily used to screen for errors. Finer diagnostics should then be carried out using a different phantom.


Subject(s)
Medical Errors , Radiometry/instrumentation , Radiotherapy Planning, Computer-Assisted/methods , Radiotherapy, Intensity-Modulated , Phantoms, Imaging
18.
J Appl Clin Med Phys ; 19(6): 274-281, 2018 Nov.
Article in English | MEDLINE | ID: mdl-30298980

ABSTRACT

Relative dose measurements with small ionization chambers in combination with an electrometer placed in the treatment room ("internal electrometer") show a large dependence on the polarity used. While this was observed previously for percent depth dose curves (PDDs), the effect has not been understood or preventable. To investigate the polarity dependence of internal electrometers used in conjunction with a small-volume ionization chamber, we placed an internal electrometer at a distance of 1 m from the isocenter and exposed it to different amounts of scattered radiation by varying the field size. We identified irradiation of the electrometer to cause a current of approximately -1 pA, regardless of the sign of the biasing voltage. For low-sensitivity detectors, such a current noticeably distorts relative dose measurements. To demonstrate how the current systematically changes PDDs, we collected measurements with nine ionization chambers of different volumes. As the chamber volume decreased, signal ratios at 20 and 10 cm depth (M20/M10) became smaller for positive bias voltage and larger for negative bias voltage. At the size of the iba CC04 (40 mm³) the difference of M20/M10 was around 1% and for the smallest studied chamber, the iba CC003 chamber (3 mm³), around 7% for a 10 × 10 cm² field. When the electrometer was moved further from the source or shielded, the additional current decreased. Consequently, PDDs at both polarities were brought into alignment at depth even for the 3 mm³ ionization chamber. The apparent polarity effect on PDDs and lateral beam profiles was reduced considerably by shielding the electrometer. Due to normalization the effect on output values was low. When measurements with a low-sensitivity probe are carried out in conjunction with an internal electrometer, we recommend careful monitoring of the particular setup by testing both polarities, and if deemed necessary, we suggest shielding the electrometer.


Subject(s)
Electronics/instrumentation , Particle Accelerators/instrumentation , Radiometry/instrumentation , Scattering, Radiation , Calibration , Electronics/methods , Humans , Radiation Dosage
19.
Phys Med Biol ; 63(7): 075002, 2018 03 22.
Article in English | MEDLINE | ID: mdl-29494344

ABSTRACT

The purpose of this work is the development of a robust and reliable three-dimensional (3D) Cartesian imaging technique for fast and flexible retrospective 4D abdominal MRI during free breathing. To this end, a non-uniform quasi random (NU-QR) reordering of the phase encoding (k y -k z ) lines was incorporated into 3D Cartesian acquisition. The proposed sampling scheme allocates more phase encoding points near the k-space origin while reducing the sampling density in the outer part of the k-space. Respiratory self-gating in combination with SPIRiT-reconstruction is used for the reconstruction of abdominal data sets in different respiratory phases (4D-MRI). Six volunteers and three patients were examined at 1.5 T during free breathing. Additionally, data sets with conventional two-dimensional (2D) linear and 2D quasi random phase encoding order were acquired for the volunteers for comparison. A quantitative evaluation of image quality versus scan times (from 70 s to 626 s) for the given sampling schemes was obtained by calculating the normalized mutual information (NMI) for all volunteers. Motion estimation was accomplished by calculating the maximum derivative of a signal intensity profile of a transition (e.g. tumor or diaphragm). The 2D non-uniform quasi-random distribution of phase encoding lines in Cartesian 3D MRI yields more efficient undersampling patterns for parallel imaging compared to conventional uniform quasi-random and linear sampling. Median NMI values of NU-QR sampling are the highest for all scan times. Therefore, within the same scan time 4D imaging could be performed with improved image quality. The proposed method allows for the reconstruction of motion artifact reduced 4D data sets with isotropic spatial resolution of 2.1 × 2.1 × 2.1 mm3 in a short scan time, e.g. 10 respiratory phases in only 3 min. Cranio-caudal tumor displacements between 23 and 46 mm could be observed. NU-QR sampling enables for stable 4D-MRI with high temporal and spatial resolution within short scan time for visualization of organ or tumor motion during free breathing. Further studies, e.g. the application of the method for radiotherapy planning are needed to investigate the clinical applicability and diagnostic value of the approach.


Subject(s)
Imaging, Three-Dimensional/methods , Magnetic Resonance Imaging/methods , Movement , Phantoms, Imaging , Radiography, Abdominal , Respiratory-Gated Imaging Techniques/methods , Adult , Artifacts , Feasibility Studies , Female , Healthy Volunteers , Humans , Male , Middle Aged , Respiration , Retrospective Studies
20.
Med Phys ; 45(2): 898-907, 2018 Feb.
Article in English | MEDLINE | ID: mdl-29193168

ABSTRACT

PURPOSE: Different detector properties will heavily affect the results of off-axis measurements outside of radiation fields, where a different energy spectrum is encountered. While a diode detector would show a high spatial resolution, it contains high atomic number elements, which lead to perturbations and energy-dependent response. An ionization chamber, on the other hand, has a much smaller energy dependence, but shows dose averaging over its larger active volume. We suggest a way to obtain spatial energy response corrections of a detector independent of its volume effect for profiles of arbitrary fields by using a combination of two detectors. METHODS: Measurements were performed at an Elekta Versa HD accelerator equipped with an Agility MLC. Dose profiles of fields between 10 × 4 cm² and 0.6 × 0.6 cm² were recorded several times, first with different small-field detectors (unshielded diode 60012 and stereotactic field detector SFD, microDiamond, EDGE, and PinPoint 31006) and then with a larger volume ionization chamber Semiflex 31010 for different photon beam qualities of 6, 10, and 18 MV. Correction factors for the small-field detectors were obtained from the readings of the respective detector and the ionization chamber using a convolution method. Selected profiles were also recorded on film to enable a comparison. RESULTS: After applying the correction factors to the profiles measured with different detectors, agreement between the detectors and with profiles measured on EBT3 film was improved considerably. Differences in the full width half maximum obtained with the detectors and the film typically decreased by a factor of two. Off-axis correction factors outside of a 10 × 1 cm² field ranged from about 1.3 for the EDGE diode about 10 mm from the field edge to 0.7 for the PinPoint 31006 25 mm from the field edge. The microDiamond required corrections comparable in size to the Si-diodes and even exceeded the values in the tail region of the field. The SFD was found to require the smallest correction. The corrections typically became larger for higher energies and for smaller field sizes. CONCLUSIONS: With a combination of two detectors, experimentally derived correction factors can be obtained. Application of those factors leads to improved agreement between the measured profiles and those recorded on EBT3 film. The results also complement so far only Monte Carlo-simulated values for the off-axis response of different detectors.


Subject(s)
Radiometry/instrumentation
SELECTION OF CITATIONS
SEARCH DETAIL